510(k) Data Aggregation

    K Number
    K203801

    Validate with FDA (Live)

    Date Cleared
    2021-04-06

    (99 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Surgical Face Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

    Device Description

    The proposed device(s) are Blue color, and Flat Pleated type mask, utilizing Tie-On (model: JB-DM04) or Ear Loops (model: JB-DM03) way for wearing, and they all has Nose Piece design for fitting the facemask around the nose.

    The proposed device(s) are manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter.

    The model JB-DM04 of proposed device, tie-on, is held in place over the user's mouth and nose by four ties welded to the facemask. The tie is made of spun-bond polypropylene.

    The model JB-DM03 of proposed device, ear loops, is held in place over the user's mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are made of Nylon and spandex.

    The nose chip contained in the proposed device(s) is in the layers of facemask to allow the user to fit the facemask around their nose, which is made of PE and iron.

    The proposed device(s) are sold non-sterile and are intended to be single use, disposable device.

    AI/ML Overview

    The provided document is a 510(k) Summary for a Disposable Surgical Face Mask, which is a medical device. The information requested pertains to the acceptance criteria and study proving a digital health or AI/ML-driven device's performance. This document does not describe such a device. Instead, it outlines the non-clinical performance testing for a physical medical device (a face mask).

    Therefore, I cannot extract the information required by your prompt, such as sample sizes for test/training sets, data provenance, expert ground truth adjudication, MRMC studies, standalone algorithm performance, or how ground truth was established for training, because these concepts are not applicable to the evaluation of a physical surgical face mask as presented in this document.

    The document focuses on:

    • Acceptance Criteria for the Face Mask: These are based on established standards for surgical masks, such as ASTM F2100 performance levels (Fluid Resistance, Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, Flammability) and ISO 10993 for biocompatibility.
    • Study Proving Device Meets Criteria: This involves non-clinical laboratory tests performed directly on the mask material or a sample of masks.

    Here's an analysis of what is present in the document, framed against your request:

    1. A table of acceptance criteria and the reported device performance:

    Test MethodologyAcceptance Criteria (level 2 & 3)Proposed Device 1 (model: JB-DM04) Reported PerformanceProposed Device 2 (model: JB-DM03) Reported PerformanceResult
    Fluid Resistance (ASTM F1862)29 out of 32 pass at 160 mmHg for Level 3; 29 out of 32 pass at 120 mmHg for Level 232 out of 32 pass at 160 mmHg32 out of 32 pass at 120 mmHgPASS
    Particulate Filtration Efficiency (ASTM F2299)≥ 98%99.9%99.9%PASS
    Bacterial Filtration Efficiency (ASTM F2101)≥ 98%>99.9%>99.9%PASS
    Differential Pressure (Delta P) (EN 14683 Annex C)< 6.0 mmH2O/cm²2.21 mmH2O/cm²2.53 mmH2O/cm²PASS
    Flammability (16 CFR 1610)Class 1Class 1Class 1PASS

    Biocompatibility Testing:

    ItemAcceptance CriteriaReported Performance (Proposed Device)Result
    CytotoxicityNo potential cytotoxicityUnder the conditions of the study, the subject device extract did not show potential toxicity to L929 cells.PASS
    IrritationNon-IrritatingUnder the conditions of the study, the device is non-irritating.PASS
    SensitizationNon-SensitizingUnder the conditions of the study, the device is non-sensitizing.PASS

    2. Sample sized used for the test set and the data provenance:
    * Sample Size (Test Set): For Fluid Resistance, 32 masks were tested. For other tests like PFE, BFE, Differential Pressure, Flammability, and Biocompatibility, specific sample sizes are not explicitly stated but are implicit in the methodology of these lab tests (e.g., usually multiple samples are tested).
    * Data Provenance: The tests were non-clinical, laboratory-based. The location where the tests were performed is not explicitly stated, but the manufacturer is Anhui Jiabao Protective Equipments Co., Ltd in China. The submission correspondent is in Shanghai, China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    * Not applicable. Ground truth for a physical medical device like a face mask is established by validated laboratory test methods and established standards (e.g., ASTM, ISO), not by human expert consensus or adjudication in this context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    * Not applicable. This is for a physical device, evaluated by objective lab tests, not human interpretation/adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    * Not applicable. This is for a physical device, not an AI/ML system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    * Not applicable. This is for a physical device, not an algorithm.

    7. The type of ground truth used:
    * Established by Objective Laboratory Test Standards: The "ground truth" for the performance claims of this medical device is the adherence to the specified performance criteria as measured by internationally recognized and validated laboratory test methods (e.g., ASTM F1862, ASTM F2299, ASTM F2101, EN 14683, 16 CFR 1610, ISO 10993 series).

    8. The sample size for the training set:
    * Not applicable as this is not an AI/ML device requiring a training set.

    9. How the ground truth for the training set was established:
    * Not applicable as this is not an AI/ML device requiring a training set.

    In summary, the provided document describes the safety and performance evaluation of a physical medical device (surgical face mask) through non-clinical laboratory testing against established industry standards, rather than the evaluation of a digital health or AI/ML-driven device.