510(k) Data Aggregation

    K Number
    K211451

    Validate with FDA (Live)

    Date Cleared
    2021-08-25

    (107 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable surgical mask (non sterile) is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non sterile.

    Device Description

    The Disposable surgical mask (non sterile) is single use, three-layer, flat-pleated style with ear straps and nose piece. The mask is manufactured with three layers, the inner and outer layers are made of nonwoven fabrics, and the middle layer is made of melt blown fabrics. The ear straps are held in place over the users' mouth and nose by two elastic ear straps welded to the facemask. The elastic ear straps are not made with natural rubber latex. The nose piece on the layers of facemask is to allow the user to fit the facemask around their nose, which is made of PE (polyethylene) with dual-Galvanized wire. The Disposable surgical mask (non sterile) will be provided in blue. The masks are sold non-sterile and are intended to be single use, disposable devices.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a Disposable Surgical Mask (non-sterile), K211451, submitted by Qinhuangdao Taizhi Medical Technology Co., Ltd. The document primarily focuses on demonstrating substantial equivalence to a predicate device (K182515, Surgical Face Mask by Wuhan Dymex Healthcare Co., Ltd.) through non-clinical testing.

    Based on the provided text, the device in question is a Disposable Surgical Mask, not an AI or imaging device. Therefore, many of the requested criteria (e.g., sample size for test set/training set, expert qualification, adjudication method, MRMC study, standalone performance, type of ground truth) typically associated with the development and validation of AI/ML-based medical devices or diagnostic tools are not applicable.

    The acceptance criteria and study proving the device meets them are related to the physical and performance characteristics of a surgical mask, as per established standards and guidance for such products.

    Here's the information extracted from the document, tailored to the nature of this device:


    Acceptance Criteria and Device Performance for Disposable Surgical Mask (K211451)

    The device demonstrates conformance to recognized standards and requirements for surgical face masks through non-clinical performance and biocompatibility testing. The testing was conducted to establish substantial equivalence to a predicate device (K182515).

    1. Table of Acceptance Criteria and Reported Device Performance

    The following table summarizes the performance testing results against established acceptance criteria for surgical masks (specifically, Level 2, ASTM F2100-19):

    ItemsPerformance ReportAcceptance Criteria (Level 2, ASTM F2100-19)Result
    Bacterial filtration efficiency (BFE) (%)99.9%≥98%Pass
    Differential pressure (mmH2O/cm²)2.8-3.6 mmH2O/cm²<6.0 mmH2O/cm²Pass
    Sub-micron particulate filtration efficiency at 0.1 micron, % (PFE)99.48% - 99.95%≥98%Pass
    Resistance to penetration by synthetic blood, Minimum pressure in mmHg for pass resultTest 1-3: 32 of 32 test articles passed at 120mmHg29 of 32 test articles passed at 120mmHgPass
    Flame spreadClass 1, Non FlammableClass 1Pass

    Biocompatibility Testing:

    ItemProposed Device FindingResult
    CytotoxicityUnder the conditions of the study, the device is noncytotoxic.Pass
    IrritationUnder the conditions of the study, the device is nonirritating.Pass
    SensitizationUnder the conditions of the study, the device is nonsensitizing.Pass

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The exact sample size for each specific non-clinical test (e.g., number of masks tested for BFE, PFE, etc.) is not explicitly stated, beyond "32 of 32 test articles" for synthetic blood penetration. However, the testing was conducted according to recognized standards (e.g., ASTM F2100-19), which would prescribe appropriate sample sizes for each test.
    • Data Provenance: The origin of the data is from required non-clinical performance and biocompatibility testing performed on the manufactured device. The location of the test facility is not specified beyond the manufacturer's location in China. This is prospective testing conducted specifically for regulatory submission.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • This is not applicable for a surgical mask. "Ground truth" in this context refers to objective scientific measurements and conformance to standardized performance criteria, not expert consensus on diagnostic interpretations. The tests themselves are the "ground truth." These tests are performed by qualified laboratory personnel following established protocols.

    4. Adjudication Method for the Test Set

    • Not applicable. Performance metrics are objectively measured against pre-defined thresholds.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • Not applicable. This device is a surgical mask, not an AI or diagnostic tool requiring human reader studies.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This device is a physical product, not an algorithm.

    7. The Type of Ground Truth Used

    • The "ground truth" for this device is based on objective measurements obtained through standardized laboratory testing for physical and performance characteristics (e.g., filtration efficiency, differential pressure, fluid resistance, flame spread) and biocompatibility, as defined by recognized standards like ASTM F2100-19.

    8. The Sample Size for the Training Set

    • Not applicable. This device is not an AI/ML product developed using training datasets.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable.