510(k) Data Aggregation

    K Number
    K212112

    Validate with FDA (Live)

    Date Cleared
    2021-11-18

    (135 days)

    Product Code
    Regulation Number
    870.2340
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Cardiologs Platform (Also known as Cardiologs ECG Analysis Platform) is intended for use by qualified healthcare professionals for the assessment of arrhythmias using ECG data in the adult and pediation. The product supports downloading and analyzing data recorded in compatible formats from any device used for the arrhythmia diagnostics such as Holter, event recorder, 12 lead ambulatory ECG devices, or other similar devices when assessment of the rhythm is necessary. The Cardiologs Platform can also be electronically interfaced and perform analysis with data transferred from other computer-based ECG systems, such as an ECG management system. The Cardiologs Platform provides ECG signal processing and analysis, QRS and Ventricular Ectopic Beat detection, QRS feature extraction, interval measurement, heart rate measurement and rhythm analysis.

    The Cardiologs Platform is not for use in life supporting or sustaining systems or ECG monitor and Alarm devices.

    The product can be integrated into computerized ECG monitoring devices. In this case, the medical device manufacturer will identify the indication for use depending on the application of their device.

    Cardiologs Platform interpretation results are not intended to be the sole means of diagnosis. It is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information.

    Device Description

    Cardiologs Holter Platform is a software only medical device with the following components made up of:

    • An interface which provides tools to measure, analyze, and review numerous ECGs;
    • An automated proprietary ECG interpretation support algorithm which measures and analyzes ECGs to provide physicians supportive information for ECG analysis.

    Cardiologs Holter Platform can be accessed through a network connection/wireless device when the electrocardiograph allows for digital ECG upload, or locally (stand-alone algorithm).

    Cardiologs Holter Platform works in the following sequence:

    • . Upload of digital ECG file to Cardiologs' secured hosting databases (for health data i.e. HIPAA compliant);
      • Manual upload: via the web-interface which allows the selection of files to upload on ● the evaluating user's computer
      • . Direct upload: with no manual intervention, in the specific cases where the user's hardware is already connected to the Cardiologs' Application Programming Interface (or API): the ECG is automatically sent to Cardiologs' servers.
    • II. Processing of the uploaded ECG file;
    • III. Analysis of the uploaded ECG performed by Cardiologs' proprietary algorithm, which labels the ECG:
      • . Delineation (detection of at least P waves, QRS complexes and T waves on the ECG signal). The output format is a sequence of elements "wave type/start time":
      • Abnormality labels: the algorithm provides probability scores on a predefined set of . abnormality labels (multi-label classification);
    • IV. Cardiologs Holter Platform displays the resulting analysis, along with the original ECG signal and a variety of tools (ruler, zoom, etc.). Results are made available either on the Cardiologs interface if the user's hardware allows for digital ECG upload, or on an API in cases where the service is available as part of a package with an ECG machine
    • V. Cardiologs Holter Platform then allows for editing and/or confirmation of the measurements and parameters by the analyzing physician.

    The resulting information is stored on a secured database.

    AI/ML Overview

    The provided text describes the Cardiologs Holter Platform (K212112) and its substantial equivalence to a predicate device. However, it does not contain specific acceptance criteria or a detailed study description with performance metrics in the format requested.

    It primarily focuses on the device's indications for use, a comparison of features with its predicate device, and a general statement about testing in compliance with recognized standards.

    Therefore, I cannot fulfill your request for a table of acceptance criteria and reported device performance, sample size, ground truth details, or comparative effectiveness study results from the provided text.

    The closest information available is:

    1. General Statement about Testing:

    • "Testing described in this 510(k) consisted of verification of all design input requirements and product specifications. All clinical input requirements were validated against a gold standard."
    • "All software validation testing was completed successfully and met all requirements."

    2. Mention of Ground Truth (for clinical input requirements):

    • "All clinical input requirements were validated against a gold standard." The specific type of "gold standard" is not elaborated.

    3. No information regarding:

    • A specific table of acceptance criteria and reported device performance.
    • Sample sizes for test sets or data provenance.
    • Number or qualifications of experts used for ground truth.
    • Adjudication methods.
    • Multi-reader multi-case (MRMC) comparative effectiveness study or effect size.
    • Standalone performance (though it mentions the platform "can be accessed through a network connection/wireless device ... or locally (stand-alone algorithm)").
    • Sample size for the training set.
    • How ground truth for the training set was established.