510(k) Data Aggregation

    K Number
    K210300

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-04-01

    (57 days)

    Product Code
    Regulation Number
    886.5928
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VueSonic One Contact Lens Cleaning System is a cleaning system for Soft and Hybrid contact lenses by mimicking digital rubbing with 3-D sonic Vibration Technology.

    The VueSonic Advance 2 Contact Lens Cleaning System is a cleaning system for Rigid contact lenses by mimicking digital rubbing with 3-D sonic Vibration Technology.

    Device Description

    The VueSonic One Contact Lens Cleaning System and VueSonic Advance 2 Contact Lens Cleaning Systems are cleaning systems for contact lenses. Both devices consist of a device body, a contact lens holder, which houses the cleaning cushions and the damping heads. The contact lens holder, cleaning cushions and damping heads are made of soft medical silicone, which will not damage the lens. The systems are designed to mimic human hand rubbing. This is done by the 3D multi-frequency modulated stereo power source and corresponding intelligent control circuit, which make a uniform three-dimensional relative movement between the damping heads and cleaning cushion.

    The proposed device consists of two versions:

    • For Rigid Lenses, VueSonic Advance 2 Contact Lens Cleaning System; and
    • For Soft and Hybrid Lenses, VueSonic One Contact Lens Cleaning System.
    AI/ML Overview

    The provided document describes the VueSonic One Contact Lens Cleaning System and VueSonic Advance 2 Contact Lens Cleaning System. It does not contain a study explicitly proving the device meets acceptance criteria. Instead, it relies on a battery of non-clinical tests and a comparison of technological characteristics with predicate devices to establish substantial equivalence.

    Therefore, many of the requested details about a specific study, acceptance criteria for performance, and ground truth establishment cannot be extracted directly from this document.

    However, I can provide the information that is present:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly define "acceptance criteria" for device performance in the form of specific thresholds or metrics for cleaning effectiveness. Instead, it lists the non-clinical tests performed, implying that satisfactory results from these tests constitute meeting the requirements for substantial equivalence. The "reported device performance" in terms of cleaning effectiveness is described through the successful completion of a "Protein and Lipid Removal Test."

    Test CategorySpecific TestRelevant Standard/GuidanceImplied Acceptance: Satisfactory Result
    Cleaning PerformanceProtein and Lipid Removal TestCitation (specific not provided)The test results demonstrate adequate protein and lipid removal, supporting the device's cleaning function (implied by "test results show that the devices are substantially equivalent").
    BiocompatibilityOcular IrritationISO 10993-10, 12Device materials are not ocular irritants.
    In Vitro Cytotoxicity TestISO 10993-5Device materials are not cytotoxic.
    SensitizationISO 10993-10Device materials do not cause sensitization.
    Systemic ToxicityISO 10993-11, 12Device materials do not cause systemic toxicity.
    Electrical Safety and EMCGeneral Requirements electromagnetic compatibilityIEC 60601-1-2The device meets electromagnetic compatibility requirements.
    General Requirements electrical safetyIEC 60601-1The device meets general electrical safety requirements.
    General Requirements home healthcare environmentIEC 60601-1-11The device meets safety requirements for home healthcare use.
    UsabilityHuman Factor Engineering and Usability Engineering Evaluation ReportIEC62366-1:2015The device is user-friendly and safe for its intended use, based on human factors principles.
    Material SafetyTotal Extractive ResiduesFDA 21 CFR 177.2600The device's materials do not release unacceptable levels of extractive residues.
    SoftwareSoftware ValidationFDA GuidanceThe software components of the device are validated and function as intended.
    Lens CompatibilityLens Compatibility TestISO 18369-2:2017The device is compatible with the specified types of contact lenses (soft, hybrid, rigid) without causing damage.

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample size for the "Protein and Lipid Removal Test" or any other non-clinical test. It also does not explicitly state the country of origin of the data or whether the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The document describes non-clinical tests, which typically rely on standardized protocols and laboratory measurements rather than expert consensus for ground truth.

    4. Adjudication method for the test set

    This information is not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned. The device is a contact lens cleaning system, not an AI-powered diagnostic or interpretive tool that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not directly applicable to a contact lens cleaning device. The "algorithm" in this context refers to the device's mechanical cleaning mechanism (3D sonic vibration technology), which operates in a "standalone" manner in that it performs its cleaning function without human intervention during the cleaning cycle. However, the overall process requires human interaction for placement of lenses and solution.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the "Protein and Lipid Removal Test," the ground truth would likely be established through quantitative laboratory measurements of protein and lipid levels on the contact lenses before and after cleaning, using standardized analytical methods. For other non-clinical tests (e.g., biocompatibility), the ground truth is based on the adherence to specified ISO standards and the absence of adverse biological reactions or safety failures.

    8. The sample size for the training set

    This information is not applicable and not provided. The device is a mechanical cleaning system and does not employ a machine learning model that would require a training set in the conventional sense.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided, as there is no mention of a machine learning model or a training set.