510(k) Data Aggregation

    K Number
    K213124

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-10-26

    (29 days)

    Product Code
    Regulation Number
    888.3660
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tornier Perform Reversed Glenospheres are indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by:

    • Rheumatoid arthritis
    • · Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
    • · Correction of functional deformity
    • · Fractures of the humeral head
    • · Traumatic arthritis
    • · Revision of the devices if sufficient bone stock remains

    Notes:

    • · All components are single use.
    • The glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation.
    Device Description

    The Tornier Perform™ Reversed Glenoid Cannulated Glenospheres are part of a reverse shoulder prosthesis consisting of cannulated cobalt chromium and titanium alloy glenospheres. The Tornier Perform Reversed Glenoid Cannulated Glenospheres are intended for replacement of the shoulder joint to reduce pain and improve shoulder mobility in comparison with preoperative status.

    The Tornier Perform Reversed Glenoid Cannulated Glenospheres must be used in association with a compatible Tornier reversed glenoid and humeral component. The cannulated glenosphere can be used as part of the glenoid implant. The glenoid implant is composed of a baseplate with central and peripheral anchoring screws and a glenosphere. An optional press-fit post is available that can be used in lieu of the central anchoring screw.

    Ancillary instruments are also provided for the implantation of the prosthesis.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the Tornier Perform™ Reversed Glenoid Cannulated Glenospheres. This document does not contain information about acceptance criteria or a study proving the device meets said criteria in the context of an AI/ML powered device.

    The document states:

    • "Non-clinical bench testing and process validations were performed to demonstrate substantial equivalence to the predicate device."
    • "No clinical studies were performed."

    Therefore, there is no information available in the provided text to answer the specific questions about acceptance criteria for an AI/ML device, performance studies, sample sizes, expert involvement, or ground truth establishment. The document describes mechanical bench testing (Glenosphere Loosening, Taper Disassembly, Glenosphere Fatigue) and simulated use to demonstrate substantial equivalence to a legally marketed predicate device, K161742.