510(k) Data Aggregation

    K Number
    K214056

    Validate with FDA (Live)

    Date Cleared
    2022-09-20

    (267 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.

    Device Description

    The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue/purple colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: S, M, L, XL. The subject device is non-sterile.

    AI/ML Overview

    The provided document describes the acceptance criteria and the results of a study for "Disposable Nitrile Inspection Gloves (Tested for Use with Chemotherapy Drugs)".

    Here's the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance (Blue Gloves)Reported Device Performance (Purple Gloves)
    ASTM D6319Physical Dimensions TestLength(mm): S: ≥220; M/L/XL: ≥230.(S: ≥220; M/L/XL: ≥230) - Pass(S: ≥220; M/L/XL: ≥230) - Pass
    Width(mm): S: 80±10; M: 95±10; L: 110±10; XL: 120±10.S: 84-86/Pass; M: 96-98/Pass; L: 104-106/Pass; XL: 114-116/PassS: 88-90/Pass; M: 99-100/Pass; L: 102-104/Pass; XL: 115-117/Pass
    Finger Thickness: ≥0.05 mm0.10-0.13 mm / Pass0.12-0.15 mm / Pass
    Palm Thickness: ≥0.05 mm0.06-0.09 mm / Pass0.07-0.09 mm / Pass
    ASTM D5151Watertightness Test for Detection of HolesAQL 2.50/125 / Pass0/125 / Pass
    ASTM D6124Powder Content< 2.0 mg0.3 mg / Pass0.1 mg / Pass
    ASTM D412Physical properties (Before Aging)Tensile Strength: ≥14 MPa15.3-19.0 MPa30.0-37.8 MPa
    Ultimate Elongation: ≥500%521-569%505-539%
    ASTM D412Physical properties (After Aging)Tensile Strength: ≥14 MPa14.3-18.0 MPa29.2-38.6 MPa
    Ultimate Elongation: ≥400%417-524%453-504%
    ISO 10993-5CytotoxicityNon- In Vitro CytotoxicityDevice extract is cytotoxic. (Further evaluated by ISO 10993-11)Device extract is cytotoxic. (Further evaluated by ISO 10993-11)
    ISO 10993-11Acute Systemic ToxicityNon-acute systemic toxicityDid not show acute systemic toxicity in vivo / PassDid not show acute systemic toxicity in vivo / Pass
    ISO 10993-10IrritationNon-irritatingNot an irritant / PassNot an irritant / Pass
    ISO 10993-10SensitizationNon-sensitizingNot a sensitizer / PassNot a sensitizer / Pass
    ASTM D6978-05 (Chemotherapy Permeation)See detailed tables below for each drugBreakthrough Detection Time in MinutesSee tables in section 6.1See tables in section 6.2

    Chemotherapy Drug Permeation (Blue Gloves):

    Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes (Acceptance implicitly >0, ideally >240 for most drugs)Reported Device Performance (>240 is generally considered passing for clinical relevance)
    Carboplatin10.0 mg/ml> 240 Minutes> 240 Minutes / Pass
    Carmustine3.3 mg/ml> 240 Minutes (implicit desired)18.4 Minutes / Caution (Low permeation time)
    Cisplatin1.0 mg/ml> 240 Minutes> 240 Minutes / Pass
    Cyclophosphamide20.0 mg/ml> 240 Minutes> 240 Minutes / Pass
    Doxorubicin HCI2.0 mg/ml> 240 Minutes> 240 Minutes / Pass
    Etoposide20.0 mg/ml> 240 Minutes (implicit desired)202.7 Minutes / Caution (Low permeation time)
    Fluorouracil50.0 mg/ml> 240 Minutes> 240 Minutes / Pass
    Paclitaxel6.0 mg/ml> 240 Minutes> 240 Minutes / Pass
    ThioTepa10.0 mg/ml> 240 Minutes (implicit desired)26.6 Minutes / Caution (Low permeation time)

    Chemotherapy Drug Permeation (Purple Gloves):

    Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes (Acceptance implicitly >0, ideally >240 for most drugs)Reported Device Performance (>240 is generally considered passing for clinical relevance)
    Carboplatin10.0 mg/ml> 240 Minutes> 240 Minutes / Pass
    Carmustine3.3 mg/ml> 240 Minutes (implicit desired)20.9 Minutes / Caution (Low permeation time)
    Cisplatin1.0 mg/ml> 240 Minutes> 240 Minutes / Pass
    Cyclophosphamide20.0 mg/ml> 240 Minutes> 240 Minutes / Pass
    Doxorubicin HCI2.0 mg/ml> 240 Minutes> 240 Minutes / Pass
    Etoposide20.0 mg/ml> 240 Minutes> 240 Minutes / Pass
    Fluorouracil50.0 mg/ml> 240 Minutes> 240 Minutes / Pass
    Paclitaxel6.0 mg/ml> 240 Minutes> 240 Minutes / Pass
    ThioTepa10.0 mg/ml> 240 Minutes (implicit desired)33.5 Minutes / Caution (Low permeation time)

    2. Sample size used for the test set and the data provenance

    The document provides details of the tests conducted but does not explicitly state the exact sample sizes for each test (e.g., number of gloves tested for physical dimensions, biocompatibility, or chemotherapy permeation). It mentions quantities like "0/125/Pass" for the watertightness test, implying 125 units were tested.

    The data provenance is not specified regarding the country of origin. The study appears to be bench testing (non-clinical), meaning it was performed in a lab setting rather than with human subjects, making it akin to a controlled experimental study rather than retrospective or prospective clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable as the document describes bench testing for physical properties, chemical permeation, and biocompatibility of gloves. These tests rely on established scientific methods and instrumental measurements, not on expert human interpretation or consensus to establish ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as the study did not involve human interpretation or subjective assessments that would require adjudication. The performance is determined by objective measurements against predefined criteria in the standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is an examination glove, not an AI-assisted diagnostic or therapeutic tool. Therefore, MRMC studies and AI-related effectiveness metrics are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not applicable. The device is a physical product (gloves) and does not involve any algorithms or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The 'ground truth' for these tests is defined by industry standards and their specified acceptance criteria. For example:

    • Physical Properties: ASTM D6319, ASTM D412 specify acceptable ranges for length, width, thickness, tensile strength, and elongation.
    • Freedom from Holes: ASTM D5151 sets the Acceptable Quality Level (AQL) for watertightness.
    • Powder Content: ASTM D6124 defines the maximum allowable powder.
    • Biocompatibility: ISO 10993 standards define methods and acceptable responses for cytotoxicity, irritation, sensitization, and systemic toxicity.
    • Chemotherapy Permeation: ASTM D6978-05 provides the standard practice for assessing breakthrough times. The "ground truth" here is the measured time until breakthrough, compared against a desired clinical threshold (e.g., >240 minutes for safe handling).

    8. The sample size for the training set

    This is not applicable. There is no "training set" as the device is a physical product and the study involves direct measurements and testing against specified standards, not machine learning model training.

    9. How the ground truth for the training set was established

    This is not applicable for the same reason as point 8.