510(k) Data Aggregation

    K Number
    K213289

    Validate with FDA (Live)

    Date Cleared
    2023-05-26

    (602 days)

    Product Code
    Regulation Number
    878.4460
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A powder-free surgeons' glove is intended to be worn by operating room personnel to protect a surgical wound from contamination. The glove system was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. Chemotherapy testing was carried out on the system as a whole, including both the under-glove together. Please note that the following drugs have extremely low permeation times: 3.0 minutes and Thiotepa: 67.0 minutes. Warning: Do not use with Carmustine or Thiotepa.

    Device Description

    Gammex® Pl Plus Glove-in-Glove™ System tested for use with Chemotherapy Drugs is a sterile and disposable device. This glove system ismade of synthetic polyisoprene rubber. Gammex® Pl Plus Glove- in-Glove™ System Tested for Use with Chemotherapy Drugs is comprised of a single-use, sterile, white outer glove, which is a disposable, powder-free surgical glove made from synthetic polyisoprene and a single-use, sterile, green underglove which is a disposable, powder-free surgical glove made from syntheticpolyisoprene. The Gammex® Pl PlusGlove-in-Glove™ system Testedfor Use with Chemotherapy Drugs is comprised of the underglove being mechanically inserted into the outer glove prior to packing and sterilization. There is no adhesive present between the two gloves. This results in quick double gloving with only one donning event.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA regarding the "Gammex® PI Plus Glove-in-Glove™ System Tested For Use with Chemotherapy Drugs." This document outlines the device's characteristics, its intended use, and the studies conducted to demonstrate its safety and effectiveness.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document includes a table of non-clinical tests performed on the Gammex® PI Plus Glove-in-Glove™ System.

    Test TitlePurpose of TestAcceptance CriteriaReported Device Performance
    ASTM D3767-03DimensionsIn accordance with ASTM D3577-19: Standard Specification for Rubber Surgical Gloves.PASS
    ASTM D3577-19Physical PropertiesFor tensile strength, ultimate elongation and stress at 500% elongation before and after accelerated aging for synthetic surgical gloves per ASTM D3577-19 : Standard Specification for Rubber Surgical Gloves.PASS
    ASTM D5151-19Freedom from holesIn accordance with ASTM D3577-19: Standard Specification for Rubber Surgical Gloves with AQL requirements of 1.5.PASS
    ASTM D6124Powder-FreeApplicable acceptance criteria for powder free ≤ 2mg per glove per ASTM D3577-19: Standard Specification for Rubber Surgical Gloves.PASS
    ANSI/AAMI/ISO 11137-1:2006SterilityRequirement of 10-6 SAL per ISO 11137-1: Sterilization for health care products – Radiation – Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices.PASS
    ASTM D6978-05(2019)Chemotherapy Drug Permeation TestIn accordance with ASTM D6978-05(2019): Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. Note: Carmustine (35.0 min) and Thiotepa (67.0 min) have extremely low permeation times, implying these are specific exceptions within the criteria or require special consideration/warnings.PASS
    ISO 10993-5 (Outer and Underglove)ISO in vitro Cytotoxicity Study (Undiluted, 1:2, 1:4, 1:8, 1:16, 1:32, 1:64 dilutions)Acceptance criteria in accordance with ISO 10993-5: Biological Evaluation of Medical Devices, Part 5: Tests for In Vitro Cytotoxicity, 2009. (Specifically, for the outer glove: non-cytotoxic at 1:4, 1:8, 1:16, 1:32, and 1:64 dilutions. For the underglove: non-cytotoxic at 1:16, 1:32, and 1:64 dilutions). Intermediate dilutions for both gloves showed varying levels of cytotoxicity.FAIL (at certain dilutions for both gloves, e.g., undiluted and 1:2 for outer; undiluted, 1:2, 1:4, 1:8 for under)
    ISO 10993-10:2010ISO Skin Irritation StudyPasses Primary Skin Irritation test per ISO 10993-10, Biological Evaluation of medical devices, Part 10: Test for irritation and skin sensitization. Not an irritant under study conditions.PASS
    ISO 10993-10:2010ISO Maximization Sensitization StudyPasses Dermal Sensitization test per ISO 10993-10, Biological Evaluation of medical devices, Part 10: Test for irritation and skin sensitization. Not a sensitizer under study conditions.PASS
    ISO 10993-11:2017ISO Acute Systemic Toxicity StudyPasses Acute Systemic Toxicity Test per ISO 10993-11, Biological Evaluation of medical devices, Part 11: Test for systemic toxicity. No mortality or evidence of acute systemic toxicity under study conditions.PASS

    Chemotherapy Drug Permeation Times (Detailed)

    Test Chemotherapy Drug & ConcentrationAverage Breakthrough Detection Time (Minutes)
    Bleomycin - 15.0 mg/ml>240
    Busulfan - 6.0 mg/ml>240
    Carboplatin - 10 mg/ml>240
    Carmustine - 3.3 mg/ml35.0
    Cisplatin - 1.0 mg/ml>240
    Cyclophosphamide - 20.0 mg/ml>240
    Cytarabine HCl - 100.0 mg/ml>240
    Dacarbazine - 10.0 mg/ml>240
    Daunorubicin HCl - 5.0 mg/ml>240
    Docetaxel - 10.0 mg/ml>240
    Doxorubicin HCl 2.0 mg/ml>240
    Epirubicin - 2.0 mg/ml>240
    Etoposide - 20.0 mg/ml>240
    Fludarabine - 25 mg/ml>240
    Fluorouracil - 50.0 mg/ml>240
    Gemcitabine - 38.0 mg/ml>240
    Idarubicin - 1.0 mg/ml>240
    Ifosfamide - 50.0 mg/ml>240
    Irinotecan - 20.0 mg/ml>240
    Mechlorethamine HCl - 1.0 mg/ml>240
    Melphalan - 5.0 mg/ml>240
    Methotrexate - 25.0 mg/ml>240
    Mitomycin C - 0.5 mg/ml>240
    Mitoxantrone - 2.0 mg/ml>240
    Oxaliplatin - 2.0 mg/ml>240
    Paclitaxel - 6.0 mg/ml>240
    Rituximab - 10.0 mg/ml>240
    Thiotepa - 10.0 mg/ml67.0
    Vincristine Sulfate - 1.0 mg/ml>240

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify sample sizes for individual tests like dimensions, physical properties, freedom from holes, powder-free, sterility, or biocompatibility studies. The general phrasing "Meets ASTM D..." or "Passes..." suggests that the sample sizes typically required by these standards were used.

    For the Chemotherapy Drug Permeation Test (ASTM D6978-05), while the results for each drug are provided (Average Breakthrough Detection Time), the document does not specify the number of samples (e.g., gloves or glove specimens) tested for each drug.

    The data provenance is prospective in the sense that these are tests conducted specifically for this 510(k) submission to demonstrate the device's performance against established standards. There is no mention of country of origin for the data; it is assumed to be from a testing facility compliant with the mentioned standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The "ground truth" for these tests is defined by the objective measurement methods and acceptance criteria specified in the referenced ASTM, ANSI, and ISO standards. These standards themselves are developed by expert consensus, but specific independent expert adjudication for this particular test set is not mentioned or typically required for this type of device and testing.

    4. Adjudication Method for the Test Set:

    An explicit adjudication method (like 2+1 or 3+1) is not applicable or mentioned for the non-clinical tests described. The tests are objective measurements against predefined criteria outlined in the consensus standards. The "Pass/Fail" results are determined directly by comparing the test outcomes to these established thresholds.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, a MRMC comparative effectiveness study was not done. The document explicitly states: "A clinical study was not conducted on the subject device." This type of study focuses on human reader performance, often with or without AI assistance, which is not relevant to the physical and chemical resistance properties of a surgical glove.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done:

    No, a standalone algorithm-only performance study was not done. This device is a physical medical device (surgical glove), not an algorithm or AI software. Therefore, studies evaluating algorithm performance are not relevant.

    7. The Type of Ground Truth Used:

    The ground truth used for the non-clinical tests is based on consensus standards and laboratory measurements.

    • For physical properties, dimensions, freedom from holes, powder residue, and sterility: The ground truth is objective measurement against the specified ASTM and ISO standards' criteria.
    • For chemotherapy drug permeation: The ground truth is the measured breakthrough time determined by the ASTM D6978-05 standard.
    • For biocompatibility (skin irritation, sensitization, acute systemic toxicity, cytotoxicity): The ground truth is the biological response observed in validated in vitro and in vivo models according to the ISO 10993 series of standards.

    8. The Sample Size for the Training Set:

    Not applicable. This device is a physical product (surgical glove) and does not involve AI or machine learning algorithms that would require a "training set."

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this device.