Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K210947

    Validate with FDA (Live)

    Device Name
    Navident
    Manufacturer
    Date Cleared
    2022-06-27

    (454 days)

    Product Code
    Regulation Number
    872.4120
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Navident is a computerized dental navigational system intended to assist preoperative planning and to guide drilling in a patient jaw during implantation surgery, using pre-acquired CT scan of the jaw. The device is intended for use by a qualified dental surgeon in the treatment of partially or fully edentulous jaws.

    Device Description

    Navident is an image-guided dental navigational system intended to assist with preoperative planning and real-time positioning of drilling tools during implantation surgery. In particular, Navident provides visual, real-time feedback on the location of the working tip of a dental handpiece. It shows the location and direction of the tip relative to a volumetric CT image of the patient's jaw registered to that anatomy, and, when available, relative to a path planned on that image. Navident is comprised of the following parts: The main system is comprised of a cart that carries a stereoscopic video camera and a laptop with pre-installed proprietary software. The Navident system also includes several types of accessories: Jaw motion Tracking Accessories, Dental Handpiece Tracking Accessories, Registration accessories, Calibrator. Navident's four core functions are: Model, Plan, Register, Guide.

    AI/ML Overview

    Here's an analysis of the provided text to extract information about the acceptance criteria and the study proving the device meets them:

    Device: Navident

    Product Code: PLV
    Regulation Number: 21 CFR 872.4120
    Regulation Name: Bone Cutting Instrument And Accessories
    Regulatory Class: Class II

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text focuses on the performance data of the device rather than explicit "acceptance criteria" presented as pass/fail thresholds against specific metrics, except for the accuracy at the drill tip. Instead, it describes various validation and testing activities that collectively demonstrate the device's acceptable performance.

    CategoryAcceptance Criteria / Standard (Implicit or Explicit)Reported Device Performance / Study Finding
    ReprocessingPer AAMI TIR 30: 2011(R) 2016 for cleaning (Proteins < 6.4 µg/cm2, Hemoglobin < 2.2 µg/cm2). Per ISO 17665-1:2006 (R) 2013 for sterilization (acceptable sterility assurance level).Cleaning: Validation demonstrated that the cleaning process included in Navident's labeling can reduce the organic soil load to an acceptable level (meeting the specified protein and hemoglobin thresholds). Sterilization: Validation confirmed components withstand steam sterilization and achieve acceptable sterility using recommended protocols.
    BiocompatibilityPer ISO 10993-1 series or justified non-performance.Patient-contacting components considered tissue contacting for < 24 hours. Tested per ISO 10993-1 or justification provided per FDA Guidance.
    SoftwarePer FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (moderate level of concern).Software documentation submitted according to the guidance.
    Electrical SafetyPer IEC 60601-1:2005 Edition 3.1 & ANSI/AAMI ES60601-1:2005 / 2012 & C1:2009/ 2012 & A2:2010/ 2012.Compliance with these standards for product safety.
    Electromagnetic Compatibility (EMC)Per IEC 60601-1-2:2014 Edition 4.0.Compliance with this standard for EMC.
    Wireless TechnologyPer FDA Guidance "Radio Frequency Wireless Technology in Medical Devices".Evaluation assessment conducted.
    System Accuracy Test (Bench)Explicit: Accuracy at the drill tip ≤1.0mm. Stability and repeatability of patient reference tag coupling to jaw.In all instances (System accuracy test, stability of Jaw Tracker S, C, B, U attachments), the Navident device functioned as intended and the results observed were as expected. The "Accuracy at the drill tip" is explicitly stated as "≤1.0mm" for the proposed Navident, matching the predicate device. Observed performance of new Trace Registration Method: Mean deviation for 136 implants: 0.67 mm (entry), 0.9 mm (apex), 0.55 mm (depth), 2.50° (angular). These values are all below the 1.0mm limit for positional accuracy and represent successful performance for the angular deviation.
    Human Factors/UsabilityMeet identified user specifications; no significant use errors, close calls, or operational difficulties.Summative evaluation demonstrated that the Navident system satisfies identified user specifications. Minor findings were thoroughly analyzed and design improvements made.
    Clinical Literature (Trace Registration Method)Demonstrate acceptable accuracy for the new registration method, comparable to or better than existing methods.Study 1: 136 implants in 59 partially edentulous patients. Mean deviation: 0.67 mm (entry point), 0.9 mm (apex), 0.55 mm (depth), 2.50° (angular deviation). 95th percentile values for trace registration (5-6 teeth) were favorable compared to fiducial marker-based registration (predicate device's method). Study 2: (Pterygoid implants) 31 dynamic navigation cases vs. 32 freehand surgery cases. Mean deviations for dynamic navigation via trace registration: 0.66 mm (coronal), 1.13 mm (apical), 0.67 mm (depth), 2.64° (angular). This was more accurate than freehand surgery.

    2. Sample Sizes and Data Provenance

    • Test Set (Clinical Data from Literature):

      • Study 1 (Trace Registration Accuracy): 136 implants placed in 59 partially edentulous patients.
      • Study 2 (Pterygoid Implants): 63 pterygoid implants in 39 partially edentulous patients (31 with dynamic navigation/trace registration, 32 with freehand surgery).
    • Data Provenance: Clinical studies were conducted outside of the US. The document states that "patient populations, user profiles, use environment, and clinical practices are considered equivalent and applicable to the US population." The studies are retrospective, as they are "published scientific literature" that was collected and analyzed.

    • Human Factors/Usability Study: 15 representative users.

    • Bench Testing: Sample sizes are not explicitly stated for individual bench tests (reprocessing, electrical safety, EMC, full system accuracy bench testing, stability/repeatability of attachments), but it mentions "representative devices" and "representative master devices."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The clinical studies referenced are "published scientific literature" and involve "qualified dental surgeon(s)" performing the procedures. The ground truth for the clinical accuracy studies (Study 1 & 2) appears to be derived from actual implant placement compared to the pre-planned position (likely measured from post-operative imaging).

    • Number of Experts: Not explicitly stated as "number of experts for ground truth establishment." The studies were performed by treating surgeons. Study 2 mentions "the same surgeon" for comparative data, implying single-surgeon data collection for parts of the comparison.
    • Qualifications of Experts: "Qualified dental surgeon(s)." One of the authors in the referenced papers (Stefanelli LV, Mandelaris GA) are typically oral and maxillofacial surgeons or periodontists who are experts in implant surgery.

    4. Adjudication Method for the Test Set

    No explicit adjudication method (e.g., 2+1, 3+1) is described for the ground truth of the clinical studies. The accuracy metrics appear to be derived from quantitative measurements comparing planned vs. actual outcomes, likely without a separate expert adjudication panel.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC study was not explicitly performed in the context of comparing human readers with and without AI assistance, as Navident is a surgical navigation system, not an AI-assisted diagnostic imaging device for human reading. The "comparison" is between different surgical techniques (dynamic navigation with trace registration vs. freehand surgery, or dynamic navigation with new trace registration vs. old fiducial registration) or system accuracy benchmarks, not human interpretation of images. However, the studies demonstrate the clinical effectiveness of the device-assisted procedure.

    • Effect Size for AI vs. Without AI Assistance:
      • Study 1 (Trace Registration vs. Fiducial-based): The 95th percentile values for the trace registration method were "favorable" compared to the fiducial-based registration method (cleared predicate). This implies an improvement, but specific effect sizes are not quantified here beyond the descriptive "favorable." This is a comparison of system performance, not human performance.
      • Study 2 (Dynamic Navigation with Trace Registration vs. Freehand Surgery): Dynamic navigation was "more accurate." The mean deviations for dynamic navigation were significantly lower across all metrics (e.g., 0.66 mm coronal vs. 1.54 mm for freehand; 2.64° angular vs. 12.49° for freehand). This provides a clear quantitative effect size for the improvement in surgical accuracy when using the device compared to freehand.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The "Full system accuracy bench testing" and the clinical accuracy studies (comparing planned vs. actual implant positions) inherently represent the performance of the algorithm and system, as they quantify the accuracy of the guidance provided by the device. While a human surgeon provides the input and executes the drilling, the accuracy metrics are a direct measure of the system's ability to guide to the planned position. The "Accuracy at the drill tip: ≤1.0mm" is a standalone performance metric.

    7. Type of Ground Truth Used

    • Clinical Studies: Ground truth for implant accuracy was based on comparison of the pre-planned position (from CT scan) with the actual post-operative implant position (derived from post-operative imaging). This is a form of outcomes data or objective measurement of surgical execution.
    • Bench Testing: Ground truth for system accuracy was established through controlled measurements conducted on test setups designed to measure the mapping accuracy and stability.

    8. Sample Size for the Training Set

    The document does not specify a separate "training set" or its size. This is a surgical navigation system, not typically a machine learning or AI model trained on a large imaging dataset in the conventional sense. The "training" for the device's development would be more akin to algorithm development and calibration rather than data-driven machine learning, although some components might use statistical models or optimizations. The clinical literature cited pertains to validation of the device's performance, not training data for an AI.

    9. How the Ground Truth for the Training Set was Established

    Given that a specific "training set" for an AI algorithm is not mentioned, the method for establishing its ground truth is also not described. The device's core functionality relies on stereoscopic optical tracking and registration algorithms, whose "ground truth" and performance are verified through precise engineering and bench testing rather than large-scale data labelling for AI training.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1