Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K210208

    Validate with FDA (Live)

    Device Name
    Titan Condoms
    Manufacturer
    Date Cleared
    2022-02-17

    (387 days)

    Product Code
    Regulation Number
    884.5300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Titan Condoms are indicated for use for contraceptive and prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).

    Device Description

    Titan Condoms are made of a natural latex sheath which completely covers the penis with a closely fitted membrane. The condom is shaped with a reservoir end and a cylinder shape. Titan condoms are available in a range of sizes, with different length and width combinations. Surface textures include smooth, dotted, and ribbed. Titan Condoms are provided prelubricated with a silicone-based lubricant.

    AI/ML Overview

    The provided text describes the acceptance criteria and a clinical study conducted for Titan Condoms, but it does not pertain to an AI/ML device. Therefore, many of the requested categories related to AI/ML device studies (such as sample sizes for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, and how training ground truth was established) are not applicable.

    Below is the information specifically related to the acceptance criteria and the study as presented in the document for the physical device, Titan Condoms.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Clinical Failure Rate < 5%64 mm condom: 0% clinical failure rate69 mm condom: 3.1% clinical failure rate
    BiocompatibilityAll tested subject devices were biocompatible.
    Compliance with ISO 4074:2015All Titan Condom versions met all requirements.
    Shelf-life5 years

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated in terms of number of condoms or participants for the clinical study. It mentions that "Each couple was to use three condoms."
    • Data Provenance: Prospective clinical study. The country of origin of the data is not specified.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable, as this is a clinical study for a physical device, not an AI/ML device requiring expert ground truth for image or data interpretation. The ground truth (clinical failure) was observed directly during use.

    4. Adjudication Method for the Test Set

    Not applicable, as this is a clinical study for a physical device, not an AI/ML device requiring adjudication for result interpretation. The clinical endpoints were clinical condom slippage and breakage during use, participant's satisfaction, and genitourinary adverse events, reported via an online individual condom use case report form.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable, as this is a clinical study for a physical device, not an AI/ML device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable, as this is a clinical study for a physical device, not an AI/ML algorithm.

    7. Type of Ground Truth Used

    The ground truth for the clinical study was based on observed clinical outcomes: clinical condom slippage, breakage during use, participant satisfaction, and genitourinary adverse events, as reported by the users themselves.

    8. Sample Size for the Training Set

    Not applicable, as this is a clinical study for a physical device. There is no AI/ML model to train.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI/ML model.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1