510(k) Data Aggregation

    K Number
    K203257

    Validate with FDA (Live)

    Date Cleared
    2020-12-23

    (49 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LUBi Water Based Personal Lubricant is a personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

    Device Description

    LUBi Water Based Personal Lubricant is a clear, colorless, moderate sweet perfume and viscous personal lubricant that is compatible with condoms made of natural rubber latex and polyisoprene. This product is not compatible with polyurethane condoms. This device is a non-sterile personal lubricant for penile, and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is sold as an over-the-counter (OTC) product in an 8.54 fl. oz. / 250 mL size. This product is provided in a clear polyethylene terephthalate (PET) cylinder bottle. The bottles are capped with silver disc tops. The individual bottles are hermetically sealed during the production process. The device is composed of water (aqua), glycerin, propylene glycol, hydroxyethyl cellulose, sodium hyaluronate, potassium sorbate, citric acid, disodium ethylenediaminetetraacetic acid (EDTA).

    AI/ML Overview

    The provided document is a 510(k) Pre-market Notification for a medical device (LUBi Water Based Personal Lubricant). It does not contain information typically found in a study proving a device meets acceptance criteria for an AI/ML-driven device.

    Therefore, many of the requested categories (such as sample size for test set, number of experts, adjudication method, MRMC study, training set information, etc.) are not applicable to the content provided, as this is a submission for a personal lubricant, not an AI/ML diagnostic or therapeutic device.

    However, I can extract the acceptance criteria and reported device performance from the document where available for the physical and chemical properties of the lubricant, and also comment on the type of studies performed for a medical device like this.


    Acceptance Criteria and Device Performance (for a Personal Lubricant)

    The document primarily focuses on the physical, chemical, and biological compatibility aspects of the LUBi Water Based Personal Lubricant.

    1. Table of Acceptance Criteria and Reported Device Performance:

    PropertyAcceptance CriteriaReported Device Performance
    AppearanceHighly viscous liquidHighly viscous liquid
    ColorClearClear
    OdorModerate sweet perfumeModerate sweet perfume
    Viscosity (per USP<911>)28,500-30,000 mPa.sMet specification (implied, as all specifications were met across shelf life)
    pH (per USP <791>)7.5-8.0Met specification (implied, as all specifications were met across shelf life)
    Osmolality (per USP<785>)1,000-1,100 mOsm/kgMet specification (implied, as all specifications were met across shelf life)
    Antimicrobial Effectiveness (per USP<51>)- No less than a 2.0 log reduction from initial count at 14 days and no increase from 14-day count at 28 days for E. coli, P. aeruginosa, S. aureus - No increase from initial calculated count at 14 and 28 days for C. albicans, A. nigerMet specification (implied, as all specifications were met across shelf life)
    Total Aerobic Microbial Count (TAMC) (per USP<61> and <1111>)<100 CFU/gMet specification (implied, as all specifications were met across shelf life)
    Total Yeast and Mold Count (TYMC) (per USP<61> and <1111>)<10 CFU/gMet specification (implied, as all specifications were met across shelf life)
    Absence of Pathogenic Organisms (per USP<62>)Absent for S. aureus, Salmonella spp., E. coli, P. aeruginosa, Clostridia, C. albicansAll absent (Met specification, implied as all specifications were met across shelf life)
    BiocompatibilityNon-cytotoxic, non-sensitizing, non-irritating, and non-systemically toxicDemonstrated to be non-cytotoxic, non-sensitizing, non-irritating, and non-systemically toxic
    Shelf-Life3 years (device specifications met throughout)3 years (all device specifications listed in Table 1 were tested and met across the shelf life duration)
    Condom CompatibilityCompatible with natural rubber latex and polyisoprene condoms; Not compatible with polyurethane condomsCompatible with natural rubber latex and polyisoprene condoms; Not compatible with polyurethane condoms (as per ASTM D7661-10)

    Regarding an AI/ML-driven device study (which this is not):

    As the given input describes a personal lubricant and not an AI/ML-driven device, the following points are largely not applicable to this specific submission. However, if this were an AI/ML device, this is how I would address the questions based on best practices:

    2. Sample size used for the test set and the data provenance: Not applicable. For an AI/ML device, this would detail the number of cases (e.g., images, patient records) used for final model evaluation, and whether they were collected prospectively or retrospectively and their geographical origin.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. For an AI/ML device, this would specify the number and specialty of clinicians (e.g., radiologists, pathologists) who provided labels for the test data, along with their experience levels.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. For an AI/ML device, this would describe the process of resolving disagreements among experts when establishing ground truth (e.g., two experts agree, or a third expert adjudicates disagreement).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This type of study evaluates how an AI system impacts human reader performance, which isn't relevant for a personal lubricant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This evaluates the AI algorithm's performance in isolation, which isn't relevant for this device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): "Ground truth" for this device comes from laboratory testing against established physical, chemical, and biological standards (e.g., USP monographs, ASTM standards, ISO 10993).

    8. The sample size for the training set: Not applicable. This device is not an AI/ML system, so there's no "training set."

    9. How the ground truth for the training set was established: Not applicable. As there's no training set in the AI/ML sense, this question does not apply. The "ground truth" for the lubricant's properties is established by recognized scientific and regulatory standards.