510(k) Data Aggregation

    K Number
    K210239

    Validate with FDA (Live)

    Date Cleared
    2021-02-24

    (26 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Citrespline™ and Citrelock™ ACL Implants are intended for soft tissue reattachment, i.e. fixation of ligament and tendon graft tissue in surgeries of the shoulder, elbow, foot/ankle, knee, and hand/wrist. More specifically:

    Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tendon Reattachment, Acromion-Clavicular Separation Repair, Deltoid Repair, Capsulolabral Reconstruction.

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Achilles Tendon reconstruction, tendon transfers in the foot and ankle.

    Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis.

    Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.

    Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction. Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty). Carpal Ligament Reconstructions and repairs, tendon transfer in the hand/wrist.

    Device Description

    The CITRESPLINE™ and CITRELOCK™ ACL Implants are soft tissue fixation devices ranging in diameter from 7 to 12 mm in 1 mm increments, and lengths of 23, 28, and 33 mm. There are two designs for the system: 1) a "twist-ribbed" cannulated, self-locking device (trade names CITRELOCK ACL and CITRELOCK) and 2) a "straight spline" style device (trade name CITRESPLINE). Both designs include a cannula through the central axis to allow for alignment into the surgically prepared site. Like the predicate devices, surgical placement of the implants in bone is facilitated with a reusable set of instruments. The CITRESPLINE and CITRELOCK ACL Implants are intended to attach soft tissue to bone for procedures in the foot/ankle, knee, shoulder, elbow, and hand/wrist.

    The CITRESPLINE and CITRELOCK ACL Implants are made from CITREGEN biomaterial, which is a homogeneous biocomposite comprised of 60 wt. -% unsintered hydroxyapatite (HA) and 40 wt.-% polyester. CTREGEN's polymer component is a citrate-based network of completely amorphous polymer chains crosslinked together to form an elastomeric material. As water penetrates the subject device, surface erosion of the polymer phase occurs through hydrolysis of the ester bonds located between the monomers and at the crosslink sites.

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter and supporting documentation for two medical devices, the Citrespline™ and Citrelock™ ACL Implants. It focuses on demonstrating substantial equivalence to a predicate device, not on proving device performance through a clinical study with acceptance criteria in the manner typical for AI/ML device submissions.

    Therefore, many of the requested details, such as sample sizes for test and training sets, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and how ground truth was established for training and testing, are not applicable to this document. This document describes the mechanical testing and material properties of the implant devices, not an AI/ML algorithm.

    However, I can extract information related to the performance data and "acceptance criteria" as presented within the context of mechanical and material testing for these implant devices, and present it in a similar structured format.

    Here's the closest interpretation of your request based on the provided text:


    Description of Acceptance Criteria and Device Performance (for Implant Devices)

    The device in question, the Citrespline™ and Citrelock™ ACL Implants, are soft tissue fixation devices intended for surgical reattachment of ligaments and tendons. The documentation focuses on demonstrating their substantial equivalence to a legally marketed predicate device (K200725) through material composition, technological characteristics, and mechanical performance testing, specifically "pull-out" strength.

    The "acceptance criteria" are implied by the phrase "possess similar or significantly higher pull out strengths relative to the primary predicate devices." The study described is a benchtop mechanical testing study, not a clinical study involving human subjects or an AI/ML algorithm.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Similar or significantly higher pull-out strengths relative to the primary predicate devices (K200725)The worst-case subject devices (e.g., smallest diameter, shortest length) demonstrated similar or significantly higher pull-out strengths compared to the primary predicate devices. This was attributed to the larger diameter and longer engagement length of the subject devices.
    Maintain satisfactory pull-out strength over time (aging)Pull-out testing was performed at time-zero and after soaking in 37°C PBS through 26 weeks. Results supported the conclusion of substantial equivalence.
    Maintain satisfactory pull-out strength after simulated physiological loadingPull-out testing was performed after 5,000 cycles of simulated physiological loading. Results supported the conclusion of substantial equivalence.
    SterilityValidations for sterility of implants were adopted from previously performed activities and confirmed.
    Packaging and Shelf LifePackaging and shelf life tests using real and accelerated time aging were performed with passing results. The labeled shelf-life is one year based on analyzed stability data.
    Endotoxin LimitsBacterial endotoxin testing (BET) showed CITREGEN test articles to meet the set endotoxin limits (<20 EUs / surgical procedure).

    2. Sample Size and Data Provenance

    • Sample Size for Test Set: The document refers to testing "worst case subject devices (e.g. smallest diameter, shortest length)" but does not specify the numerical sample size for the mechanical testing.
    • Data Provenance: The testing appears to be benchtop laboratory testing, conducted as part of the verification and validation (V&V) activities for regulatory submission. The country of origin for the data is not explicitly stated but would presumably be at the manufacturer's testing facilities (USA, as the manufacturer is Acuitive Technologies, Inc. based in Allendale, NJ). This is prospective data collection for regulatory submission.

    3. Number of Experts and Qualifications for Ground Truth

    • Not Applicable. This document describes mechanical device testing for implants, not an AI/ML algorithm that requires expert labeling for ground truth. The "ground truth" here is the measured mechanical strength and material properties from laboratory tests.

    4. Adjudication Method for the Test Set

    • Not Applicable. Adjudication is relevant for human-labeled data, not for direct mechanical measurements.

    5. MRMC Comparative Effectiveness Study

    • Not Applicable. This is a mechanical device, not an AI/ML system requiring human-in-the-loop performance measurement.

    6. Standalone Performance (Algorithm Only)

    • This concept is Not Applicable in the context of a physical implant device. The "performance" is the mechanical integrity and material properties of the device itself.

    7. Type of Ground Truth Used

    • The "ground truth" for the device's performance is established through direct measurement of physical properties (e.g., pull-out strength) under controlled laboratory conditions, material composition analysis (CITREGEN biomaterial), and other standard engineering and biocompatibility tests (sterility, endotoxin, shelf-life).

    8. Sample Size for the Training Set

    • Not Applicable. This is a mechanical device, not an AI/ML algorithm that requires a training set. The "design" evolution of the device would be based on engineering principles and previous device performance, not a "training set" in the AI sense.

    9. How the Ground Truth for the Training Set was Established

    • Not Applicable. See point 8.