510(k) Data Aggregation

    K Number
    K213239

    Validate with FDA (Live)

    Date Cleared
    2022-06-02

    (245 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Single use injection needle is to be used in conjunction with an endoscopic injections, such as for the treatment of esophageal and gastric varies and for submucosal dye marking in the GI tract.

    Device Description

    Endoscopic Injection Needle consists of Handle, Rod, Needle, Needle holder, Sheath, Catheter, Adapter, Sheath holder, Sheath protection pipe, Tube cap.

    The Endoscopic Injection Needle is available in three needle sizes (20 gauges 22 gauge and 25 gauge), five needle lengths (4mm,5mm,8mm and 10mm),five working length(1200mm,1600mm, 1950mm,2000mm, 2300mm). There are 5 series (SN18-05 series, SN18-07 series, SN18-09 series, SN19-05 series and SN19-07 series) total ninety (90) specification injection needles.

    The Endoscopic Injection Needle is sterile, single-use device. It is expected to be used in conjunction with an endoscope to perform endoscopic injections into tissue in the GI tract during an endoscopic procedure.

    The Endoscopic Injection Needle is designed to pass through the endoscope's forceps to mark the lesions of the digestive tract and use it for injection. Compatibility with endoscopes, working lengths are 1600mm, 1950mm,2000mm, 2300mm, and the minimum working clamp channel is φ 2.8 mm.

    AI/ML Overview

    The provided text describes the Endoscopic Injection Needle (K213239) and its acceptance criteria, along with a summary of the testing performed to demonstrate its safety and effectiveness.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the successful completion of the tests listed. The "Reported Device Performance" column indicates that the device met these criteria through successful testing.

    Test ItemAcceptance Criteria (Implied)Reported Device Performance
    Integrity Checks
    AppearanceIntegrity, cleanliness, and hygieneConfirmed to be integrity and cleanliness and hygiene
    DimensionComplied with company's requirementsComplied with company's requirements
    LeakageNo gas leakageNo gas leakage
    Needle ProtectionNeedle shall not be exposed when pulling rod backNeedle not exposed when pulling rod back
    Material/Mechanical Properties
    Resistance to CorrosionNo corrosive marks in immersed part of the needleNo corrosive marks found
    StiffnessNeedle tube is not easy to deformConfirmed not easy to deform
    Resistance to BreakageNeedle tube is not broken during useConfirmed not broken during use
    Penetrating ForceNeedle tip is sharp (implies sufficient sharpness)Confirmed needle tip is sharp
    Connection StrengthAll joints are rigid connectConfirmed all joints are rigid connect
    Luer Lock ConnectorsMeets the requirement of ISO 80369-7Meets ISO 80369-7
    Functional Testing
    Basic Functional TestingSuccessfully inserted and withdrawn into/out of endoscope; stylet passes syringe successfullySuccessfully inserted and withdrawn, stylet passes through
    Chemical/Sterilization Properties
    Chemical PropertiespH, Total content of heavy metals, Potassium permanganate reductive substance, Evaporative residues and Ultraviolet Absorbance meet related requirementsMet related requirements
    SterilityDevice shall be sterile (SAL: 10-6)Sterile (by ethylene oxide, SAL: 10-6)
    EO Residue≤10 μg/g≤10 μg/g
    PyrogenNo pyrogenNo pyrogen
    BiocompatibilityConform with ISO10993-1, -4, -5, -10, -11 (Hemolysis, cytotoxicity, sensitization, intracutaneously irritation, acute systemic toxicity, material medicated pyrogenicity)Verified with ISO 10993 (non-toxic and biocompatible)
    Shelf LifeMaintain sterility and performance for 2 years2-year shelf life established

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not explicitly state the sample size used for each specific bench test item. It generally refers to "bench testing" being performed.
    • Data Provenance: The data is from "Summary of non-clinical and performance testing- Bench testing" carried out by Beijing ZKSK Technology Co., Ltd. The tests were performed according to recognized consensus standards and the company's technical requirements. This is retrospective testing performed by the manufacturer on its own device prototypes/production samples. The country of origin for the testing is implied to be China, where the manufacturer is located.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This device is a physical medical instrument, not an AI/software device that relies on expert interpretation for ground truth in its performance evaluation as described in the context of AI/ML devices. The "ground truth" for the bench tests is based on objective measurements and adherence to engineering and biological standards.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, this is not an AI/ML device requiring human adjudication for performance evaluation. Adjudication methods are typically relevant for subjective expert assessments or when there are disagreements in interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The device is an Endoscopic Injection Needle, a physical instrument, and its evaluation focuses on bench testing and mechanical/biocompatibility performance, not on AI-assisted human reading.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    Not applicable. This is not an algorithm or AI device; it's a physical medical instrument. Therefore, the concept of "standalone performance" for an algorithm doesn't apply.

    7. The Type of Ground Truth Used

    The ground truth for the performance testing is based on objective measurements, adherence to recognized consensus standards (e.g., ISO 7864, ISO 9626, ISO 10993 series, ISO 11607-1, ISO 80369-7), and the manufacturer's internal design requirements and specifications. For biocompatibility, the ground truth is established by the biological responses observed against the criteria in ISO 10993.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set. The device's safety and effectiveness are established through physical and biological testing as described above.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.