510(k) Data Aggregation

    K Number
    K212827

    Validate with FDA (Live)

    Date Cleared
    2021-12-06

    (94 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JR Medic Blue Nitrile Examination Gloves Powder Free tested for use with Chemotherapy drugs is a disposable device intended for medical purpose that is worn on the examination between patient and examiner.

    Additionally, the gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.

    The tested chemotherapy drugs and their breakthrough detection times are as follows:

    Tested Chemotherapy Drug Name & ConcentrationMinimum Breakthrough Detection Time
    Carmustine (BCNU) (3.3 mg/ml)35.0 Minutes
    Carboplatin (10 mg/ml)>240 Minutes
    Cisplatin (1 mg/ml)>240 Minutes
    Cyclophosphamide (Cytoxan) (20 mg/ml)>240 Minutes
    Dacarbazine (10.0 mg/ml)>240 Minutes
    Doxorubicin HCl (2 mg/ml)>240 Minutes
    Etoposide (20 mg/ml)>240 Minutes
    Fluorouracil (50 mg/ml)>240 Minutes
    Ifosfamide (50 mg/ml)>240 Minutes
    Methotrexate (25 mg/ml)>240 Minutes
    Mitomycin C (0.5 mg/ml)>240 Minutes
    Mitoxantrone (2 mg/ml)>240 Minutes
    Paclitaxel (6 mg/ml)>240 Minutes
    Thiotepa (10 mg/ml)64.9 Minutes
    Vincristine Sulfate (1 mg/ml)>240 Minutes
    Device Description

    The subject device in 510(K) notification is JR Medic blue nitrile examination gloves powder free tested for use with Chemotherapy drugs.

    The subject device is a patient examination glove made from acrylonitrile-butadiene copolymer dispersion, blue color, powder free and non sterile (as per 21CFR 880.6250, class I).

    The subject device meets all the current specifications listed under the ASTM Specification D 6319 -2019, Standard Specification for Nitrile Examination Gloves for Medical Application. This device also complies with requirements for standard practice for assessment of resistance of medical gloves to permeation by chemotherapy drugs as per ASTM D6978- 05(2019)

    AI/ML Overview

    I am sorry, but the provided text does not describe a study involving an AI device or algorithm, and therefore all the requested information about acceptance criteria, study design, sample sizes, ground truth, expert involvement, and MRMC studies cannot be extracted.

    The document is a 510(k) premarket notification for "JR Medic Blue Nitrile Examination Powder Free Gloves tested for use with chemotherapy drugs." It details the non-clinical testing performed on these medical gloves to demonstrate their safety and effectiveness, specifically their resistance to permeation by chemotherapy drugs, in accordance with established ASTM (American Society for Testing and Materials) standards.

    The document focuses on the physical, chemical, and biocompatibility properties of the gloves, not on the performance of a software algorithm or AI.