    K Number
    K210333

    Validate with FDA (Live)

    Device Name
    DiaPaste
    Date Cleared
    2021-03-25

    (48 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Aqueous ointment material that temporarily fills the root canal for the following indications: Apexification Temporary root filling Root canal filling for primary teeth Vital pulpotomy Temporary pulp capping

    Device Description

    Diapaste is a calcium hydroxide paste with barium sulfate, used as a temporary root canal filling material.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called DiaPaste, a root canal filling resin. It focuses on demonstrating substantial equivalence to a predicate device rather than detailing extensive clinical studies with acceptance criteria for device performance in diagnosing or predicting conditions.

    Therefore, I cannot extract the information required for a table of acceptance criteria and reported device performance related to diagnostic or predictive capabilities, nor can I provide information on sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, training set details, or how ground truth was established for a diagnostic AI.

    However, I can provide the non-clinical performance data and acceptance criteria as presented in the document for the physical properties of the device:

    1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical Physical Properties)

    TestAcceptance Criteria (Requirement)Reported Device PerformanceResult
    FlowabilityEach disc to have a diameter of not less than 17mm18mm (mean value)Pass
    Film thicknessNo more than 50µm23µm (mean value)Pass
    RadiopacityEquivalent to not less than 3mm of AluminumNot less than 3mm of Aluminum (approx. 4.1mm)Pass

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The performance data listed (Table 1) pertains to non-clinical physical and material property tests of the device itself, not to a test set of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided. The ground truth for physical properties (flowability, film thickness, radiopacity) is established by standardized laboratory measurements rather than expert human interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided. Adjudication methods are typically relevant for human interpretation of medical images or clinical data, not for material property testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and not provided. This document describes a paste used for root canal filling, not an AI-powered diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable and not provided. This document describes a physical medical device, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the physical property tests, the "ground truth" is derived from established laboratory measurement standards and protocols (e.g., ISO 6876). For biocompatibility, the ground truth is established through standardized biological evaluation tests (e.g., ISO 10993 series), confirming that the material does not cause adverse biological responses.

    8. The sample size for the training set

    This information is not applicable and not provided. This document describes a physical medical device, not an algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided. As noted above, this is a physical device, not an algorithm with a training set.