    K212438

    Validate with FDA (Live)

    Device Name
    Comfortpro
    Date Cleared
    2021-12-07

    (125 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Comfortpro latex examination powder free gloves is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    COMFORTPRO Latex Examination Powder Free Gloves are manufactured to meet all the current specifications listed under the ASTM Specification D3578-19, Standard Specification for Rubber Examination Gloves. They are made from Natural Rubber Latex. These gloves are Natural in color (No color added) and are powder free.

    AI/ML Overview

    This is a 510(k) summary for a medical device, specifically, a latex examination glove. The document describes the device's characteristics and compares them against a predicate device and established performance standards.

    Here's the breakdown of the acceptance criteria and the study that proves the device meets them, structured as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicStandardAcceptance Criteria (Predicate)Device Performance (Current)Comparison
    Dimensions - LengthASTM D3578-19Min 230 mm for all sizesOverall: Length > 230 mm X-Small: 245 mm Small: 238 mm Medium: 233 mm Large: 240 mmSimilar
    Dimensions - Width (Medium size)ASTM D3578-19Min 95 +/- 10 mm (for medium size)Overall: Min 95+/-10 mm (for medium size) X-Small: 77 mm Small: 81 mm Medium: 93 mm Large: 101 mmSimilar
    Physical Properties - Tensile StrengthASTM D3578-19Before Ageing: Tensile Strength > 18 MPa After Ageing: Tensile Strength > 14 MPaBefore Ageing: Tensile Strength > 18 MPa Actual values: X-Small: 29.6 MPa, Small: 25 MPa, Medium: 26 MPa, Large: 24.9 MPa After Ageing: Tensile Strength > 14 MPa Actual values: X-Small: 25.5 MPa, Small: 24.1 MPa, Medium: 22.2 MPa, Large: 24.4 MPaSimilar
    Physical Properties - Ultimate ElongationASTM D3578-19Before Ageing: Ultimate Elongation > 650% After Ageing: Ultimate Elongation > 500%Before Ageing: Ultimate Elongation > 650% Actual values: X-Small: 810%, Small: 760%, Medium: 680%, Large: 780% After Ageing: Ultimate Elongation > 500% Actual values: X-Small: 820%, Small: 760%, Medium: 650%, Large: 760%Similar
    ThicknessASTM D3578-19Palm: > 0.08 mm Finger: > 0.08 mmPalm: > 0.08 mm Actual values: X-Small: 0.102 mm, Small: 0.087 mm, Medium: 0.10 mm, Large: 0.108 mm Finger: > 0.08 mm Actual values: X-Small: 0.114 mm, Small: 0.109 mm, Medium: 0.122 mm, Large: 0.121 mmSimilar
    Powder Free ResidueASTM D3578-19$\leq$ 2 mg/glove0.68 mg/glove (for Medium)Similar
    Water Tight (1000 ml)ASTM D5151-19 AQL-2.5PassesPassesSimilar
    Extractable ProteinASTM D5712-15200 µg/dm² Max187.1 µg/dm² (for Medium)Meets criteria
    Biocompatibility - Primary Skin IrritationISO 10993-10:2010(E)Under the condition of study, not an irritantUnder the condition of study, not an irritantSame
    Biocompatibility - Dermal SensitizationISO 10993-10:2010(E)Under the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizerSame
    Biocompatibility - In vitro CytotoxicityISO 10993-5:2009(E)Under the conditions of the study, non-cytotoxicUnder the conditions of the study cytotoxic for 100% (neat) test item extract. A follow-up acute systemic toxicity test demonstrated the extract did not present an acute systemic toxicity concern.Meets criteria
    Biocompatibility - Acute Systemic ToxicityISO 10993-11:2017(E)Under the conditions of study, the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concernSame

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each specific test (e.g., how many gloves were tested for length or tensile strength). However, the tests are conducted according to ASTM D3578-19, ASTM D5151-19, ASTM D6124-06, and ASTM D5712-15, which describe the standard test methods and typically include specifications for sampling.

    The data provenance is from non-clinical bench testing and biocompatibility studies of the "Comfortpro" device. The manufacturer is THAI RUBBER INDUSTRY COMPANY LIMITED, located in Thailand. The tests are therefore likely performed by the manufacturer or a third-party lab on behalf of the manufacturer, and the data is prospective as it's generated specifically for this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of device (patient examination glove) does not involve a "ground truth" established by human experts in the typical sense of medical image analysis or diagnostic studies. The performance is assessed against established international consensus standards (ASTM and ISO) for physical properties and biocompatibility. The "ground truth" is not based on expert interpretation but on the quantitative measurement and qualitative assessment defined by these standards.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, the performance is measured against objective standards, not subjective human assessment requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-powered diagnostic device or a device where human interpretation is a primary performance factor.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device (glove), not an algorithm or AI system.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context is defined by international consensus standards (ASTM D3578-19, ASTM D5151-19, ASTM D6124-06, ASTM D5712-15, ISO 10993-10:2010(E), ISO 10993-5:2009(E), ISO 10993-11:2017(E)). These standards specify the test methods and the acceptable ranges for the physical and biological properties of the gloves.

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set for this type of device.