Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K220064

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-07-22

    (193 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SYNERGY MARBLE is a hand-held device for over-the counter aesthetic purposes. The device red light and amber light are intended for the use in treating wrinkles on the face, and the blue light is intended for the treatment of the mild to moderate inflammatory acne.

    Device Description

    The SYNERGY MARBLE is an over-the-counter, battery powered hand-held light emitting diode (LED) device that emits light energy in the red, blue and amber spectrum for the treatment of wrinkles or mild to moderate inflammatory acne on the face. The SYNERGY MARBLE components include an applicator, a charging station and an adaptor. The treatment surface is applied directly to the skin to ensure consistent administration of light during each treatment. The device does not contain any user repairable components. The device is sold as Over the Counter (OTC).

    AI/ML Overview

    The provided 510(k) summary for the SYNERGY MARBLE (Model: Opasm) device does NOT include a robust clinical study or performance-based acceptance criteria for its intended uses (treating wrinkles and mild to moderate inflammatory acne).

    Instead, the submission relies on non-clinical testing and a demonstration of substantial equivalence to previously cleared predicate devices. Therefore, a table of acceptance criteria and reported device performance with associated study details cannot be fully constructed as requested.

    However, based on the information provided, we can infer some "acceptance criteria" through the lens of substantial equivalence and present the non-clinical testing conducted.


    Inferred Acceptance Criteria and Reported Device Performance (Based on Substantial Equivalence and Non-Clinical Testing)

    Acceptance Criteria (Inferred from Predicate & Standards)Reported Device Performance (SYNERGY MARBLE)
    Safety:
    Electrical Safety (IEC 60601-1)Conforms to ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009(R)2012, and A2:2010(R)2012 (Consolidated Text)
    Electromagnetic Compatibility (IEC 60601-1-2)Conforms to IEC 60601-1-2 Edition 4.0 2014-02
    Usability (IEC 60601-1-6)Conforms to IEC 60601-1-6 Edition 3.1 2013-10
    Home Healthcare Environment (IEC 60601-1-11)Conforms to IEC 60601-1-11 Edition 2.0 2015-01
    Photobiological Safety (IEC 62471)Conforms to IEC 62471 First edition 2006-07
    Battery Safety (IEC 62133-2)Conforms to IEC 62133-2, Edition1.0 2017-02
    Biocompatibility:
    In Vitro Cytotoxicity (ISO 10993-5)Conforms to ISO 10993-5:2009/(R) 2014
    Irritation and Skin Sensitization (ISO 10993-10)Conforms to ISO 10993-10 Third Edition 2010-08-01
    Systemic Toxicity (ISO 10993-11)Conforms to ISO 10993-11 Third edition 2017-09
    Performance (Inferred from Substantial Equivalence):
    Indications for Use"SYNERGY MARBLE is a hand-held device for over-the counter aesthetic purposes. The device red light and amber light are intended for the use in treating wrinkles on the face, and the blue light is intended for the treatment of the mild to moderate inflammatory acne."
    Power and WavelengthsRed (625nm±3nm), Blue (465nm±3nm), Amber (605nm±3nm)
    Energy DensityRed: 55.3mW/cm²±10%Blue: 48.2mW/cm²±10%Amber: 38.3mW/cm²±10%
    Treatment Time3 minutes per target area; 2 treatments per week for 6 weeks
    Product CodeOHS, OLP (Same as predicates)
    Regulation Number21 CFR 878.4810 (Same as predicates)
    Device ClassII (Same as predicates)

    Study Details (Based on the provided 510(k) Summary)

    The provided document describes non-clinical testing for safety and performance related to electrical safety, EMC, usability, photobiological safety, battery safety, and biocompatibility. It does NOT describe a clinical study to directly prove the device meets acceptance criteria for its intended therapeutic effects (treating wrinkles or acne).

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not applicable. The document refers to non-clinical bench testing and conformity to international consensus standards, rather than a test set of patient data.
      • Data Provenance: The data provenance is from laboratory bench testing (non-clinical) and conformity assessments against international standards. No specific country of origin is mentioned for these lab tests, but the manufacturer is YRS Group Inc. (US) and the consultant is from Shanghai, China.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. As this was non-clinical testing against established standards, there was no "ground truth" established by experts in a clinical context. Results were compared against the requirements of the standards themselves.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. There was no clinical test set requiring adjudication.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. This device is a light-based therapy device, not an AI-assisted diagnostic or therapeutic device requiring human reader assessment.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not applicable. This device is a direct-use therapy device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the non-clinical testing, the "ground truth" was defined by the specific requirements and limits set forth in the international consensus standards (e.g., maximum current leakage for electrical safety, specific wavelengths and energy densities, cytotoxicity limits). There was no clinical ground truth established for effectiveness.
    7. The sample size for the training set:

      • Not applicable. This review concerns non-clinical testing and substantial equivalence, not a machine learning model that requires a training set.
    8. How the ground truth for the training set was established:

      • Not applicable.

    Conclusion from document:

    The SYNERGY MARBLE device gained clearance by demonstrating substantial equivalence to existing legally marketed predicate devices (K202055 - Looper and K190443 - MMSphere™) through a comparison of technological characteristics and intended uses, alongside conformity to relevant non-clinical safety (electrical, EMC, photobiological, battery, usability, home use) and biocompatibility standards. The submission explicitly states: "The conclusion drawn from the non-clinical tests is that the device is as safe, as effective, and performs as well as the legally marketed predicate device K202055 and reference device K190443." This implies that clinical effectiveness for wrinkle reduction and acne treatment is primarily supported by the established effectiveness of the predicate devices, rather than a new clinical study.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1