510(k) Data Aggregation

    K Number
    K212260

    Validate with FDA (Live)

    Date Cleared
    2022-04-13

    (267 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sensuva Premium Silicone Personal Lubricant is a silicone-based personal lubricant for penile, anal and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex, polyisoprene and polyurethane condoms.

    Device Description

    Sensuva Premium Silicone Personal Lubricant is a non-sterile, clear, colorless, thin liquid, silicone-based personal lubricant. This device is for penile, anal, and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. It is compatible with natural rubber latex, polyisoprene and polyurethane condoms. Its formulation consists of dimethicone, dimethiconol, and cyclopentasiloxane. Sensuva Premium Silicone Personal Lubricant is sold as an over-the-counter (OTC) device and is provided in polyethylene terephthalate bottles. The 1.93 fl. oz./57 mL bottles are fitted with 20/410 smooth black pump sprayers and clear polypropylene caps. The 4.23 fl. oz./125 mL and 8.12 fl. oz./240 mL bottles are capped with 24/410 smooth black lotion pumps.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a personal lubricant, not an AI/ML medical device. Therefore, the requested information regarding AI/ML device performance, training data, ground truth establishment, and expert involvement is not applicable.

    However, I can extract the acceptance criteria and the studies performed for the Sensuva Premium Silicone Personal Lubricant as described in the document.

    Acceptance Criteria and Device Performance for Sensuva Premium Silicone Personal Lubricant

    The device is a non-sterile, clear, colorless, thin liquid, silicone-based personal lubricant.

    1. Table of acceptance criteria and the reported device performance:
    PropertyAcceptance Criteria (Specification)Reported Device Performance (from studies described)
    AppearanceThin liquidMaintained its specifications during shelf-life
    ColorClear, colorlessMaintained its specifications during shelf-life
    OdorCharacteristicMaintained its specifications during shelf-life
    Viscosity (cps)300-630 cpsMaintained its specifications during shelf-life
    Total aerobic microbial count (TAMC) per USP <61> and <1111>Less than 100 cfu/gMaintained its specifications during shelf-life
    Total yeast and mold count (TYMC) per USP <61> and <1111>Less than 10 cfu/gMaintained its specifications during shelf-life
    Presence of Pathogens per USP <62> (Pseudomonas aeruginosa, Staphylococcus aureus, Salmonella/Shigella, Escherichia coli, Candida albicans)AbsentMaintained its specifications during shelf-life
    BiocompatibilityPass (no adverse biological effects)Passed
    Shelf-Life8.4 monthsMaintained specifications for 8.4 months
    Condom CompatibilityCompatibleCompatible with natural rubber latex, polyisoprene, and polyurethane condoms
    1. Sample size used for the test set and the data provenance:

      • The document does not explicitly state the sample sizes for the individual non-clinical tests (biocompatibility, shelf-life, condom compatibility).
      • The data provenance is from non-clinical performance data (laboratory testing) as described. The country of origin of the data is not specified. All studies appear to be prospective in nature (i.e., tests were conducted on the device to assess its performance against defined criteria).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • This question is not applicable as this is not an AI/ML device or a diagnostic device requiring expert interpretation for ground truth. The acceptance criteria for the physical, chemical, and biological properties are established by recognized standards (e.g., USP, ISO, ASTM).
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable as this is not an AI/ML device or a diagnostic device requiring expert adjudication. The results of the non-clinical tests are objectively measured against defined specifications in established standards.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI/ML device.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is not an AI/ML device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this device's performance is based on established standard specifications and methodologies for physical, chemical, and biological properties relevant to personal lubricants as per recognized guidelines (e.g., USP <61>, <62>, <1111>, ISO 10993 series, ASTM D7661-18). For biocompatibility, the truth is whether the device elicits a toxic or sensitizing response according to the ISO standards. For condom compatibility, it's whether the lubricant degrades the condom material beyond an acceptable threshold per ASTM. For shelf-life, it's whether the product meets its specifications over time under accelerated aging conditions.
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device.
    8. How the ground truth for the training set was established:

      • Not applicable. This is not an AI/ML device.