510(k) Data Aggregation

    K Number
    K210142

    Validate with FDA (Live)

    Device Name
    Fiber Dust
    Manufacturer
    Date Cleared
    2021-03-11

    (51 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fiber Dust is intended for incision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.

    Urology

    • · Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH]
    • · Laser Resection of the Prostrate (LRP)
    • · Laser Enucleation of the Prostate (LEP)
    • · Laser Ablation of the Prostate (LAP)
    • · Transurethral Incision of the Prostate (TUIP)
    • Condylomas
    • · Urethral strictures
    • Lesions of external genitalia
    • · Bladder neck incisions (BNI)
    • · Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors
    • · Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi
    • · Treatment of distal impacted fragments remaining in the ureters following lithotripsy

    Lithotripsy and Percutaneous Urinary Lithotripsy Indications

    • · Endoscopic fragmentation of ureteral, bladder and renal calculi including cystine, calcium oxalate,
    • monohydrate and calcium oxalate dehydrate stones
    • Endoscopic fragmentation of renal calculi
    • · Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed

    Gastroenterology
    Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including:

    • · Appendectomy
    • · Angiodysplasia
    • · Polyps
    • · Colorectal cancer
    • · Biopsy
    • · Telangiectasias
    • Gall Bladder calculi
    • · Telangiectasias of the Osler-Weber-Renu disease
    • Biliary/Bile duct calculi
    • Vascular Malformation
    • Ulcers
    • · Gastritis
    • · Gastric ulcers
    • · Esophagitis
    • · Duodenal ulcers
    • · Esophageal ulcers
    • Non Bleeding Ulcers
    • · Varices
    • Pancreatitis
    • Colitis
    • Haemorrhoids
    • Mallory-Weiss tear
    • ·Cholecystectomy
    • · Gastric Erosions
    • ·Benign and Malignant Neoplasm

    Gynecology
    Open and laparoscopic gynecological surgery (incision, excision, ablation, vaporization, coagulation and hemostasis) of soft tissue.

    Device Description

    The Fiber Dust laser device includes a diode laser that emits a wavelength of 1940 nm with a laser output power up to 60W. Laser radiation is delivered to the patient via a quartz optical fiber having a diameter up to 1000 μm.

    The main subsystems of the device are the diode laser, the power electronics, the optical delivery system, the control electronics, and the cooling system. Software controls the device functions and allows the user select device settings. Laser emission is triggered by a footswitch.

    AI/ML Overview

    The provided text is a 510(k) summary for the "Fiber Dust" laser device, which is a medical device. This document focuses on demonstrating substantial equivalence to a predicate device through performance testing against recognized consensus standards. It does not contain information about the use of AI/ML or a study that establishes acceptance criteria for an AI-powered device's performance metrics like sensitivity, specificity, or AUC, as typically seen in submissions for AI/ML medical devices.

    Therefore, the requested information cannot be fully extracted based on the provided text. Specifically:

    1. A table of acceptance criteria and the reported device performance: This information is not provided in a format relevant to AI/ML device performance (e.g., no metrics like sensitivity, specificity, or accuracy based on a clinical ground truth). The document states that "non-clinical performance standards testing support that the device can be used safely and effectively," but these refer to engineering and safety standards, not AI performance metrics.

    2. Sample size used for the test set and the data provenance: Not applicable, as no AI/ML test set is described. Performance testing primarily refers to engineering and safety standards compliance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.

    4. Adjudication method for the test set: Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. The device is a surgical laser, not an AI-assisted diagnostic or decision-support tool for human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.

    7. The type of ground truth used: For the purposes of this submission, the ground truth relates to the device meeting engineering and safety standards (e.g., electrical safety, electromagnetic compatibility, laser safety). There is no "ground truth" as it would be defined for an AI/ML diagnostic or predictive model.

    8. The sample size for the training set: Not applicable, as there is no mention of an AI model being trained.

    9. How the ground truth for the training set was established: Not applicable.

    Summary of available information related to "acceptance criteria" and "study":

    The acceptance criteria for the Fiber Dust device are based on compliance with recognized consensus standards for medical electrical equipment and laser safety.

    • IEC 60601-1:2012, ed 3.1 (General Requirements for Basic Safety And Essential Performance)
    • IEC 60601-1-2:2014, ed 4 (Electromagnetic Disturbances - Requirements And Tests)
    • IEC 60601-2-22: 2012, ed 3.1 (Particular Requirements For Basic Safety And Essential Performance Of Surgical, Cosmetic, Therapeutic And Diagnostic Laser Equipment)
    • IEC 60825-1:2014, ed 3.0 (Safety of laser products – Part 1: Equipment classification and requirements)
    • IEC60601-1-6:2013, ed 3.1 (Usability)

    Additionally, "Software Verification and Validation Testing" was conducted following FDA guidance for software in medical devices.

    The "study" or testing that proves the device meets these criteria refers to the performance testing against these standards, the results of which "support that the device can be used safely and effectively." These are non-clinical tests typically conducted in a laboratory or engineering environment. The document states: "The results of the non-clinical performance standards testing support that the device can be used safely and effectively."

    In conclusion, the provided text describes the clearance of a traditional medical device (a surgical laser) based on engineering and safety standards compliance, not an AI/ML-driven device with performance metrics based on clinical data or expert ground truth.