Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K211995

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-08-27

    (60 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sonendo Filling Material 5C is intended for permanent obturation of the root canal following root canal treatment.

    Device Description

    The Sonendo Filling Material 5C is a two-part filling material developed for permanent root canal filling. It is a water-soluble acrylate which is radiopaque and hardens with a chemical curing system. The Sonendo Filling Material 5C is packaged in plastic syringes with a syringe mixer and a dispensing tip. The syringe and dispensing tips are the only patient contacting devices packaged with the Sonendo Filling Material 5C and are commercially available.

    AI/ML Overview

    Based on the provided text, the device in question is a dental root canal filling material, and the provided document is an FDA 510(k) summary for its clearance. This document type focuses on demonstrating "substantial equivalence" to a legally marketed predicate device, rather than proving a medical device's performance through a clinical comparative effectiveness study against specific acceptance criteria like an AI algorithm would.

    Therefore, the requested information regarding acceptance criteria, sample sizes, expert ground truth establishment, adjudication methods, multi-reader multi-case studies, standalone performance, and training set details cannot be fully provided from this document. This document specifies that no human clinical testing was conducted to support substantial equivalence.

    However, I can extract the relevant performance data that was conducted, which focuses on device properties compared to industry standards and the predicate device.

    Here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    For a root canal filling material, performance is evaluated based on physical and chemical properties as defined by ISO standards, rather than diagnostic accuracy metrics. The "acceptance criteria" are implied by the ISO standard requirements, and "device performance" is the measured value for the Sonendo Filling Material 5C compared to the predicate and the standard.

    Element (Acceptance Criteria per ISO 6876:2012)Sonendo Filling Material 5C (Subject Device)Predicate Device (EndoREZ Dual Cure)Comparison / Meets Criteria?
    Flow (≥ 17mm)33.45 mm17.31 mmPassed ISO 6876 test requirements
    Working Time (implied by predicate & standard)32.98 mm17.38 mmPassed ISO 6876 test requirements
    Setting Time (Between 1 - 24 hours)60 minutes, 7 seconds25 minutes, 46 secondsPassed ISO 6876 test requirements
    Film Thickness (≤ 50 μm)5 μm5 μmPassed ISO 6876 test requirements
    Solubility & Disintegration (≤ 3.0% by mass; No Disintegration)2.03% (No disintegration)0.77% (No disintegration)Passed ISO 6876 test requirements
    Radio-opacity (≥ 3 mm of Aluminum)3.10 mm Al6.6 mm AlPassed ISO 6876 test requirements

    Note: The "Working Time" and "Setting Time" acceptance criteria are stated as "Passed ISO 6876 test requirements" rather than specific numeric thresholds for the subject device.

    2. Sample sizes used for the test set and the data provenance

    • Test Set Sample Size: Not explicitly stated in terms of number of "samples" for each test. The ISO testing involves specific laboratory procedures using test specimens, but the quantity of these specimens is not detailed.
    • Data Provenance: The tests were conducted to comply with ISO standards. The location of the testing is not specified, but this is a U.S. FDA submission. These are laboratory/benchtop tests, not clinical data from patients.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. For this type of device (root canal filling material) and the testing performed (physical/chemical properties per ISO standards), human expert "ground truth" as you would define for an AI diagnostic algorithm is not relevant. The "ground truth" is the objective measurement of the material's properties according to the specified ISO standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This concept relates to reconciling discrepancies in human expert interpretations, which is not part of the ISO standard physical/chemical testing for a dental material.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. The document explicitly states: "No human clinical testing was conducted to support substantial equivalence." An MRMC study is a clinical study involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This device is a physical dental material, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The "ground truth" for the performance data presented is based on measurements derived from standardized laboratory tests according to ISO 6876:2012, such as flow, setting time, film thickness, solubility, and radio-opacity. Biocompatibility was assessed against ISO 10993 standards.

    8. The sample size for the training set

    • Not applicable. This is not an AI/machine learning device; it does not have a "training set."

    9. How the ground truth for the training set was established

    • Not applicable. As above, no training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1