    K Number
    K213441

    Validate with FDA (Live)

    Date Cleared
    2022-03-09

    (135 days)

    Product Code
    Regulation Number
    888.3080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PathLoc Lumbar Interbody Fusion Cage System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). This device is to be used with autogenous bone graft and/or allogenous bone graft composed of cancellous and/or corticocancellous bone. PathLoc Lumbar Interbody Fusion Cage System is to be used with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.

    Device Description

    The PathLoc Lumbar Interbody Fusion Cage System implants are interbody fusion devices intended for use as an aid in spinal fixation. They are made of Titanium 6AL-4V Alloy (ASTM F136). These hollow, rectangular implants are offered in a variety of widths, lengths, heights and lordotic angles designed to adapt to a variety of patient anatomies. The implants can be expanded in height after insertion in the unexpanded state using the system instrumentation. The implants have serrations on the superior and inferior surfaces designed for fixation.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the "PathLoc Lumbar Interbody Fusion Cage System." It outlines the regulatory approval process and describes the device, its intended use, and the studies conducted to demonstrate its substantial equivalence to previously approved devices.

    Here's an analysis of the provided information concerning acceptance criteria and study proof:

    Important Note: The provided document is for a lumbar interbody fusion cage system, which is a mechanical implant. The questions posed (e.g., "effect size of how much human readers improve with AI vs without AI assistance," "standalone (i.e. algorithm only without human-in-the-loop performance) was done," "ground truth used (expert concensus, pathology, outcomes data, etc.)" with specific numbers of experts and adjudication methods) are typical questions for the validation of AI/ML-based medical devices for image analysis or diagnostics. These types of studies are not applicable to a mechanical orthopedic implant like the PathLoc Lumbar Interbody Fusion Cage System.

    The "studies" described in this document are bench testing to assess the mechanical properties and biocompatibility of the implant, demonstrating its structural integrity and safety for implantation. There are no "human readers," "AI assistance," or "ground truth based on expert consensus" in the context of this device's validation.

    Therefore, many of the requested points in your prompt cannot be answered from the provided text because they pertain to a different type of medical device (e.g., AI diagnostic software) and not a physical implant.

    Let's address what is available in the document:


    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of acceptance criteria with numerical targets and the exact reported performance values. Instead, it states:

    Acceptance Criteria (General Description)Reported Device Performance (General Statement)
    "The PathLoc Lumbar Interbody Fusion Cage System met all pre-defined acceptance criteria""Bench testing to evaluate the mechanical properties of the PathLoc Lumbar Interbody Fusion Cage System showed a higher or similar mechanical value than predicate marketed devices."
    Compliance with ASTM F2077, ASTM F2267 standards for static and dynamic mechanical properties.Testing was performed for: Static Axial Compression, Static Compression-Shear, Static Torsion, Static Subsidence, Dynamic Axial Compression, Dynamic Compression-Shear. The results supported substantial equivalence.
    Use of Ti-6Al-4V ELI titanium alloy (ASTM F136) for material composition."The device is made of Titanium 6AL-4V Alloy conforming ASTM F136 and is manufactured in an identical manner to the predicate device." "This this is the same material used in the predicate devices."

    The "acceptance criteria" are implied to be meeting or exceeding the performance of the legally marketed predicate devices and conforming to relevant ASTM standards. The detailed numerical values of these tests are not provided in this summary.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: This refers to the number of physical samples (implants) tested in the mechanical bench studies. The document does not specify the exact number of samples tested for each mechanical test. Standard ASTM test methods typically prescribe a minimum number of samples (e.g., 5 or 6 per test condition), but this detail is not present.
    • Data Provenance: Not applicable in the context of clinical patient data. This is bench testing of manufactured devices. The manufacturer is L&K BioMed Co., Ltd. from South Korea. The testing was reported as "non-clinical testing."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This is a mechanical device. "Ground truth" in this context is established by standardized mechanical testing methods (ASTM standards) and engineering principles, not by medical experts interpreting data like images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. See point 3. This device is not an AI diagnostic tool and does not involve human readers interpreting cases.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. See point 3. This is not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Type of "Ground Truth": For this mechanical device, the "ground truth" or standard for evaluation is:
      • ASTM Standard Methods: Compliance with the mechanical testing specifications outlined in ASTM F2077 and ASTM F2267.
      • Predicate Device Performance: Demonstrating mechanical performance (e.g., static and dynamic strength, subsidence) that is "higher or similar" to the legally marketed predicate devices.

    8. The sample size for the training set

    • Not Applicable. There is no "training set" in the context of a mechanical implant. This term applies to machine learning models.

    9. How the ground truth for the training set was established

    • Not Applicable. See point 8.

    In summary, the provided document details the regulatory clearance for a physical medical implant. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" for such a device primarily revolve around mechanical performance bench testing (to ensure it can withstand forces in the body without failure) and biocompatibility (to ensure it doesn't cause adverse reactions when in contact with human tissue). The questions you've posed are highly relevant for digital health devices, particularly those involving AI/ML for diagnostics or image analysis, but they do not align with the type of device described in this FDA 510(k) summary.