510(k) Data Aggregation

    K Number
    K212213

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-02-09

    (209 days)

    Product Code
    Regulation Number
    866.3975
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Xpert Xpress MVP test, performed on the GeneXpert Instrument Systems, is an automated qualitative in vitro diagnostic test for the detection of DNA targets from anaerobic bacteria associated with bacterial vaginosis (BV), Candida species associated with vulvovaginal candidiasis, and Trichomonas vaginalis. The Xpert Xpress MVP test uses clinician-collected and self-collected vaginal swabs (collected in a clinical setting) from patients who are symptomatic for vaginitis/vaginosis. The Xpert Xpress MVP test utilizes real-time polymerase chain reaction (PCR) for the amplification of specific DNA targets and utilizes fluorogenic target-specific hybridization probes to detect and differentiate DNA from:

    • . Organisms associated with bacterial vaginosis (detected organisms not reported individually)
      • o Atopobium spp. (Atopobium vaginae, Atopobium novel species CCUG 55226)
      • Bacterial Vaginosis-Associated Bacterium 2 (BVAB2) o
      • Megasphaera-1 o
    • Candida spp. (C. albicans, C. tropicalis, C. parapsilosis, C. dubliniensis, species not differentiated)
    • Candida glabrata/Candida krusei (species not differentiated)
    • . Trichomonas vaginalis

    The Xpert Xpress MVP test is intended to aid in the diagnosis of vaginal infections in women with a clinical presentation consistent with bacterial vaginosis, vulvovaginal candidiasis, or trichomoniasis.

    Device Description

    The Xpert® Xpress MVP test is an automated in vitro diagnostic test for qualitative detection of DNA targets from anaerobic bacteria associated with bacterial vaginosis, Candida species associated with vulvovaginal candidiasis, and Trichomonas vaginalis, the agent of trichomoniasis. The Xpert Xpress MVP test is performed on GeneXpert Instrument Systems.

    The GeneXpert Instrument Systems automate and integrate sample preparation, nucleic acid extraction and amplification, and detection of the target sequences in simple or complex samples using real-time PCR assays. The systems consist of an instrument, computer, and preloaded software for running tests and viewing the results. The systems require the use of single-use disposable cartridges that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained, cross-contamination between samples is minimized.

    The Xpert Xpress MVP test includes reagents for the detection of DNA from BV organisms, Candida species, and Trichomonas vaginalis from vaginal swab samples. A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge utilized by the GeneXpert System instrument. The SPC is present to control for adequate processing of the sample and to monitor for the presence of potential inhibitor(s) in the PCR reaction. The SPC also ensures that the PCR reaction conditions (temperature and time) are appropriate for the amplification reaction and that the PCR reagents are functional. The PCC verifies reagent rehydration, PCR tube filling, and confirms that all reaction components are present in the cartridge including monitoring for probe integrity and dye stability.

    The Xpert Xpress MVP test is designed for use with the following specimens collected from symptomatic individuals: self-collected vaginal swabs (collected in a clinical setting) and clinician-collected vaginal swabs. The swab transport reagent included in the Xpert Swab Specimen Collection Kit is designed to collect and preserve patient specimens to allow transport to the laboratory prior to analysis with the Xpert Xpress MVP test.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study proving the device meets those criteria, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as distinct numerical targets for sensitivity, specificity, and agreement rates in the provided document. However, the performance outcomes of the Xpert Xpress MVP test are presented and can be interpreted as the device meeting the performance standards considered acceptable for its intended use, especially given the FDA's 510(k) clearance based on "substantial equivalence." The document compares the device's performance to an FDA-cleared predicate device.

    For the purpose of this table, "Acceptance Criteria" will be inferred from the reported performance, as it highlights what the device achieved and what was deemed sufficient for clearance.

    Test ParameterAcceptance Criteria (Implied)Reported Performance (Xpert Xpress MVP)Sample TypeComparator/Reference MethodGround Truth Type
    Bacterial Vaginosis (BV)High PPA and NPAPPA: 93.8% (531/566)Clinician-collected (CVS)FDA-cleared NAATNAAT results
    NPA: 93.8% (808/861)Clinician-collected (CVS)FDA-cleared NAATNAAT results
    PPA: 94.0% (533/567)Self-collected (SVS)FDA-cleared NAATNAAT results
    NPA: 92.9% (794/855)Self-collected (SVS)FDA-cleared NAATNAAT results
    Candida groupHigh Sensitivity and SpecificitySensitivity: 98.0% (396/404)Clinician-collected (CVS)Yeast culture + mass spectrometryCulture + MS
    Specificity: 94.6% (984/1040)Clinician-collected (CVS)Yeast culture + mass spectrometryCulture + MS
    Sensitivity: 97.5% (393/403)Self-collected (SVS)Yeast culture + mass spectrometryCulture + MS
    Specificity: 92.1% (954/1036)Self-collected (SVS)Yeast culture + mass spectrometryCulture + MS
    Candida glabrata/krusei (C. glab-krus)High Sensitivity and SpecificitySensitivity: 93.6% (44/47)Clinician-collected (CVS)FDA-cleared NAAT (for discrepants)Culture + MS (primary), NAAT (discrepant)
    Specificity: 99.6% (1392/1397)Clinician-collected (CVS)FDA-cleared NAAT (for discrepants)Culture + MS (primary), NAAT (discrepant)
    Sensitivity: 97.8% (45/46)Self-collected (SVS)FDA-cleared NAAT (for discrepants)Culture + MS (primary), NAAT (discrepant)
    Specificity: 99.4% (1384/1393)Self-collected (SVS)FDA-cleared NAAT (for discrepants)Culture + MS (primary), NAAT (discrepant)
    Sensitivity (Contrived): 99.0% (98/99)Clinician-collected (CVS)Not specified (presumably internal reference)Contrived specimens
    Specificity (Contrived): 96.4% (27/28)Clinician-collected (CVS)Not specified (presumably internal reference)Contrived specimens
    Trichomonas vaginalis (TV)High PPA and NPAPPA: 97.3% (73/75)Clinician-collected (CVS)Patient Infected Status (PIS) algorithm (FDA-cleared NAAT + TV culture)PIS algorithm
    NPA: 99.6% (1332/1337)Clinician-collected (CVS)Patient Infected Status (PIS) algorithm (FDA-cleared NAAT + TV culture)PIS algorithm
    PPA: 97.3% (72/74)Self-collected (SVS)Patient Infected Status (PIS) algorithm (FDA-cleared NAAT + TV culture)PIS algorithm
    NPA: 99.8% (1330/1333)Self-collected (SVS)Patient Infected Status (PIS) algorithm (FDA-cleared NAAT + TV culture)PIS algorithm

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Clinical Study/Test Set):
      • Patients: 1,478 female patients (14 to ≥ 50 years of age).
      • Vaginal Swabs Tested: 2,947 (one self-collected vaginal swab (SVS) and five clinician-collected vaginal swab (CVS) specimens per patient).
    • Data Provenance:
      • Country of Origin of the Data: United States (multi-site clinical study with 12 sites from geographically diverse locations in the U.S.).
      • Retrospective or Prospective: Prospective observational, method comparison clinical study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number or qualifications of experts (e.g., radiologists, pathologists) used to establish the ground truth for the clinical test set. The ground truth for the clinical study was established by comparator methods (FDA-cleared NAATs, yeast culture followed by mass spectrometry, and a PIS algorithm). These methods are analytical laboratory tests, not dependent on expert visual review.

    4. Adjudication Method for the Test Set

    • The document states: "When applicable, investigation of discrepant results was performed by testing specimens with another FDA-cleared NAAT."
    • This indicates a form of adjudication for discrepant results, where a third, independent, FDA-cleared NAAT was used to resolve disagreements between the Xpert Xpress MVP test and the initial comparator method. The specific rule (e.g., 2 out of 3 agreement) for this discrepancy resolution is not detailed, but the use of an independent NAAT as a tie-breaker or confirming tool is implied.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done.
    • This study evaluates an in vitro diagnostic (IVD) device that detects nucleic acid sequences from microorganisms using real-time PCR. It is not an imaging-based AI device that would typically involve human readers interpreting images. Therefore, the concept of human readers improving with AI vs. without AI assistance does not apply to this type of device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, the primary performance evaluation of the Xpert Xpress MVP test in the clinical study was standalone.
    • The device is an automated, qualitative in vitro diagnostic test that performs sample preparation, nucleic acid extraction and amplification, and detection, and provides results "within 60 minutes." The clinical performance tables (Table 5-13) represent the direct output of the device compared to reference methods, without human interpretation of the device's signal directly impacting the final result reported by the device itself.
    • Human involvement is in specimen collection, loading the cartridge, and reviewing the system's final reported result for the pathogen. The device's diagnostic output for a given sample is fully automated.

    7. The Type of Ground Truth Used

    The ground truth varied by the target organism:

    • Bacterial Vaginosis (BV): An FDA-cleared nucleic acid amplification test (NAAT).
    • Candida group and Candida glabrata/krusei: Yeast culture followed by mass spectrometry for species identification. For Candida glabrata/krusei, there was also a "contrived" study, meaning the ground truth was based on known concentrations of the organisms.
    • Trichomonas vaginalis (TV): A Patient Infected Status (PIS) algorithm that included results from an FDA-cleared NAAT and TV culture.
    • Discrepancy Resolution: For all targets, a second FDA-cleared NAAT was used for investigation of discrepant results, effectively serving as an adjudication method to establish the clinical ground truth for those specific samples.

    8. The Sample Size for the Training Set

    The document describes the clinical study as a "performance evaluation" and "method comparison clinical study" used to demonstrate substantial equivalence. It does not explicitly reference or describe a separate "training set" for the device's algorithm in the context of machine learning, because this is a molecular diagnostic test based on PCR, not an adaptable AI algorithm that is trained on data in the traditional sense. The development of such a device involves assay design and optimization rather than machine learning training sets.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no explicit mention of a "training set" in the context of an AI/ML algorithm for this PCR-based diagnostic device, the concept of establishing ground truth for a training set as typically described for AI/ML devices doesn't apply. The development and validation of the device would have involved extensive laboratory (non-clinical) studies, including analytical sensitivity (LoD), analytical reactivity (inclusivity), analytical specificity (exclusivity), microbial interference, competitive interference, interfering substances, and carry-over contamination studies (as described in Section 5.4), where the "ground truth" for these studies would be precisely controlled laboratory-prepared samples with known concentrations of target organisms and potential interferents.