Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities.

    The SYSTEM 1E Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C, and rinses the load with extensively treated* potable water. After completion of a cycle, critical devices should be used immediately; semi-critical devices should be used immediately or may be handled and stored in a manner similar to that of high level disinfected endoscopes. Critical devices not used immediately should be processed again before use.

    The SYSTEM 1E Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.

    • The extensive treatment of EPA potable water consists of:
      1. Pre-filtration through two pre-filters:
      • · Pre-filter A is a gross depth filter that removes approximately 2.5 micron or larger particles/contaminants.
      • · Pre-filter B is a surface filter that removes particles/contaminants > 0.1 micron.
      1. UV Irradiation:
      • · During transit through the UV water treatment chamber, a UV dose sufficient to achieve a > or equal to 6-log reduction of MS2 virus is delivered to the water.
    • 3.0.1 micron filtration:
      • · The water prepared by pre-filtration is filtered through redundant, 0.1-micron (absolute rated) membranes to remove bacteria. fungi and protozoa > 0.1 micron.

    The SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800 is intended for liquid chemical sterilization of cleaned, immersible, and reusable semi-critical medical devices and their accessories in healthcare facilities.

    The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution(> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.

    The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.

    The SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900 is intended for liquid chemical sterilization of cleaned, immersible, and reusable sensitive medical devices and their accessories in healthcare facilities.

    The SYSTEM 1 endo Processor automatically dilutes the S40 Sterilant Concentrate to its use dilution (> 1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 45.5 to 60°C and rinses the load with 0.2 micron filtered water.

    The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.

    Device Description

    The SYSTEM 1E Liquid Chemical Sterilant Processing System is a liquid chemical sterilization system, utilizing peracetic acid to process totally immersible, heat sensitive, flexible and rigid endoscopes and their accessories, and microsurgical instruments. The system consists of the SYSTEM 1E Processor and the S40 Sterilant Concentrate, interchangeable processing trays/containers and Quick Connects. The current submission is provided to describe modifications for:

    • . Obsolescence and replacement of interface board and connector
    • Obsolescence and replacement of silicone tubing ●
    • S40 Sterilant Concentrate design modification ●

    The SYSTEM 1E Processor is an automated, self-contained device which creates and maintains the conditions necessary for liquid chemical sterilization in 6 minutes. Following processing, the liquid chemically sterilized articles are rinsed with extensively treated water produced by passing EPA potable tap water through pre-filters, an ultraviolet light treatment subsystem, and then through two 0.1micron filter membranes. The processor, which is computer controlled and continually monitored, provides printed documentation of each cycle.

    S40 Sterilant Concentrate is a single use chemical sterilant concentrate developed for use in the SYSTEM 1E Processor. The active ingredient in S40 Sterilant Concentrate, peracetic acid, is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.

    The interchangeable processing trays/containers are made to accommodate a variety of instrument types, models and procedure specific sets. Each container is designed to maintain instruments in appropriate position while specific Quick Connects for the SYSTEM 1E Processor, if required, facilitate delivery of the liquid chemical sterilant use dilution and rinse water to internal channels.

    The SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800 is a liquid chemical sterilization system, utilizing peracetic acid to process totally immersible semi-critical medical devices and their accessories. The system consists of the SYSTEM 1 endo Processor and S40 Sterilant Concentrate, interchangeable Processing Trays/Containers, and Quick Connects. The current submission is provided to describe modifications for:

    • Obsolescence and replacement of silicone tubing
    • S40 Sterilant Concentrate design modification .

    The SYSTEM 1 endo Processor is an automated. self-contained device that uses S40 Sterilant Concentrate to create and maintain the conditions necessary for liquid chemical sterilization in 6 minutes. Following processing, the liquid chemically sterilized articles are rinsed with 0.2 micron filtered potable water and are ready for use or may be prepared for storage. The processor, which is computer controlled and continually monitored, provides printed documentation of each cycle.

    The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate. Upon loading the single-use cup, the active ingredient in S40 - peracetic acid - is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.

    The interchangeable processing trays/containers are made to accommodate a variety of semi-critical instrument types and models. Each container is designed to maintain instruments in position while specific SYSTEM 1 endo Quick Connects, if required, facilitate delivery of the sterilant use-solution and rinse water to internal channels.

    The SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900 is a liquid chemical sterilization system, utilizing peracetic acid to process totally immersible semi-critical heat-sensitive medical devices and their accessories. The system consists of the SYSTEM 1 endo Processor and S40 Sterilant Concentrate, interchangeable Processing Trays/Containers, and Quick Connects. The current submission is provided to describe modifications for:

    • Obsolescence and replacement of silicone tubing
    • S40 Sterilant Concentrate design modification

    The SYSTEM 1 endo Processor is an automated, self-contained device that uses S40 Sterilant Concentrate to create and maintain the conditions necessary for liquid chemical sterilization in 6 minutes. After LCS processing, the liquid chemically sterilized articles are rinsed with 0.2 micron filtered potable water and are ready for use or may be prepared for storage.

    The SYSTEM 1 endo Processor uses only S40 Sterilant Concentrate. Upon loading the single-use cup, the active ingredient in S40 - peracetic acid - is combined with inert ingredients (builders) to form a use dilution which inhibits corrosion of metals, polymers and other materials.

    The interchangeable processing trays/containers are made to accommodate a variety of semi-critical instrument types and models. Each container is designed to maintain instruments in position while specific SYSTEM 1 endo Quick Connects, if required, facilitate delivery of the sterilant use-solution and rinse water to internal channels.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from STERIS Corporation to the FDA regarding several Liquid Chemical Sterilant Processing Systems. It's important to understand that this document describes a sterilization device, not an AI/ML-driven diagnostic or image analysis device. Therefore, many of the requested criteria related to AI/ML performance (e.g., sample size for training/test sets, expert adjudication, MRMC studies, standalone algorithm performance, ground truth for training data) are not applicable to this type of medical device submission.

    The acceptance criteria provided are for the functional and physical performance of a chemical sterilant processing system and its associated sterilant, not for an AI model's diagnostic accuracy.

    Here's a breakdown of the requested information based on the provided document:


    Device Description and Purpose:

    The devices are the "SYSTEM 1E Liquid Chemical Sterilant Processing System," "SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800," and "SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900," along with the "S40 Sterilant Concentrate." These systems are intended for liquid chemical sterilization of cleaned, immersible, and reusable heat-sensitive medical devices (critical and semi-critical) in healthcare facilities. They utilize peracetic acid as the sterilant.

    The submission is specifically for modifications to existing cleared devices (predicate K210737), primarily involving:

    • Obsolescence and replacement of interface board and connector (for SYSTEM 1E)
    • Obsolescence and replacement of silicone tubing (for all models)
    • S40 Sterilant Concentrate design modification (for all models)

    1. Table of Acceptance Criteria and Reported Device Performance

    The tables below synthesize the acceptance criteria and the results as presented in the "Summary of verification activities" (Table 3 in each section of the document).

    For SYSTEM 1E Liquid Chemical Sterilant Processing System (Page 15, Table 3)

    ChangeTestAcceptance CriteriaResult
    Obsolescence and replacement of interface board and connectorFunctional testingThe modification does not affect the operation of the device.Pass
    Reliability testingThe modification does not affect the operation of the device.Pass
    EMC/EMI and electrical safetyThe device must meet UL 61010-1:2010 Third Edition and UL 61010-2-040:2015-7-07Pass
    Obsolescence and replacement of silicone tubingFunctional testingThe modification does not affect the operation of the device.Pass
    Form and fitThe replacement must fit within the device.Pass
    Physical propertiesThe replacement must be equivalent to the current tubing.Pass
    S40 Sterilant Concentrate design modificationStability of sterilantThe design modification does not affect the stability of the sterilant.Pass
    Delivery of sterilantThe design modification does not affect delivery of the sterilant.Pass

    For SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6800 (Page 25, Table 3)

    ChangeTestAcceptance CriteriaResult
    Obsolescence and replacement of silicone tubingFunctional testingThe modification does not affect the operation of the device.Pass
    Form and fitThe replacement must fit within the device.Pass
    Physical propertiesThe replacement must be equivalent to the current tubing.Pass
    S40 Sterilant Concentrate design modificationStability of sterilantThe design modification does not affect the stability of the sterilant.Pass
    Delivery of sterilantThe design modification does not affect delivery of the sterilant.Pass

    For SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900 (Page 36, Table 3)

    ChangeTestAcceptance CriteriaResult
    Obsolescence and replacement of silicone tubingFunctional testingThe modification does not affect the operation of the device.Pass
    Form and fitThe replacement must fit within the device.Pass
    Physical propertiesThe replacement must be equivalent to the current tubing.Pass
    S40 Sterilant Concentrate design modificationStability of sterilantThe design modification does not affect the stability of the sterilant.Pass
    Delivery of sterilantThe design modification does not affect delivery of the sterilant.Pass

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes (e.g., number of units tested, number of cycles performed) for the functional, reliability, EMC/EMI, form and fit, physical properties, stability, or delivery tests. It simply states "Functional testing," "Reliability testing," etc., and confirms the Pass result.

    Data provenance (e.g., country of origin, retrospective/prospective) is not mentioned as these were laboratory-based engineering and performance tests for a sterilization device, not clinical data for an AI/ML diagnostic.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This criterion is not applicable. The "ground truth" for this device relates to its ability to function as a sterilizer and its components meeting engineering specifications. This is established through predefined performance requirements (e.g., UL standards for electrical safety, physical property equivalence, stability of sterilant), not through expert consensus on medical images or diagnoses. Validation would be performed by engineers and quality control personnel according to established test protocols.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This criterion is not applicable. Adjudication methods like 2+1 are used for resolving disagreements among human readers or between AI and human reads in diagnostic studies. For a sterilization device, test results are typically objective (e.g., does it pass/fail a function test, does it meet a physical specification).


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This criterion is not applicable. This is not an AI-assisted diagnostic device. No human readers or interpretations are involved in its primary function, which is sterilization.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This criterion is not applicable. This is not an algorithm, but a physical medical device.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context refers to the engineering and performance specifications for the device and its components. These include:

    • Predefined functional requirements (e.g., device operation, sterilant delivery).
    • Compliance with recognized electrical safety and electromagnetic compatibility (EMC/EMI) standards (e.g., UL 61010-1, UL 61010-2-040).
    • Physical properties matching existing components.
    • Chemical stability of the sterilant.

    For the basic germicidal efficacy of the sterilant mentioned in the predicate device comparison (Tables 2 on pages 11-13, 22-23, and 32-34), the ground truth for microbial efficacy (e.g., "Meets efficacy requirements," "Solution is bactericidal") would be established through laboratory microbiological testing (e.g., AOAC Official Methods, EPA Viricidal Testing), which involves culturing specific microorganisms and observing their reduction or elimination after exposure to the sterilant. This is not an AI/ML-related ground truth.


    8. The sample size for the training set

    This criterion is not applicable. This is not an AI/ML device; there is no "training set."


    9. How the ground truth for the training set was established

    This criterion is not applicable. There is no AI/ML training set for this device. Ground truth for its performance is established through adherence to physical and chemical engineering specifications and established microbiological testing protocols.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1