Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K203474

    Validate with FDA (Live)

    Date Cleared
    2021-01-28

    (64 days)

    Product Code
    Regulation Number
    888.3030
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RibFix Advantage System is indicated for the fixation, and fusion of rib fractures and osteotomies of normal and osteoporotic bone.

    Device Description

    The RibFix Advantage System consists of bridges (with locking posts) and locking caps for the thoracoscopic fixation and stabilization of ribs. These implants are manufactured from commercially pure titanium (per ASTM F67) and titanium alloys (Ti-6A1-7Nb per ASTM F1295 and Ti-6Al-4V per ASTM F136 or ASTM F1472). The system is comprised of the following implants:

    • . Titanium alloy RibFix Advantage Bridge with threaded locking posts
    • . Commercially pure titanium locking caps
    • . Commercially pure titanium washers (optional)
      When fully assembled, the bridge is placed on the underside of the rib, the threaded locking posts extends through pre-drilled holes in the rib, and the locking caps are fixed to the locking post on the anterior side of the rib. Washers may be placed on the posts prior to assembling with locking caps. The combined threaded locking post/locking cap provides for fixation of the bridge. The devices are sold non-sterile and intended to be sterilized by the user prior to implantation. The system also includes reusable instrumentation to aid in implantation of the devices.
    AI/ML Overview

    This document (K203474) describes a 510(k) premarket notification for the "Biomet Microfixation RibFix Advantage System." This system is for the fixation and fusion of rib fractures and osteotomies. The FDA determined the device is substantially equivalent to legally marketed predicate devices.

    Acceptance Criteria and Device Performance:

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (K183317, AdvantageRib System) and provides non-clinical performance data related to MR safety. It does not present acceptance criteria in the typical format of accuracy, sensitivity, or specificity commonly used for AI/software devices. Instead, the acceptance criteria are implicitly that the device performs equivalently or better than the predicate device in relevant non-clinical tests.

    The document states: "Testing identified in this summary has all passed acceptance criteria established by the predicate device where applicable."

    Here's a table based on the provided information, interpreting "acceptance criteria" as meeting the standards set for safe operation in an MRI environment:

    Acceptance Criteria CategorySpecific Standard/RequirementReported Device Performance
    MR SafetyASTM F2052-15: Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance EnvironmentAll testing passed
    (implicitly, the device demonstrated acceptable displacement force as per the standard)
    ASTM F2213-17: Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance EnvironmentAll testing passed (implicitly, the device demonstrated acceptable torque as per the standard)
    ASTM F2182-11a: Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance ImagingAll testing passed (implicitly, the device demonstrated acceptable RF induced heating as per the standard)
    ASTM F2119-07 (Reapproved 2013): Evaluation of MR Image Artifacts from Passive ImplantsAll testing passed (implicitly, the device demonstrated acceptable MR image artifacts as per the standard)
    Functional EquivalenceThe design, materials, sterilization method, and indications for use are identical between the subject device and the predicate device (K183317, AdvantageRib System).The subject device maintains identical indications for use, design, materials, and sterilization methods compared to the predicate, with minor updates (package insert, labels, cleaning instructions, sterilization parameters, contraindications statement).
    Predicated PerformanceDevice must be at least as safe and effective as the predicates.The submission concludes that "the proposed device is at least as safe and effective as the predicates" based on the non-clinical data and similarities.

    Study Details:

    1. Sample sizes used for the test set and the data provenance:

      • The document primarily refers to non-clinical testing on the device itself for MR safety. It does not describe a clinical study or a test set of patient data.
      • For the MR safety evaluations, the "sample" would refer to the physical devices tested. The document does not specify the number of devices or components tested, but implies that a sufficient number were tested to satisfy the ASTM standards.
      • Data provenance: Not applicable in the context of clinical data. The non-clinical tests were conducted in a lab setting.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable as this is a non-clinical device clearance. Ground truth is established by the accepted ASTM standards for MR compatibility, which are developed by expert committees in their respective fields (e.g., engineers, physicists, medical professionals specializing in MRI safety).
    3. Adjudication method for the test set:

      • Not applicable. The adjudication method relates to resolving disagreements among human reviewers on ground truth, which is not relevant for non-clinical engineering tests.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This is a physical medical device (implants for rib fixation), not an AI/software device. Hence, an MRMC study is not relevant, and there's no "human readers improve with AI" component.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No. This is a physical medical device. Not applicable to standalone algorithm performance.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for demonstrating device safety and equivalence is based on established engineering standards (ASTM) for material properties, mechanical performance, and MR compatibility. For substantial equivalence, it's also based on the established safety and efficacy profile of the predicate device.
    7. The sample size for the training set:

      • Not applicable. This is not an AI/machine learning device requiring a training set.
    8. How the ground truth for the training set was established:

      • Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1