Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K211194

    Validate with FDA (Live)

    Device Name
    Molekule Air Pro
    Manufacturer
    Date Cleared
    2021-08-20

    (121 days)

    Product Code
    Regulation Number
    880.6500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Molekule Air Pro air purifier is a device intended for medical purposes that is used to capture 95% of particulate matter and destroy bacteria, mold, and viruses by exposure to ultraviolet radiation when operated in Auto Mode Standard or manual mode at fan speed 2 or higher.

    The Molekule Air Pro air purifier has been demonstrated to entrain and destroy the following bioaerosols under the following exposure/working conditions:

    Test ItemAverage Net Log Reduction / Time @ Fan Speed 6. Room Temperature Test
    BacteriaEscherichia Coli$4.20 \pm 0.11$ / 90 minutes
    Bacteria EndosporesBacillus Subtilis$4.02 \pm 0.23$ / 30 minutes
    Mold SporesAspergillus Brasiliensis$4.15 \pm 0.06$ / 60 minutes
    VirusMS2 bacteriophage$4.38 \pm 0.15$ / 30 minutes
    Test ItemSingle Pass Mechanical Filtration Efficiency
    Particulate Matter0.3 to 1.0 micron size particles95% or greater according to ASHRAE 52.2
    Device Description

    The Molekule Air Pro air purifier is an ultraviolet air purifier that employs a photo electrochemical oxidation (PECO) ultraviolet air purification technology that destroys bacteria and viruses in air. The Molekule Air Pro may be used in medical facilities and in the home. A fan within the unit draws air up from vents in the bottom of the unit, up through the PECO filter (where it is exposed to low energy UV-A 320 - 400 nm light) and out through the vents on the top of the unit. PECO is an air purification technology that oxidizes microorganisms, including bacteria, molds and viruses. PECO works by shining UV-A light on the surface of the catalytic filter to initiate a chemical reaction that generates hydroxyl radicals. These radicals combine with microbiological contaminants that are captured on the filter. Once combined, a chemical reaction takes place destroying the contaminants.

    The Molekule Air Pro is a freestanding device. It is a standalone device with an LCD display and a capacitive touchscreen user interface. Integrated WLAN provides a secondary means for controlling the device from the Molekule Android or iOS application.

    Air Pro also includes a particulate matter sensor, particulate matter indicator, and Auto Protect Mode, which controls fan speed based on particulate levels detected in the use environment.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Molekule Air Pro:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test ItemAcceptance CriteriaReported Device Performance (Average Net Log Reduction / Time @ Fan Speed 6. Room Temperature Test)
    Bioaerosol Destruction
    Bacteria (Escherichia Coli)4 log reduction (99.99%)$4.20 \pm 0.11$ / 90 minutes
    Bacteria Endospores (Bacillus Subtilis)4 log reduction (99.99%)$4.02 \pm 0.23$ / 30 minutes
    Mold Spores (Aspergillus Brasiliensis)4 log reduction (99.99%)$4.15 \pm 0.06$ / 60 minutes
    Virus (MS2 bacteriophage)4 log reduction (99.99%)$4.38 \pm 0.15$ / 30 minutes
    Particulate Matter Filtration
    0.3 to 1.0 micron size particles95% or greater according to ASHRAE 52.2Required filtration efficiency 95% or greater on 0.3 to 1.0 micron size particles was achieved.

    Performance at Different Fan Speeds (MS2 bacteriophage):

    Fan SpeedAcceptance Criteria (4 log reduction)Reported Device Performance (Average net log reduction / time)
    Speed 24 log reduction (99.99%)$4.54 \pm 0.07$ / 120 mins
    Speed 34 log reduction (99.99%)$4.58 \pm 0.23$ / 60 mins
    Speed 64 log reduction (99.99%)$4.38 \pm 0.15$ / 30 mins

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes (number of units tested or number of replicates for each test) for the bioaerosol destruction and particulate filtration tests. It mentions "three PECO filters" were used for the fractional efficiency test.

    The data provenance is not specified beyond being "nonclinical tests" conducted to support substantial equivalence. Information on the country of origin or whether the data was retrospective or prospective is not provided.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not applicable and not provided in the document. The "ground truth" for these performance tests is based on objective measurements of bioaerosol reduction and particle filtration efficiency, not on expert interpretation of observational data.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided. Adjudication methods are typically used in studies involving human interpretation or clinical outcomes, which are not relevant for these specific engineering and microbiology performance tests.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study was not done. The Molekule Air Pro is an air purification device, and its performance is evaluated through laboratory tests for bioaerosol destruction and particle filtration, not through human reader interpretation of medical images or data.

    6. Standalone (Algorithm Only) Performance

    Yes, a standalone performance evaluation was done. The performance testing described (bioaerosol destruction and particulate filtration) evaluated the device's capabilities without human intervention in its operation for the described tests. The device does operate in "Auto Mode Standard" or "manual mode," but the performance data presented is for the device operating independently to achieve purification.

    7. Type of Ground Truth Used

    The ground truth used for these studies is based on:

    • Direct Measurement of Bioaerosol Reduction: This involves aerosolizing a known quantity of microorganisms and measuring the reduction in their concentration over time within a sealed chamber due to the device's operation.
    • Standardized Filtration Efficiency Testing: This involves testing the filter's ability to remove particles of specific sizes according to the ASHRAE 52.2-2012 standard using specialized equipment.

    8. Sample Size for the Training Set

    This information is not applicable and not provided. The Molekule Air Pro is a physical device that performs air purification. It is not an AI/ML algorithm that requires a "training set" in the conventional sense. The "training" for such a device would involve its engineering design, material selection, and optimization, not data-driven machine learning.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and not provided, as there is no "training set" for an AI/ML algorithm in the context of this device. The development of the device's technology (PECO) and its components (filters, UV LEDs, fan) would be based on scientific principles, engineering design, and iterative testing, not on establishing ground truth for a data-driven model.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1