Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K210884

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-07-27

    (124 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    1Derful™ Zirconia blanks are indicated for use in prosthetic dentistry to create porcelain (ceramic) prostheses (dentures, crowns and bridges) in the anterior applications and not used solely as an implant. 1Derful™ Zirconia blanks are intended to be milled and fully sintered by Dental Professional or Dental Laboratory before use.

    Device Description

    1Derful™ Zirconia are disc shaped dental porcelain zirconia oxide blanks that come in various sizes that are used in custom restorations by the dental laboratory. The dental laboratory will further process the blank by milling the blank based upon the anatomically rendering of the patient's teeth (done at the dental office) through "Computer Aided Drafting/ Computer Aided Machining (CAD/CAM). Once the custom rendered blank is milled the product is fully sintered and colored (if required) and fitted to the patient's teeth as dentures, crowns or bridges. 1Derful™ Zirconia blanks are composed of zirconia ceramics (ZrO2) based on yttria- stabilized tetragonal zirconia (Y-TZP). Shading oxide will also be used to create the need shade in the disc. These three oxide materials are Fe2O3, ErO3, Co3O4 & Er2O3.

    AI/ML Overview

    The provided text describes the non-clinical testing performed for the 1DERFUL™ HS & 1DERFUL™ HT Zirconia blanks. It does not mention any clinical studies, multi-reader multi-case studies, or human-in-the-loop performance. Therefore, information regarding those aspects cannot be extracted.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test/PropertyAcceptance Criteria (Standard Requirement)Reported Device Performance (1DERFUL™ HS & 1DERFUL™ HT)Outcome
    Flexural Strength Test per ISO 6872:2015 Section 7.3Minimum (mean) 800 MPa for class 6 fixed prostheses, lowest value> 1000 MPaPass
    Freedom from extraneous materials per ISO 6872:2015 Section 5.2 active conc. of not more than 1.0 Bq g-1 of Uranium238< 1.0 Bq g-1 of Uranium238< 0.03Pass
    Chemical solubility per ISO 6872:2015 Section 7.6Maximum 100 µg/cm²5.257 µg/cm² (from table on page 5) AND 18.1 µg/cm² (from table on page 8)Pass
    Fracture Toughness per ISO 6872:2015 Section Annex A5.0 MPa m^0.5 (min.) for Class 55.0 MPa m^0.5Pass
    Sintered density (in-house calculated d=m/v)≥ 6.09 g/cm³ (from page 6)6.08 g/cm³ (from table on page 5) AND 6.08 g/cm³ (from table on page 6)Pass
    Grain Size determined per ISO 13356:2008(Not explicitly stated in "Acceptance Criteria" column, but the standard mandates the determination)0.41 µmPass
    Amount of monoclinic phase (X-ray diffraction, ASTM F1873-98)(Not explicitly stated in "Acceptance Criteria" column, but the standard mandates the determination, and the predicate has Tetragonal 98.5%, Monoclinic 1.5% max)ZrO2– Zirconium Oxide Tetragonal P42/mmc 98.5%ZrO2– Baddeleyite Monoclinic P21/a 1.5% (max)Pass
    Linear Thermal Expansion Coefficient per ISO 6872:2015(Not explicitly stated as a minimum/maximum, but the standard requires the test)10.1 µm/m°CPass
    Fe2O3 content(Not explicitly stated in "Acceptance Criteria" column, but is tested as a component with a limit of ≤0.01%)≤0.01Pass

    Note: There are two different "Chemical solubility" values reported for the 1DERFUL™ device. One is 5.257 µg/cm² in the comparison table on page 5, and another is 18.1 µg/cm² in the detailed testing results table on page 8. Both are below the 100 µg/cm² limit.
    Also, the Sintered Density acceptance criteria on page 6 states "≥ 6.09 g cm3", while the reported performance is 6.08 g/cm³ on both page 5 and 6, and it's marked as "Pass". This might indicate a slight discrepancy or that the "≥" means "at least" and 6.08 is considered acceptable based on the standard.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified for individual tests. The text only says "1Derful, Inc performed a series of tests".
    • Data Provenance: Not explicitly stated. The tests were performed by "1Derful, Inc" and are non-clinical, likely internal laboratory testing rather than data from a country of origin or retrospective/prospective studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This was non-clinical laboratory testing, not a study involving experts establishing ground truth for diagnostic accuracy.

    4. Adjudication method for the test set

    Not applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. The document explicitly states: "Clinical tests have not been performed." Therefore, no MRMC study (which typically involves human readers assessing cases) was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a physical material (zirconia blanks) for dental prostheses, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical testing was established by adherence to recognized international standards and guidance documents, such as ISO 6872:2015 and ISO 13356:2008. The performance metrics (e.g., flexural strength, chemical solubility, fracture toughness) were measured and compared against the limits specified in these standards.

    8. The sample size for the training set

    Not applicable. This is a physical device, not an AI/ML model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1