    K Number
    K214076

    Validate with FDA (Live)

    Device Name
    EsCem
    Manufacturer
    Date Cleared
    2022-12-19

    (357 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Final cementation of all-ceramic, metal-based inlays, onlays, posts, crown and bridge

    Device Description

    EsCem is a dual-cured resin cement that can be self-etched and self-adhesive. It is classified as type 2 and class 3 according to ISO 4049. It is easy to handle with 'one step' system that does not need pre-treatment process, and it is easy to remove excess cement, so it is possible to perform quick operation. It is also easy to identify because it has a radio-opacity(1.2mm Al). And it offers various shades for aesthetics and color stable.

    AI/ML Overview

    This document describes the performance testing for a dental cement called EsCem, manufactured by SPIDENT Co., Ltd. The purpose of the testing is to demonstrate the substantial equivalence of EsCem to a legally marketed predicate device, G-CEM LinkAce (K120243), manufactured by GC America Inc.

    1. Acceptance Criteria and Device Performance

    The performance testing was conducted according to various sections of the ISO 4049:2019, Dentistry Polymer-based restorative materials standard. The table below lists the acceptance criteria and the reported device performance for EsCem. Since specific numerical results for EsCem are not provided in the document (only whether it met the requirement), the "Reported Device Performance" column reflects that it met the acceptance criteria. For comparison, the document states that both the subject device and the predicate device meet the requirements of ISO 4049.

    PerformanceTest methodAcceptance RequirementReported Device Performance (EsCem)
    Film thicknessISO 4049:2019 - 7.5 Measurement of film thickness of luting materialsShall be no greater than 50 μm.Met requirement
    Working timeISO 4049:2019 - 7.7 Working time, Class 1 and Class 3 luting materialsDuring its formation there shall be no detectable change in its homogeneity.Met requirement
    Setting timeISO 4049:2019 - 7.8 Setting time, Class 1 and Class 3 materialsShall be no more than 10 min (600 sec).Met requirement
    Flexural strengthISO 4049:2019 - 7.11 Flexural strengthShall be equal to or greater than 50 MPa.Met requirement
    Water sorptionISO 4049:2019 - 7.12 Water sorption and solubilityShall be equal to or less than 40 µg/mm³.Met requirement
    SolubilityISO 4049:2019 - 7.12 Water sorption and solubilityShall be equal to or less than 7.5 µg/mm³.Met requirement
    Radio-opacityISO 4049:2019 - 7.14 Radio-opacityShall be equal to or greater than that of the same thickness of aluminum.Met requirement (1.2mm Al stated)
    Sensitivity to lightISO 4049:2019 - 7.9 Sensitivity to light, Class 2 materialsThe material shall remain physically homogeneous.Met requirement
    Color stabilityISO 4049:2019 - 7.13 Shade and colour stability after irradiation and water sorptionNo more than a slight change in color shall be observed.Met requirement

    2. Sample Size for Test Set and Data Provenance

    The document does not specify the exact sample sizes used for each performance test (e.g., number of specimens for film thickness, flexural strength, etc.). The provenance of the data is retrospective bench testing, conducted by SPIDENT Co., Ltd. in South Korea.

    3. Number of Experts and Qualifications for Ground Truth

    This study is a bench performance testing for a dental cement, not an AI/ML medical device. Therefore, the concept of "ground truth established by experts" in the context of interpretation (e.g., radiological reads) does not apply. The ground truth for this device is based on established international consensus standards (ISO 4049:2019) for physical and chemical properties of dental materials.

    4. Adjudication Method for Test Set

    Not applicable, as this is bench testing against a standard, not a human reader study requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is a bench study comparing a new device against a standard and a predicate device's performance through physical and chemical property testing. It is not an MRMC study comparing human reader performance with or without AI assistance.

    6. Standalone (Algorithm Only) Performance

    Not applicable. This is a physical dental device (cement), not an algorithm or software.

    7. Type of Ground Truth Used

    The ground truth used for this study is based on established industry standards, specifically ISO 4049:2019 (Dentistry Polymer-based restorative materials). This standard defines the acceptable ranges and properties for dental cements. Additionally, the comparison to a legally marketed predicate device (G-CEM LinkAce) serves as a comparative ground for substantial equivalence.

    8. Sample Size for Training Set

    Not applicable. This is a physical dental device subjected to bench testing, not an AI/ML algorithm that requires a training set.

    9. How Ground Truth for Training Set was Established

    Not applicable, as there is no training set for this type of device.