510(k) Data Aggregation

    K Number
    K210856

    Validate with FDA (Live)

    Date Cleared
    2021-07-28

    (127 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Purism Disposable Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

    Device Description

    The Purism Disposable Surgical Masks are manufactured with 3 layers, the inner facing layer and outer facing layer are composed of non-woven fabric, the middle layer is made up of melt-blown fabric. The surgical masks are single use, flat-folded with ear loops. The model of proposed device, ear loops, is held in place over the users' mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are not made with natural rubber latex. The Surgical Masks are sold non-sterile and are intended to be disposable mask.

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for a medical device: "Purism Disposable Surgical Mask". This document does not describe an AI/ML-driven medical device, nor does it detail a study involving AI assistance for human readers or a standalone AI algorithm. It's a standard regulatory submission for a physical medical product (a surgical mask) and focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance and biocompatibility testing.

    Therefore, many of the requested items related to AI/ML device studies (such as MRMC studies, human reader improvement with AI, ground truth establishment for training sets, etc.) are not applicable to this document.

    However, I can extract information related to the acceptance criteria and the study (non-clinical testing) that proves the device meets these criteria, as detailed in the document.

    Here's the breakdown based on the provided text, addressing the applicable points:


    Device Name: Purism Disposable Surgical Mask
    Regulatory Class: Class II
    Product Code: FXX
    Regulation Name: Surgical apparel (21 CFR 878.4040)
    Predicate Device: K182514 Disposable Surgical Face Mask (Xiantao Zhibo Non-woven Products Co., Ltd)

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents non-clinical performance testing and biocompatibility testing.

    Non-Clinical Performance Testing:

    Performance Testing StandardPurposeAcceptance CriteriaReported Device Performance
    Fluid ResistanceTo evaluate the resistance of medical face masks to penetration by the impact of a small volume (~2 mL) of a high-velocity stream of synthetic blood.At least 29 out of 32 specimens per lot show passing results at 160 mmHg94 out of 96 pass at 160mmHg
    Particulate Filtration Efficiency (PFE) ASTM F2299To determine the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres.≥ 98% Average PFE for all samples tested96/96 Samples Pass at Average of 99.90%
    Bacterial Filtration Efficiency (BFE) ASTM F2101To measure the bacterial filtration efficiency (BFE) of medical face mask materials, employing a ratio of the upstream bacterial challenge to downstream residual concentration to determine filtration efficiency of medical face mask materials.≥ 98% Average BFE for all samples tested96/96 Samples Pass at Average of 99.88%
    Differential Pressure (Delta P) EN 14683To determine the pressure required to breathe through the final manufactured face mask.Samples must be < 6.0 mm H2O/cm²96/96 Samples Pass at Average of 4.1mmH2O/cm²
    Flammability 16 CFR 1610The purpose of this standard is to reduce danger of injury and loss of life by providing, on a national basis, standard methods of testing and rating the flammability of textiles and textile products for clothing use, thereby prohibiting the use of any dangerously flammable clothing textiles.All samples must be Class I96/96 Samples Pass for Class 1

    Biocompatibility Testing:

    Performance Testing (Biocompatibility)PurposeAcceptance CriteriaReported Device Performance
    Cytotoxicity ISO 10993-5Under the research conditions, determine whether the target device extract is cytotoxic.The sample is non-cytotoxic.Under the conditions of the study, the device is noncytotoxic.
    Irritation ISO 10993-10Under the research conditions, determine whether the non-polar and polar extracts of the target device are sensitive.The sample is non-irritating.Under the conditions of the study, the device is nonirritating.
    Sensitization ISO 10993-10Under the research conditions, determine whether the non-polar and polar extracts of the target device are irritating.The sample is non-sensitizing.Under the conditions of the study, the device is non-sensitizing.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set (Non-Clinical):
      • For Fluid Resistance: 96 specimens were tested (indicated by "94 out of 96 pass" and acceptance criterion "29 out of 32 specimens per lot"; 96 suggests 3 lots of 32, or a total of 96 samples tested to ensure sufficient statistical power).
      • For Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, and Flammability: 96 samples were tested for each.
    • Data Provenance: The testing was conducted by or on behalf of Chuzhou Daddy's Choice Science and Technology Co., Ltd., based in Chuzhou City, Anhui, China. The data is from non-clinical laboratory testing specifically for this 510(k) submission. It is by nature "prospective" for the purpose of this submission, as it was generated to demonstrate compliance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not an AI/ML device requiring expert interpretation of images or other data for ground truth. The "ground truth" for these tests is the quantitative measurement results from standardized laboratory procedures.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not an AI/ML device requiring human adjudication of data. Compliance is determined by objective laboratory measurements against predefined acceptance criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device. No MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML device. No standalone algorithm performance was assessed.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For non-clinical performance and biocompatibility testing, the "ground truth" is defined by the standardized test methods and their objective quantitative results. For example:

    • Fluid Resistance: Visual observation of synthetic blood penetration after impact.
    • Filtration Efficiencies: Quantitative measurement of particle/bacteria concentration reduction.
    • Differential Pressure: Measurement of pressure differential across the mask material.
    • Flammability: Classification based on burn characteristics.
    • Biocompatibility: In-vitro (cytotoxicity) or in-vivo (irritation, sensitization) biological responses observed in laboratory models, interpreted against established biological safety evaluation criteria (e.g., ISO 10993 series).

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.


    In summary, the provided document describes a traditional medical device (surgical mask) and its non-clinical testing for 510(k) clearance, which relies on standardized physical and chemical performance tests rather than AI/ML methodologies.