510(k) Data Aggregation

    K Number
    K220080

    Validate with FDA (Live)

    Device Name
    CogNet QmTRIAGE
    Manufacturer
    Date Cleared
    2022-09-29

    (262 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MedCognetics (CogNet) QmTRIAGE™ software is a passive notification-only, parallel-workflow software tool used by MQSA qualified interpreting physicians to prioritize patients with suspicious findings in the medical care environment. QmTRIAGE™ utilizes an artificial intelligence algorithm to analyze 2D FFDM screening mammograms and flags those that are suggestive of the presence of at least one suspicious finding at the exam level. QmTRIAGE™ produces an exam level output to a PACS/Workstation for flagging the suspicious study and allows for worklist prioritization.

    MQSA qualified interpreting physicians are responsible for reviewing each exam on a display approved for use in mammography, according to the current standard of care. The OmTRIAGE™ device is limited to the categorization of exams, does not provide any diagnostic information beyond triage and prioritization, does not remove images from the interpreting physician's worklist, and should not be used in lieu of full patient evaluation, or relied upon to make or confirm diagnosis.

    The QmTRIAGE™ device is intended for use with complete 2D FFDM mammography exams acquired using validated FFDM systems only.

    Device Description

    The MedCognetics (CogNet) QmTRIAGE is a non-invasive computer-assisted triage and notification software as a medical device (SaMD) that analyzes 2D FFDM screening mammograms using a machine learning algorithm and notifies a PACS/workstation of the presence of findings suspicious of cancer in a study. The passive-notification enables radiologists to prioritize their worklist and assists them in viewing prioritized studies using the standard PACS or workstation viewing software. The device aim is to aid in the prioritization and triage of radiological medical images only. It is a software tool for MQSA interpreting physicians reading mammograms and does not replace complete evaluation according to the standard of care.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the MedCognetics CogNet QmTRIAGE device, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    CriteriaAcceptance Criteria (Implied)Reported Device Performance
    AUROCHigh AUROC value for overall performance.0.9569 (95% CI: 0.9364-0.9738)
    SensitivityExceed the standard of care (e.g., BCSC study reported values).87%
    SpecificityExceed the standard of care (e.g., BCSC study reported values).89%
    Triage AccuracyDemonstrated accuracy across various cohorts (lesion type, breast density, age, race).Measured and validated. Specific values not provided in this summary, but the study notes confounding factors were considered.

    Note: The document explicitly states the reported Sensitivity (87%) and Specificity (89%) "exceeded the standard of care as reported in the Breast Cancer Surveillance Consortium (BCSC) study," implying this was the acceptance benchmark.

    Study Details

    1. Sample Size used for the test set and data provenance:

      • Sample Size: 800 anonymized 2D FFDM mammograms.
      • Data Provenance: Retrospective cohort from various countries, specifically mentioned USA and Germany. (Training data was obtained from various sites worldwide including North America, South America, Europe, Africa, and Southeast Asia, but the test set is specified as USA and Germany.)
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not specify the number of experts used to establish ground truth for the test set.
      • Qualifications: The ground truth for cancer cases was based on biopsy confirmation, and for negative cases (BI-RADS 1 and 2), it was established by a two-year follow-up of a negative diagnosis. While MQSA qualified interpreting physicians are mentioned in the intended use for reviewing exams, their direct role in establishing ground truth for the test set is not explicitly detailed.
    3. Adjudication method for the test set:

      • The document does not explicitly state an adjudication method (e.g., 2+1, 3+1). The ground truth relies on objective medical outcomes: biopsy confirmation for positive cases and 2-year follow-up for negative cases.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not described. The study focuses on the standalone performance of the AI algorithm for triage. The device is described as a "passive notification-only, parallel-workflow software tool" to "prioritize patients," not as an aid that directly impacts a reader's diagnostic performance measured against a baseline.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance study was done. The reported AUROC, Sensitivity, and Specificity values are for the algorithm's performance independent of human-in-the-loop interaction. The device's role is to "prioritize their worklist and assists them in viewing prioritized studies," implying its standalone functionality in flagging.
    6. The type of ground truth used:

      • Pathology/Biopsy: For positive cases (399 cases positive for cancer), ground truth was established by biopsy confirmation.
      • Outcomes Data/Clinical Follow-up: For negative cases (401 cases negative for breast cancer, BI-RADS 1 and 2), ground truth was established by a two-year follow-up of a negative diagnosis.
    7. The sample size for the training set:

      • The exact sample size for the training set is not provided in the summary. It only states that "Data sets used for training the algorithm were independent of the testing datasets and were obtained from various sites worldwide."
    8. How the ground truth for the training set was established:

      • The document does not explicitly detail how the ground truth for the training set was established. It only mentions the training data was obtained from various sites worldwide and was independent of the testing dataset. Given the nature of breast cancer screening AI, it can be inferred that similar methods (biopsy, follow-up, or expert review) would have been used, but it's not stated.