Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K213156

    Validate with FDA (Live)

    Date Cleared
    2021-12-08

    (79 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Becton Dickinson Pen Needle is intended for use with pen injector devices for the subcutaneous injection of drugs.

    Device Description

    The BD Pen Needle is designed for use with pen injectors for subcutaneous injection of a desired dose of drugs approved for delivery using a pen needle. It consists of a needle, base, and shield assembly. The BD Nano™ 2nd Gen Pen Needle is offered in a 34-gauge size and 4mm length XTW cannula. The device is non-toxic and non-pyrogenic. It is a single-use disposable device that is provided sterile.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the BD Nano™ 2nd Gen Pen Needle, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test/CharacteristicAcceptance Criteria (Standard/Requirements)Reported Device Performance
    Bench Functional Performance
    Cannula StraightnessISO 11608-2:2012, ISO 9626:2016, BD Internal SpecificationsMet requirements
    Cannula StiffnessISO 11608-2:2012, ISO 9626:2016, BD Internal SpecificationsMet requirements
    Cannula Breakage ResistanceISO 11608-2:2012, ISO 9626:2016, BD Internal SpecificationsMet requirements
    Flow Rate MeasurementISO 11608-2:2012, BD Internal SpecificationsMet requirements
    Cannula PatencyISO 11608-2:2012, BD Internal SpecificationsMet requirements
    Patient End Penetration ForceISO 11608-2:2012, BD Internal SpecificationsMet requirements
    Cannula Pull ForceISO 11608-2:2012, BD Internal SpecificationsMet requirements
    Dose AccuracyISO 11608-2:2012, BD Internal SpecificationsMet requirements
    Material/Design Characteristics
    BiocompatibilityNon-toxicDevice is non-toxic
    PyrogenicityNon-pyrogenicDevice is non-pyrogenic
    Sterility (SAL)10^-6Achieved 10^-6 SAL
    Sterilization MethodNot explicitly stated as "acceptance criteria," but gamma irradiation is usedGamma irradiation (same as predicate)
    Usability/FunctionalityFor use with pen injector devices for subcutaneous injection of drugsMet requirements for intended use
    Shelf Life5 yearsSupported by testing (implied by "support the 5-year shelf life")

    2. Sample Size Used for the Test Set and Data Provenance

    • The document does not specify the exact sample sizes (number of devices tested) for the bench functional performance tests. It states "Testing included functional performance per ISO 11608-2:2012... and testing following BD's Internal Specifications." While standards typically mandate minimum sample sizes, the specific numbers for this device are not provided in this summary.
    • Data Provenance: The data is from "Non-Clinical Testing" performed by Becton, Dickinson and Company as part of its design control process. This indicates the testing was conducted internally by the manufacturer. It is prospective testing designed to validate the new device. The country of origin of the data is not explicitly stated.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • This information is not applicable as the study described is a non-clinical, bench performance study of a medical device (a pen needle), not a study involving human diagnosis, interpretation of medical images, or clinical outcomes that would require expert-established ground truth. The "ground truth" here is defined by meeting physical and performance specifications outlined in international standards (ISO) and the manufacturer's internal specifications.

    4. Adjudication Method for the Test Set

    • This information is not applicable for the same reasons as #3. Adjudication methods like 2+1 or 3+1 are used to reconcile discrepancies among human readers or evaluators, which is not relevant to bench testing of device performance against pre-defined specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a multi-reader, multi-case comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic aids or clinical decision support systems, where the goal is to assess the impact of AI on human reader performance. The BD Nano™ 2nd Gen Pen Needle is a physical medical device (pen needle), not an AI system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a physical pen needle, not an algorithm or AI system. Its performance is evaluated through bench testing against physical and functional standards, not as a standalone algorithm.

    7. The Type of Ground Truth Used

    • The "ground truth" for this device's performance testing is based on pre-defined technical specifications, international standards (ISO 11608-2:2012, ISO 9626:2016), and the manufacturer's (BD's) internal specifications. For example, a flow rate measurement would be compared against a specified range or minimum value defined by the standard or the manufacturer, making that standard the "ground truth."

    8. The Sample Size for the Training Set

    • Not applicable. Training sets are used in machine learning for AI algorithms. This device is a physical medical device, and its performance is evaluated through non-clinical testing, not by training an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable for the same reasons as #8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1