510(k) Data Aggregation

    K Number
    K203815

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-06-09

    (162 days)

    Product Code
    Regulation Number
    888.3350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ELEOSTM Limb Salvage System Hip Components:

    Indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions :

    1. Non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusion acetabuli. and painful hip dysplasia;

    2. Inflammatory degenerative joint disease such as rheumatoid arthritis;

    3. Correction of functional deformity

    4. Revision procedures where other treatments or devices have failed; and,

    5. Treatment of fractures that are unmanageable using other techniques.

    Limb salvage system is also indicated for procedures where radical resection and replacement of the proximal, distal and or total femur is required with the following conditions:

    1. Patients suffering from severe arthropathy of the hip that does not respond to any conservative therapy or better alternative surgical treatment:

    2. Surgical intervention for severe trauma, revision hip arthroplasties, and/or Oncology indications.

      1. Metastatic diseases
        ELEOSTM Limb Salvage System Knee Components:

    Indicated for cemented use in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with the following conditions:

    1. Non-inflammatory degenerative joint disease including osteoarthritis, or avascular necrosis;

    2. Inflammatory degenerative joint disease including rheumatoid arthritis;

    3. Correction of functional deformity;

    4. Revision procedures where other treatments or devices have failed; and

    5. Treatment of fractures that are unmanageable using other techniques.

    Limb salvage system is also indicated for procedures where radical resection and replacement of the proximal tibia is required with the following conditions:

    1. Patients suffering from severe arthropathy of the knee that does not respond to any conservative therapy or better alternative surgical treatment;

    2. Surgical intervention for severe trauma, revision knee arthroplasties, and/or Oncology indications.

    3. Metastatic diseases

    Device Description

    The Onkos Surgical ELEOS™ Limb Salvage System consists of components that are used in the reconstruction of the lower limb. The reconstruction are proximal femur, distal femur, total femur, proximal tibia, and hinged knee. The ELEOS Limb Salvage components are femoral head, proximal femur, mid-section, stem, distal hinge femur, tibial hinge assembly, axial pin, tibial poly spacer, tibial sleeve, male mid-section, resurfacing hinge femur, and proximal tibia, optional components include modular collars, patella, stem extensions, tibial wedges and augments. Instrumentation is provided non-sterile in surgical trays which are to be re-processed per the validated instructions stated below.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (ELEOS Limb Salvage System with BioGrip) seeking FDA clearance. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel performance against specific acceptance criteria through a clinical study.

    Therefore, the document does not contain the acceptance criteria or a study designed to prove the device meets those criteria in the way you've outlined for a clinical performance study of, for example, an AI algorithm.

    Instead, the document details performance data provided in support of substantial equivalence to a predicate device. This data focuses on mechanical and biocompatibility testing to ensure the new device (a line extension with modular collars and porous technology) is as safe and effective as the existing, cleared predicate device.

    Here's a breakdown of what is available based on your request, with an explanation of why other points are not applicable in this context:


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) submission based on substantial equivalence, there isn't a direct "acceptance criteria" table with specific performance metrics (like sensitivity/specificity for an AI algorithm) that the device must meet to prove its clinical utility. Instead, the performance data provided demonstrates that the device performs similarly or better mechanically and is biocompatible compared to predicate devices, adhering to recognized standards.

    Biocompatibility Testing:

    Acceptance Criteria (General Goal)Reported Device Performance
    Biocompatible for permanent contactPassed tests per ISO 10993 series and ANSI/AAMI ST72 for bacterial endotoxins.
    Endotoxin levels within limitsMet 20 endotoxin units (EU)/device limit for implanted medical devices.

    Mechanical Testing:

    Acceptance Criteria (General Goal)Reported Device Performance
    Equivalent mechanical properties to predicateCompleted testing according to ASTM F1044, F1147, F1160, F1854, F1978, ISO 7206-4, and ISO 13179-1:2014. Results support substantial equivalence to predicate.
    Adequate strength and durabilityImplicitly demonstrated by passing the listed ASTM and ISO standards for orthopedic implants. The specific numerical results (e.g., shear strength values) are not provided in this summary.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not applicable in the context of this 510(k) submission. The "test set" in this case refers to the physical device components subjected to laboratory testing. The document does not describe a clinical study with patients or data from medical records.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. Ground truth, in the sense of expert consensus on clinical diagnoses or outcomes, is not established for the mechanical and biocompatibility tests performed. These tests rely on standardized protocols and measurements, not expert interpretation of clinical data.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies to resolve discrepancies in expert interpretations of data. The performance analyses presented here are laboratory-based.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This information is not applicable. This submission is for a physical orthopedic implant system, not an AI algorithm. No MRMC study or AI assistance is mentioned.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable. This submission is for a physical orthopedic implant system, not an algorithm.

    7. The Type of Ground Truth Used

    For biocompatibility, the "ground truth" is defined by the toxicological characteristics and cellular responses measured against established ISO standards.
    For mechanical testing, the "ground truth" is defined by the physical limits and properties established in the ASTM and ISO standards for orthopedic implants.

    8. The Sample Size for the Training Set

    This information is not applicable. There is no "training set" in the context of this device's development as it is not an AI/machine learning device. The design and testing are based on engineering principles and material science.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reasons as point 8.