510(k) Data Aggregation

    K Number
    K220149

    Validate with FDA (Live)

    Date Cleared
    2022-04-19

    (90 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The GR10X Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This system is not intended for mammographic applications.

    Device Description

    This device consists of X-ray control, high-voltage generator, X-ray tube support, irradiation equipment, detectors, patient photography stand, digital imaging device, viewer SW, etc. The detectors and the viewer SW are cleared under the following 510ks.

    • VIVIX-S VW(K200418) . Model Names: FXRD-2530VW, FXRD-2530VW PLUS, FXRD-3643VW, FXRD-3643VW PLUS, FXRD-4343VW, FXRD-4343VW PLUS, with imaging areas of 25cm x 30cm, 36cm x 43cm, 43cm x 43cm, respectively.
      A high frequency inverter (Inverter) X-sensor voltage device designed to generate X-rays by combination of tube voltage, tube current, irradiation time, etc. so that it can be filmed at various angles for diagnosis of the patient's skeletal, respiratory, and urinary systems. The digital imaging system is used to obtain the images taken by the X-ray unit as radiation from the X-ray unit passes through the human body and is transmitted by the detector intercepts x-ray photons and the scintillator emits visible spectrum photons that illuminate an array of photo (a-SI)detectors that create electrical signals. After the electrical signals are generated, it is converted to digital value, and the Software, VXvue, acquires and processes the data values from the detector. The resulting digital images will be displayed on monitors.
    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the GR10X Digital X-ray Imaging System (Models GR10X-40K, GR10X-50K). The submission aims to demonstrate substantial equivalence to a predicate device (Ysio, K081722).

    Based on the document, here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a quantitative format for the clinical study. Instead, the "acceptance criteria" for the clinical study is an overarching statement that the device "provides images of equivalent diagnostic capability to the predicate devices."

    The reported device performance from the clinical study is:

    • "the study confirmed that the GR10X Digital X-ray Imaging System provides images of equivalent diagnostic capability to the predicate devices, the Yiso and its results demonstrate substantial equivalence."

    For non-clinical performance (technical specifications), here's a table comparing the subject device's detectors (VIVIX-S VW Series, VIVIX-S 1717V) to the predicate's detector (Trixell Pixium 4343RCE):

    Technical SpecificationsPredicate Device (Trixell Pixium 4343RCE)Subject Device (VIVIX-S VW Series, VIVIX-S 1717V) Reported PerformanceAcceptance Criteria (Implicit)Comparison Result
    Dimensions423.3mm x 425.4 mmVIVIX-S 4343VW: 460 mm x 460 mm VIVIX-S 3643VW: 384mm × 460mm VIVIX-S 2530VW: 287mm x 350mm VIVIX-S 1717V: 460mm x 460mmSubstantially EquivalentSubstantially Equivalent
    Resolution2860 x 2874 pixelsVIVIX-S 4343VW: 3072 x 3072 pixels VIVIX-S 3643VW: 2560 x 3072 pixels VIVIX-S 2530VW: 2048 x 2560 pixels VIVIX-S 1717V: 3072 x 3072 pixelsSubstantially EquivalentSubstantially Equivalent
    Pixel size148 µmVIVIX-S 4343VW: 140µm VIVIX-S 3643VW: 140µm VIVIX-S 2530VW: 124µm VIVIX-S 1717V: 140µmSubstantially EquivalentSubstantially Equivalent
    Semiconductor MaterialAmorphous silicon, a-SiAmorphous silicon, a-SiSubstantially EquivalentSubstantially Equivalent
    ScintillatorCesium iodide (CsI)Cesium iodide (CsI), Gadolinium Oxide (Gadox)Substantially EquivalentSubstantially Equivalent
    Acquisition Depth16 bit16 bitSubstantially EquivalentSubstantially Equivalent
    DQE @ 0.05 lp/mm (2 µGy)67%FXRD-4343VAW: 47% FXRD-4343VAW PLUS: 62% FXRD-3643VAW: 45.5% FXRD-3643VAW PLUS: 61% FXRD-2530VAW: 49% FXRD-2530VAW PLUS: 61% FXRD-1717NAW: 50% FXRD-1717NBW: 27%Substantially EquivalentSubstantially Equivalent
    MTF @ 1 lp/mm62%FXRD-4343VAW: 76% FXRD-4343VAW PLUS: 60% FXRD-3643VAW: 74% FXRD-3643VAW PLUS: 61% FXRD-2530VAW: 76% FXRD-2530VAW PLUS: 60% FXRD-1717NAW: 66% FXRD-1717NBW: (value cut off)Substantially EquivalentSubstantially Equivalent

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions a "single-blinded concurrence study" as a "clinical test" but does not provide any specific quantitative details regarding the sample size of the test set, or the provenance (country of origin, retrospective/prospective nature) of the data used in this clinical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not specify the number of experts used or their qualifications for establishing ground truth in the clinical study. It only refers to a "single-blinded concurrence study."

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document states "a single-blinded concurrence study was conducted." It does not provide details on the specific adjudication method used (e.g., 2+1, 3+1, or if it was based on individual expert assessment without formal adjudication). The term "concurrence" implies agreement among readings, but the method for achieving or resolving discrepancies is not described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The document describes a "single-blinded concurrence study" comparing the GR10X system to predicate devices. This study aims to confirm "equivalent diagnostic capability." It is not explicitly described as a Multi-Reader Multi-Case (MRMC) comparative effectiveness study involving AI assistance for human readers. The device is an X-ray imaging system, not an AI-powered diagnostic aid for human readers. Therefore, no effect size of human readers improving with AI vs without AI assistance is reported or relevant based on the information provided.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to an X-ray imaging system, not an AI algorithm. Therefore, the concept of "standalone (algorithm only without human-in-the-loop performance)" does not apply in the context of this device's evaluation. The evaluation is focused on the image quality and diagnostic capability of the imaging system itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document mentions a "single-blinded concurrence study" to confirm "equivalent diagnostic capability." While this implies expert interpretation, it does not explicitly state the type of ground truth used (e.g., expert consensus, pathology, outcomes data). The phrase "concurrence study" suggests that the comparison was based on human expert interpretations of images from both the subject and predicate devices.

    8. The sample size for the training set

    The document describes a clinical study for product clearance and does not mention any "training set." This type of submission (510k for a general X-ray system) typically involves demonstrating performance against a predicate device, and not machine learning model training. Therefore, there is no information on a training set sample size.

    9. How the ground truth for the training set was established

    As there is no mention of a "training set" in the context of a machine learning model, this question is not applicable based on the provided text.