Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K213992

    Validate with FDA (Live)

    Date Cleared
    2023-05-31

    (526 days)

    Product Code
    Regulation Number
    876.5820
    Age Range
    0 - 65
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blood Tubing Set is a sterile, single use, disposable indicated for use with a prescribed hemodialyzer. For use with a compatible hemodialyzer, as per the labeling.

    The Blood Tubing Set is intended for acute and chronic hemodialysis therapy.

    The Blood Tubing Set is intended to be used with Fresenius Medical Care 2008® Series K, K2 and T Hemodialysis Machines.

    Device Description

    The CombiSet Bloodline is a single-use, ethylene oxide (EO) sterilized blood tubing set. The CombiSet Bloodline is part of the extracorporeal hemodialysis circuit. During hemodialysis, the extracorporeal circuit transports arterial blood from the patient's arterial access (e.g., fistula or catheter), through a hemodialyzer, and back to the patient's venous access.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the CombiSet Hemodialysis Blood Tubing Set. It does NOT describe a study involving an AI device or algorithm. Therefore, I cannot provide the requested information about acceptance criteria, study details, sample sizes, expert involvement, or comparative effectiveness with AI.

    The document focuses on the substantial equivalence of the CombiSet Hemodialysis Blood Tubing Set to a predicate device (CombiSet SMARTECH Hemodialysis Blood Tubing Sets, K201207) through various performance, sterilization, and biocompatibility tests.

    Here's the relevant information about the non-AI device's performance testing:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test ConductedTest Objective (Acceptance Criteria Implicit)Reported Device Performance
    Structural IntegrityDemonstrate that the bloodline can withstand 1.5X the labeled maximum positive and negative pressuresMet criteria
    Pump Segment PerformanceEvaluate performance characteristics of the bloodline over the range of the inlet pressures (normally 0 mmHg to -250 mmHg). Flow rate settings will cover up to 600 mL/min.Met criteria
    Visual Inspection for Endurance and Simulated Use TestsEndurance: Demonstrate that bloodlines perform with no tubing failures (kinking, collapsing, or disconnection) at the maximum labeled flow rate and pressures for not less than 18 hr. Simulated Use: Demonstrate that bloodlines perform with no tubing failures (kinking, collapsing, or disconnection) under simulated use conditions for not less than 4 hr.Met criteria
    Needle Access Port TestDemonstrate that the needle access ports can withstand 1.5X the labeled pressures after being punctured with the largest gauge needle recommended in the labeling (21 gauge) - 6X for maximum positive pressure, 11X for maximum negative pressure per ISO 8638:2010.Met criteria
    DIN ConnectorsDemonstrate that the DIN connectors do not leak when subjected to fluid pressure of 300-330 kPa.Met criteria
    Male and Female Luer Connectors TestsDemonstrate that the applicable components of the bloodline meet the dimensional and performance requirements of ISO 80369-7 2016 (Sections 5 and 6).Met criteria
    Visual Inspection for Packaging (Shipping study)Demonstrate that the shipping case, packaging configuration, and palletization pattern maintain the product's structural integrity during manual handling and motorized freight.Met criteria
    Tensile TestingDemonstrate that all bonded engagements in the bloodline between components, and between components and tubing can withstand a tensile force of 15 lbf.Met criteria
    Spike Flow RateDemonstrate that the spike, a component of the bloodline, can deliver not less than 1000 mL of a sodium chloride solution in 10 min under a static head of 1 m.Met criteria
    Spike Insertion ForceDemonstrate that the spike, a component of the bloodline, is capable of piercing and penetrating the closure of an infusion device without coring and with force not exceeding 200 N, when inserted at a rate of 500 mm/min.Met criteria
    Spike Leak TestDemonstrate that the spike, a component of the bloodline, will not leak after piercing an infusion device, remaining pierced for 5 hr, and then having an applied internal pressure of 20 kPa for 15 sec.Met criteria
    Spike Disconnection ForceDemonstrate that the spike, a component of the bloodline, is capable of being removed from the insertion point when a removal force is applied at a rate of 100 mm/min. Establish the removal force value.Met criteria
    Labeling Content per FDA Guidance (Blood Tubing Set) and ISO 8638Verify the Instructions for Use, color-coded components, unit labels, shipping carton graphics, and case labels for the bloodline meet the requirements of ISO 8638:2010 and FDA Guidance.Met criteria
    Readability of Barcode with Human Readable Identification CodesDemonstrate that the barcode information on the outer container labels and unit labels for the bloodlines is capable of being scanned.Met criteria
    Level Detector TestDemonstrate that the venous chamber of the bloodline interfaces correctly with the hemodialysis machine (2008 series) such that the venous clamp will activate (close) when the fluid level inside the venous chamber falls below the sensor heads.Met criteria
    Air-Capture Chamber Fill LevelDemonstrate that the recommended fill level of the air-capture chambers is marked.Met criteria
    Blood Filter Retention TestDemonstrate that the blood filter of the bloodline remains in the assembly position after tests at 1.5X the maximum recommended positive pressure, with a flow rate of 600 mL/min for not less than 18 hr.Met criteria
    Transparency of Transducer ProtectorsDemonstrate that the machine side of the TPs is clear to allow for visual inspection of blood contamination during use.Met criteria
    Transducer Protector Leak TestDemonstrate that the TP is capable of maintaining a secure and leak-free connection to the hemodialysis machine.Met criteria
    Structural Integrity (Vented Caps)Demonstrate that the vented caps can withstand 1.5X the labeled maximum positive and negative pressures.Met criteria
    Viral Retentiveness Test for Transducer ProtectorsDemonstrate that the membrane inside the transducer protector prevents the passage of bacteriophage (Φ X174) from the patient side to the machine side up to a pressure of 750 mmHg for 1 hr.Met criteria
    Tubing Compliance TestDemonstrate that tubing is capable of being occlusively clamped by the venous line clamp of the dialysis machine.Met criteria
    Clamp Occlusion TestDemonstrate that tubing is capable of being occlusively clamped by the bloodline clamps.Met criteria

    Note: The document explicitly states: "All testing met predetermined acceptance criteria."

    2. Sample size used for the test set and the data provenance:

    The document does not provide specific sample sizes (e.g., number of units tested) for each performance or biocompatibility test. It mentions that "samples" were tested.

    • Data Provenance: The studies were conducted by Fresenius Medical Care Renal Therapies Group, LLC, likely internally or through contracted labs, as part of their 510(k) submission to the FDA. The country of origin for the data is not specified beyond being part of a US FDA submission. It is implicitly prospective testing of the manufactured device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable as the document describes performance testing of a physical medical device (blood tubing set), not an AI device requiring expert ground truth for classification or diagnosis.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable as the document describes performance testing of a physical medical device. The results of the physical tests (e.g., pressure resistance, leak tests) are objective measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This document is for a physical medical device (blood tubing set), not an AI device. No human reader studies or AI assistance were involved.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This document is for a physical medical device, not an AI device or algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the device's performance is based on established engineering standards and regulatory requirements (e.g., ISO 8638:2010, ISO 80369-7:2016, AAMI/ANSI/ISO 10993-7:2008, ANSI/AAMI/ST72:2019, ISO 10993-1:2018, ASTM F1671-13). The device's ability to resist pressure, prevent leaks, sustain flow rates, and demonstrate biocompatibility according to these standards serves as the ground truth.

    8. The sample size for the training set:

    Not applicable. This document describes the testing of a physical medical device, not an AI device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1