510(k) Data Aggregation

    K Number
    K212597

    Validate with FDA (Live)

    Date Cleared
    2022-05-28

    (285 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Amazing+ Latex Examination Powder Free Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Amazing Latex Examination Powder Free Gloves are manufactured to meet all the current specifications listed under the ASTM Specification D3578-2019, Standard Specification for Rubber Examination Gloves. They are made from Natural Rubber Latex. These gloves are natural in color (No color is added) and are powder free.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device: "Amazing+ Latex Examination Powder Free Gloves." This document describes the device and demonstrates its substantial equivalence to a legally marketed predicate device.

    However, the request asks for information related to a study that proves a device meets acceptance criteria, specifically in the context of an AI-powered medical device (e.g., sample size for test set, data provenance, expert adjudication, MRMC study, standalone performance, ground truth establishment for training and test sets).

    The provided 510(k) Summary does not describe an AI medical device. It describes physical specifications and performance of latex examination gloves. Therefore, many of the requested fields are not applicable to this type of device.

    Despite this, I will extract and present the relevant information that is available in the document, and clearly state when information is not applicable to this physical device.


    Device: Amazing+ Latex Examination Powder Free Gloves
    K Number: K212597
    Predicate Device: JR MEDIC Blue Latex Examination Powder Free Gloves (K192329)


    1. A table of acceptance criteria and the reported device performance

    CharacteristicStandard/Test MethodAcceptance CriteriaReported Device PerformanceComparison to Predicate
    Dimensions - LengthASTM D3578-2019Min 230 mm for all sizesX-Small 305 mm, Small 306 mm, Medium 307 mm, Large 308 mm, X-Large 310 mmSimilar
    Dimensions - WidthASTM D3578-2019X-Small 70+/-10 mm, Small 80+/-10mm, Medium 95+/-10 mm, Large 111+/-10 mm, X-Large 115+/-10 mmX-Small 76 mm, Small 85 mm, Medium 96 mm, Large 106 mm, X-Large 116 mmSimilar
    Dimensions - ThicknessASTM D3578-2019Palm > 0.08 mm min, Finger > 0.08 mm min for all sizesPalm: 0.16mm (all sizes), Finger: 0.22mm (all sizes)Similar
    Physical Properties - Tensile StrengthASTM D3578-2019Before Ageing: > 18 Mpa Min for all sizes. After Ageing: > 14 Mpa Min for all sizes.Before Ageing: X-Small 22.07 Mpa, Small 22.15 Mpa, Medium 22.22 Mpa, Large 22.30 Mpa, X-Large 22.32 Mpa. After Ageing: X-Small 18.49 Mpa, Small 18.56 Mpa, Medium 18.67 Mpa, Large 18.74 Mpa, X-Large 18.76 Mpa.Similar
    Physical Properties - Ultimate ElongationASTM D3578-2019Before Ageing: > 650% Min for all sizes. After Ageing: > 500% Min for all sizes.Before Ageing: X-Small 858%, Small 869%, Medium 874%, Large 880%, X-Large 882%. After Ageing: X-Small 841%, Small 848%, Medium 854%, Large 860%, X-Large 862%.Similar
    Physical Properties - Stress at 500% ElongationASTM D3578-2019Before Ageing: 5.5 Mpa Max for all sizes.Before Ageing: X-Small 5.1 Mpa, Small 5.1 Mpa, Medium 5.2 Mpa, Large 5.2 Mpa, X-Large 5.2 Mpa.N/A
    Water Tightness (Holes)ASTM D5151-2019AQL 2.5Gloves Pass AQL 1.5Same
    Powder Free ResidueASTM D6214<2 mg/glove MaxX-Small 0.21 mg/glove, Small 0.21 mg/glove, Medium 0.22 mg/glove, Large 0.22 mg/glove, X-Large 0.22 mg/gloveSimilar
    Biocompatibility - Primary Skin IrritationISO 10993-10:2010(E)Under the condition of study, not an irritantUnder the condition of study, not an irritantSame
    Biocompatibility - Dermal SensitizationISO 10993-10:2010(E)Under the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizerSame
    Biocompatibility - In vitro CytotoxicityISO 10993-5:2009(E)Under the conditions of the study, cytotoxicUnder the conditions of the study, cytotoxic (Note: This might seem counter-intuitive, but for certain test types, "cytotoxic" as a test result may simply indicate a reaction within the test conditions, not necessarily a failure from a biocompatibility standpoint for its intended use, especially if it's consistent with a cleared predicate. The key is that it's "Same" as the predicate's finding.)Same
    Biocompatibility - Material mediated PyrogenicityISO 10993-11:2017(E) / USP 41<151>Under the conditions of the study non pyrogenicUnder the conditions of the study, non-pyrogenicSame
    Biocompatibility - Acute Systemic Toxicity TestISO 10993-11:2017(E)Under the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concernSame
    Extractable ProteinASTM D 5712-95 (Re approved 2010) Standard Test Method for the Analysis of Protein in Natural Rubber200 µg/ dm² Max for all sizes43.65 µg/dm² (all sizes)N/A (Not compared to predicate in table, but meets criteria)

    Information Not Applicable to this Device (as it's a physical product, not an AI/software device):

    Given that the device is "Amazing+ Latex Examination Powder Free Gloves," the following sections of your request are not applicable as they pertain to AI/software functionality and clinical studies that are not performed for basic Class I examination gloves:

    • 2. Sample size used for the test set and the data provenance: This device is a physical product. Testing involves laboratory measurements on physical samples (e.g., gloves from a production lot) to assess properties like dimensions, strength, and integrity, not dataset sizes. The provenance is from "Amazing Rubber Products Pvt. Ltd." in Cochin, Kerala, India. The tests are non-clinical, likely performed according to ISO/ASTM standards.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for physical properties is established by standardized measurement methods (e.g., ASTM standards), not expert opinion or consensus.
    • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. Physical measurements do not require adjudication.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI device.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. Ground truth for device performance is based on the specified ASTM and ISO standards for physical and chemical properties and biocompatibility testing.
    • 8. The sample size for the training set: Not applicable. There is no "training set" for physical examination gloves; manufacturing process controls ensure consistent product.
    • 9. How the ground truth for the training set was established: Not applicable. See above.

    Summary of Non-Clinical Performance Data (Reiterated from document):

    The performance test data of the non-clinical tests meet the following standards:

    • ASTM D 3578-2019 Standard Specification for Rubber Examination Gloves
    • ASTM D 5151-2019 Standard Test Method for Detection of Holes in Medical Gloves
    • ASTM D 6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
    • ASTM D 5712-95 (Reapproved 2010) Standard Test Method for the Analysis of Protein in Natural Rubber

    Clinical Testing Summary:

    • "This section is not applicable because clinical data not needed for gloves." (As stated in the document).