    K Number
    K203456

    Validate with FDA (Live)

    Date Cleared
    2021-06-23

    (212 days)

    Product Code
    Regulation Number
    872.3630
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Straumann CARES M-Series CAD/CAM System is indicated for the design and fabrication of single or multiple-unit implant-borne prosthetics for the restoration of partially or fully edentulous mandibles and maxillae. The system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners or Extra-Oral Scanners, CAD software, CAM software, restoration material blanks, milling machines and associated tooling and accessories. The system is used to design and fabricate CAD/CAM milled coping, crown and bridge restorations to be cemented onto Straumann® Variobase® Abutments, as well as milled abutments to be affixed to the endosseous dental implants of the Straumann® Dental Implant System using a basal screw.

    Device Description

    The Straumann CARES M-Series CAD/CAM System is intended for the design and fabrication of dental restorations by dental laboratories by means of a digital workflow. The workflow is unchanged from the primary predicate K171649. This premarket notification is introducing the Straumann Variobase Abutments for the BLX implant to abutment interface to the previously cleared workflow.

    The Straumann CARES M-Series CAD/CAM System employs optical impression files that document the topographical characteristics of teeth, traditional dental impressions, or stone models. The Straumann CARES Visual CAD software then allows the design of the desired restorations. The CAM software converts the digital restoration design into the tooling and tool path commands needed to fabricate the restoration. The CAM software also allows multiple restoration files to be combined (nested) in order to maximize the use of dental material blanks. The milling command file is encrypted prior to transfer to the M-Series mill; this encryption ensures that files generated using other CAD or CAM software cannot be used with the M-Series mill. The user will load the milling command file into the M-Series mill where it is decoded. The user loads the appropriate dental material blank and initiates the milling operation.

    This premarket notification includes restorations (copings, and bridges) manufactured from Zolid HT/Zolid SHT materials for cementation on Straumann Variobase Abutments for the BLX implant system. The BLX dental implant platforms include RB (Regular Base) and RB/WB (Regular Base/Wide Base). The combination of the coping, crown, or bridge and the Variobase Abutment component make up a two-piece abutment assembly, which is used in conjunction with endosseous dental implants for single or multiple tooth dental prostheses.

    AI/ML Overview

    Here's the information about the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/TestAcceptance CriteriaReported Device Performance
    Dynamic Fatigue Test (Straumann BLX Variobase with Zolid SHT and Zolid HT crowns)Demonstrated equivalence to the primary predicate and reference devices according to FDA guidance document "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments".The tests demonstrated that the Straumann BLX Variobase with Zolid SHT and Zolid HT crowns are equivalent to the primary predicate and reference devices.
    Milling System ValidationConfirmation that dimensions of milled restoration were the same as the intended CAD design from CARES Visual.Leveraged from K171649, confirming the dimensions of the milled restoration were the same as the intended CAD design from CARES Visual. (Subject devices were not considered a new worst case, so existing validation was referenced).
    Simulated Use ValidationConfirmation of scan, design, and production capability of the subject devices in CARES Visual.Leveraged from K171649, confirming the scan, design, and production capability of the subject devices in CARES Visual. (Subject devices were not considered a new worst case, so existing validation was referenced).
    Sterilization Process ValidationValidation according to applicable recommendations in the FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015".The sterilization process for the Straumann Variobase as recommended in the labeling was validated.

    2. Sample Size Used for the Test Set and Data Provenance

    Specific sample sizes for the test set of the dynamic fatigue tests, milling system validation, and simulated use validation are not explicitly stated in the provided document.

    The document mentions that the studies were "leveraged from K171649" (the primary predicate device) and that "the subject devices were not considered a new worst case." This implies that the validation data was likely historical or previously collected for the predicate device.

    Regarding data provenance:

    • Country of Origin: Not specified.
    • Retrospective or Prospective: Not explicitly stated, but "leveraged from K171649" suggests a retrospective application of existing data and validation reports.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. The studies mentioned are primarily engineering and performance tests (dynamic fatigue, dimensional accuracy, sterilization efficacy) and do not involve human expert interpretation of device output in a diagnostic or clinical context to establish ground truth.

    4. Adjudication Method for the Test Set

    This information is not applicable as the studies described are performance and engineering validations, not studies involving human interpretation or clinical adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study involving human readers and AI assistance was not mentioned in the document. The device described is a CAD/CAM system for fabricating dental prosthetics, which is a manufacturing process, not an AI-powered diagnostic or interpretive tool that would typically be evaluated with MRMC studies.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The "Milling System Validation" and "Simulated Use Validation" could be considered forms of standalone performance evaluation for the CAD/CAM system's components (software outputs, milling accuracy).

    • Milling System Validation: Confirmed "the dimensions of the milled restoration were the same as the intended CAD design from CARES Visual." This evaluates the algorithm's (CAD/CAM software's) ability to translate digital design into physical form, without human intervention in the final output quality check for the purpose of this specific validation.
    • Simulated Use Validation: Confirmed "the scan, design, and production capability of the subject devices in CARES Visual." This also assesses the system's performance in a simulated workflow.

    The dynamic fatigue tests evaluate the physical product, not directly the algorithm's performance.

    7. The Type of Ground Truth Used

    • Dynamic Fatigue Test: Ground truth is established by the specified FDA guidance document regarding performance requirements for dental implants and abutments. This involves objective physical testing against performance standards.
    • Milling System Validation: The "intended CAD design from CARES Visual" serves as the ground truth against which the milled restoration is compared for dimensional accuracy.
    • Simulated Use Validation: The expected functional capability of the system (scan, design, production) serves as the ground truth.
    • Sterilization Process Validation: The ground truth is adherence to the validated sterilization parameters and expected sterility assurance levels as defined by the FDA guidance document.

    8. The Sample Size for the Training Set

    This information is not provided in the document. The device is a CAD/CAM system, and while CAD/CAM software development might involve training (e.g., machine learning for design optimization or material properties), the submission focuses on validating the manufacturing process and the final product's performance, not the training of an AI model in the typical sense. The "training set" for the CAD/CAM system would implicitly be the engineering data and design principles embedded in the software.

    9. How the Ground Truth for the Training Set Was Established

    As above, this information is not provided. If there were any AI/ML components in the CAD/CAM software that involved "training," the method for establishing ground truth for that training is not discussed. For the broader CAD/CAM system's engineering and design, the ground truth would typically be established through established engineering principles, material science, and preclinical testing to ensure precise and functional designs.