510(k) Data Aggregation

    K Number
    K212897

    Validate with FDA (Live)

    Date Cleared
    2021-11-04

    (55 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IPL Hair Removal Device (Model: AI01) is intended for the removal of unwanted body hair.

    Device Description

    IPL Hair Removal Device, model: Al01 is a device for the permanent reduction of hair growth based on Intense Pulsed Light (IPL). It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Xenon Arc Flashlamp and a skin sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot emit the treatment light pulses. The device incorporates Intense Pulse Light (IPL) technology. The purpose of the light is to heat the root where the hair grows. Device includes a main unit, an adaptor, and accessories, accessories include goggle.

    AI/ML Overview

    The provided text is a 510(k) summary for the IPL Hair Removal Device, model: AI01. It describes acceptance criteria for safety and effectiveness based on non-clinical testing and comparison to a predicate device, as well as the study that demonstrates the device meets these criteria. No clinical study was performed.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Since this is a 510(k) submission and not a clinical study report, the acceptance criteria are primarily related to meeting recognized consensus standards for safety and biocompatibility, and demonstrating substantial equivalence to a predicate device. The "reported device performance" refers to the successful completion of these tests and alignment with the predicate.

    Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
    Safety - ElectricalANSI AAMI ES60601-1Complied
    Safety - Electromagnetic CompatibilityIEC 60601-1-2Complied
    Safety - Home Healthcare EnvironmentIEC 60601-1-11Complied
    Safety - Performance of Non-Laser Light SourceISO 60601-2-57: 2011Complied
    Safety - PhotobiologicalIEC 62471: 2006Complied
    Biocompatibility - CytotoxicityISO 10993-5:2009Meets requirement ("positive results" for validation implies successful testing if "positive" refers to the test for meeting the standard)
    Biocompatibility - Irritation & SensitizationISO 10993-10:2010Meets requirement ("positive results" for validation implies successful testing if "positive" refers to the test for meeting the standard)
    Software Verification & ValidationIEC 62304:2006+A1:2015 and FDA Guidance for SoftwarePerformed in accordance with guidance
    Substantial Equivalence - Indications for UseSame as predicate (unwanted body hair removal)Same (with a note on OTC vs. Prescription, deemed not to raise new safety questions)
    Substantial Equivalence - TechnologyIntense Pulse Light (IPL)Same
    Substantial Equivalence - Technical SpecificationsWavelength, Spot Size, Max. Fluence, Pulse Duration, Output energy, Pulsing ControlSame as predicate

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    The document states: "There was no clinical testing performed." Therefore, there is no sample size for a clinical test set, nor is there data provenance from human subjects. The testing was non-clinical (device bench testing, biocompatibility, and software validation). The country of origin for the device manufacturer is China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    Not applicable, as no clinical testing was performed for hair removal effectiveness. The "ground truth" for the non-clinical tests is based on adherence to recognized international standards and comparison to the predicate device's established safety and performance characteristics.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable, as no clinical testing with human subjects or expert adjudication of results was performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This device is an IPL hair removal device, not an AI-assisted diagnostic or imaging device that would involve human readers or MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    Not applicable. The device is a physical IPL hair removal device, not a standalone algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
    The "ground truth" for this 510(k) submission is primarily:

    • Adherence to Recognized Standards: The device's components and overall system passed various safety, biocompatibility, and software validation tests according to established international standards (e.g., IEC 60601 series, ISO 10993 series, IEC 62304).
    • Substantial Equivalence to Predicate Device: The device's technical characteristics, operating principle, and indications for use are demonstrated to be substantially equivalent to a legally marketed predicate device (K183217). The "ground truth" here is the established safety and effectiveness profile of the predicate.

    8. The sample size for the training set
    Not applicable. No machine learning or AI algorithm requiring a training set for clinical performance was described in the context of hair removal effectiveness. The software referred to in the document is for device control and functionality, not for performing a diagnostic or predictive task.

    9. How the ground truth for the training set was established
    Not applicable, as there was no training set mentioned in the context of clinical performance.