510(k) Data Aggregation

    K Number
    K203424

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-12-17

    (392 days)

    Product Code
    Regulation Number
    870.4350
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Quantum Perfusion Blood Oxygenator ECC VT200-E1 and Quantum Perfusion Blood Oxygenator ECC VT160-E1 are diffusion membrane oxygenators, designed to oxygenate and remove carbon dioxide from venous blood during cardiac surgery requiring cardiopulmonary bypass and to measure blood pressure and temperature during the procedure. The device is limited to 6 hours of use.

    Quantum Perfusion Blood Oxygenator ECC VT200-E2 and Quantum Perfusion Blood Oxygenator ECC VT160-E2 are diffusion membrane oxygenators, designed to oxygenate and remove carbon dioxide from venous blood during cardiac surgery requiring cardiopulmonary bypass and to measure blood pressure during the procedure. The device is limited to 6 hours of use.

    Devices are intended for adult patients.

    Device Description

    The Quantum Perfusion Blood Oxygenator ECC (acronym VT-E) diffusion membrane device is designed to oxygenate blood and remove carbon dioxide from venous blood during cardiac surgical procedures requiring cardiopulmonary bypass for a maximum duration of 6 hours..

    Blood enters the oxygenator through the blood inlet connector, flows through a blood chamber, touching the outer surface of hollow fibers membrane; while the sweep gas flows into the hollow fiber membrane. The hollow fibers are made of Polymethy|pentene (PMP). In this chamber, carbon dioxide moves from the blood to the gas compartment, while oxygen enters into the red blood cells. Then, blood exits the oxygenator with the desired level of oxygen content and saturation, and carbon dioxide content. Sweep gas composition and flow rate are used to control saturation, and oxygen and carbon dioxide content of blood at the outlet of the oxygenator.

    The device is non-toxic, non-pyrogenic, sterilized by ethylene oxide and packaged in a single box.

    All the device surfaces in contact with a phosphorylcholine-based biocompatible coating.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the Quantum Perfusion Blood Oxygenator ECC (various models). It details the device's characteristics, intended use, and comparison to predicate devices for demonstrating substantial equivalence.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance data in a detailed quantitative manner for all parameters. Instead, it states that "All testing passed by meeting the established requirements set for the devices." The performance data is summarized under "NON-CLINICAL TESTING" as a list of tests performed.

    However, we can infer some of the performance characteristics from "Table 5-1 —Main characteristics" on page 7 and "Table 5-2 – Comparative Data" on page 9-10, which provide technical specifications of the device and its comparison to predicate devices. These act as the de facto acceptance criteria and reported performance values.

    Here's an attempt to construct a table based on the provided information, noting that specific "acceptance criteria" values are often inferred from the device's technical specifications and the predicate device's characteristics, which the new device aims to match or perform similarly to. The "Reported Device Performance" here are the self-declared specifications of the applicant's device.

    Characteristic / Performance ParameterAcceptance Criteria (Inferred/Predicate Value)Reported Device Performance (Quantum Perfusion Blood Oxygenator ECC)
    Max Operating Pressure - Blood Side760 mmHg / 100 kPa / 1 bar / 14.5 psi (This is a maximum limit, implying the device must withstand at least this)760 mmHg / 100 kPa / 1 bar / 14.5 psi
    Static Priming Volume (VT200-E1, VT200-E2)Predicate: 215 ml195 ml
    Static Priming Volume (VT160-E1, VT160-E2)Reference Predicate: 150 ml160 ml
    Exchange Surface Area (VT200-E1, VT200-E2)Predicate: 1.8 m²1.75 m²
    Exchange Surface Area (VT160-E1, VT160-E2)Reference Predicate: 1.2 m²1.45 m²
    Blood Flow (VT200-E1, VT200-E2)Predicate: 7 l/min1-7 l/min
    Blood Flow (VT160-E1, VT160-E2)Reference Predicate: 5 l/min0.5-5 l/min
    Maximum Gas FlowNot explicitly stated for predicate; device has a max gas flow.10 l/min
    Connections - Blood Side (IN/OUT)Predicate: 3/8" (9.525mm)3/8" (9.525mm)
    Temperature Probe - Blood OUT (VT200-E1, VT160-E1)Not explicitly stated for predicate; device has this feature.YSI 400
    Blood OUT Pressure Sensor - RangeNot explicitly stated for predicate; device has this feature.-200 to 600 mmHg / -27 to 80 KPa / -0.27 to 0.8 bar / -3.8 to 11.6 psi
    Material - FiberPredicate/Reference: Polymethylpentene (PMP)Polymethylpentene (PMP)
    Material - CoatingPredicate: Softline (synthetic polymer-based); Reference: PhosphorylcholinePhosphorylcholine
    Material - HousingPredicate: Polycarbonate (PC); Reference: PolycarbonateTritan copolyester
    Single-usePredicate: YesYes
    Sterile ConditionPredicate: EtO SterileEtO Sterile
    BiocompatibilityComplies with ISO 10993-1:2018 guidanceEvaluation performed, meets standard.
    Mechanical IntegrityEstablished requirements met through testing.Testing passed.
    Mechanical Resistance of ConnectorsEstablished requirements met through testing.Testing passed.
    Coating Coverage and DurabilityEstablished requirements met through testing.Testing passed.
    Operating Priming VolumeEstablished requirements met through testing.Testing passed.
    Device Pressure DropEstablished requirements met through testing.Testing passed.
    Gas Transfer PerformancesEstablished requirements met through testing.Testing passed.
    Mechanical Blood Cell DamageEstablished requirements met through testing.Testing passed.
    Ease of Prime and Air HandlingEstablished requirements met through testing.Testing passed.
    Filtration EfficiencyEstablished requirements met through testing.Testing passed.
    Temperature Probe and Pressure Sensor VerificationEstablished requirements met through testing.Testing passed.
    Product Shelf LifeComplies with EP/UPS requirements.Evaluation performed, meets standard.
    EtO Sterilization Process ValidationComplies with ISO 11135:2014.Validation performed, meets standard.
    Packaging ValidationComplies with ISO 11607-1:2019.Validation performed, meets standard.

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample sizes for the non-clinical tests (e.g., how many devices were tested for priming volume, pressure drop, etc.). It mentions that in-vitro testing was performed.

    • Data Provenance: The manufacturer is Qura S.r.l. in Mirandola, Modena, Italy. The testing was non-clinical (in-vitro), so there is no patient data provenance in terms of country or retrospective/prospective nature. The tests were performed to demonstrate compliance with international standards (ISO, EP/UPS) and FDA guidance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This section is not applicable as the studies described are non-clinical, in-vitro performance, and biocompatibility tests for medical devices, not diagnostic studies requiring expert adjudication of images or patient outcomes. The "ground truth" here is defined by physical and chemical measurements against engineering specifications and international standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as there is no human-in-the-loop diagnostic task or medical imaging analysis requiring adjudication. Adjudication methods are typically used in clinical studies or expert reviews of diagnostic interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is a blood oxygenator, not an AI-powered diagnostic tool. No MRMC study was performed or is relevant for this type of device. The submission explicitly states: "No clinical data have been included in the current Traditional 510(k) submission to support substantial equivalence to legally marketed predicate devices."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is not an algorithm or AI system. Its "performance" is mechanical, physical, and biological (biocompatibility, gas exchange efficiency). These were tested in a standalone (device-only) manner as non-clinical, in-vitro tests, as described in the "NON-CLINICAL TESTING" section.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance evaluation is established through:

    • Engineering specifications and design parameters.
    • Relevant international standards (e.g., ISO 7199 for oxygenators, ISO 11135 for sterilization, ISO 11607-1 for packaging, ISO 10993-1 for biocompatibility).
    • FDA guidance documents specific to cardiopulmonary bypass oxygenators.
    • Performance characteristics of the predicate and reference devices, against which substantial equivalence is claimed.

    8. The sample size for the training set

    This is not applicable. This is a physical medical device, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    This is not applicable for the reason stated in point 8.