510(k) Data Aggregation

    K Number
    K203087

    Validate with FDA (Live)

    Date Cleared
    2020-12-03

    (51 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CombiDiagnost R90 is a multi-functional general RF system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography

    Device Description

    The CombiDiagnost R90 is a multi-functional general Radiography/Fluoroscopy (R/F) system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography.

    The CombiDiagnost R90 is a remote-controlled fluoroscopy system in combination with high-end digital radiography. The system is suitable for routine X-ray examinations and special examinations on patients in standing, seated or lying positions. The CombiDiagnost R90 retrieves images by means of a Cesium Iodide flat panel detector.

    Philips fluoroscopy systems (standard configuration) consist of the Basic unit ("geometry" or "table unit"). Workstation Eleva Workspot (with integrated generator control, hand switch, keyboard, mouse, touch screen and PC), dual screen-monitor, Spot film device (digital camera or flat panel detector), Dynamic detector, Fixed detector, X-ray Generator Velara, X-ray tube assembly, Receptor (Flat panel detector). The optional component like Skyplate wireless portable detectors small and large, Ceiling Suspension (CSM3), Vertical Wall stand (VS2), Ceiling Suspension for monitors, monitor trolley, Remote control for RF viewer, accessories for "Stitching on the Table", are also available.

    The Eleva software of the proposed CombiDiagnost R90 is based on a workstation i.e. Eleva Workspot (computer, keyboard, display and mouse) that is used by an operator to preset examination data and to generate, process and handle digital x-ray images. As part of the radiographic system, the Eleva software is intended to acquire, process, store, display and export digital Fluoroscopy and radiographic images.

    The Eleva Software system is decomposed into software components. These components are clustered in three component collections like the image handling focused Back-end (BE), the acquisition focused Front-end (FE) and Image Processing (IP).

    • The Front-end Software is intended to acquire images.

    • The Back-end Software is intended to query patient data from Radiology Information System (RIS), store, display and export digital radiographic images to Picture Archiving and Communication System (PACS)

    • The Image Processing Software is intended to perform the pre and post processing on the acquired raw images.

    The CombiDiagnost R90 uses the same workflow from the currently marketed and predicate device, CombiDiagnost R90 (K163210) with only the following modifications:

    • additional optional components (like the reference monitor, remote control), -
    • -Eleva Workspot updated to incorporate new imaging features mainly from the previously approved reference device, DigitalDiagnost C90 (K202564) along with functional clusters like Digital Subtraction Imaging and stitching on the table
    • updates to improve usability and serviceability. -
    AI/ML Overview

    The provided text does not contain detailed acceptance criteria or a study that directly proves the device meets specific performance metrics in a way that would typically be described for an AI/CADe device. The document is a 510(k) summary for a general radiography/fluoroscopy system (CombiDiagnost R90) and focuses on demonstrating substantial equivalence to a predicate device.

    However, based on the information provided, I can infer some aspects related to acceptance and testing.

    Here's an attempt to structure the information according to your request, acknowledging the limitations from the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table of explicit, quantitative acceptance criteria for the new features (like UNIQUE 2 image processing or Bone Suppression) or their specific performance against those criteria. Instead, it relies on demonstrating that these features, which are either identical to or upgrades from previously cleared devices, maintain or improve safety and effectiveness.

    The closest to "acceptance criteria" are the standards and guidance documents followed:

    Acceptance Criteria (Implied)Reported Device Performance
    Compliance with FDA recognized standards and guidance documents (e.g., ISO, IEC safety and performance standards for X-ray equipment, FDA guidance for solid-state X-ray imaging devices).Non-clinical verification and validation tests, as well as image quality testing, were performed according to the listed standards and guidance documents. Test results demonstrate that the CombiDiagnost R90 meets the acceptance criteria and is adequate for its intended use. (Page 10)
    Maintenance/Improvement of Image Quality (for new imaging features)- UNIQUE 2 (radiography modality only): Intended to provide improved image processing, reduced noise, and improved contrast compared to UNIQUE for radiography only. It does not alter the clinical workflow, hence there is no change to the safety or effectiveness of the device. (Page 8) - The Eleva software is upgraded to incorporate new features for better imaging, and tests verified no impact on safety or effectiveness. (Page 7) - The Image Acquisition remains the same on a basic level, while incorporating new features. (Page 7)
    Mitigation of risks associated with changes (software, optional components)Safety risk assessment report conducted. Risks associated with changes are considered in the risk management report and activities, showing all risks are sufficiently mitigated and overall residual risks are acceptable. (Page 5, 10)
    No negative impact on safety and effectiveness compared to predicate.Comparisons in Table 1 and Table 2 demonstrate that minor differences in technological characteristics (like the OS upgrade, additional reference monitor, firmware tool) or the inclusion of features from a reference device (UNIQUE 2, Bone Suppression) do not affect the safety or effectiveness of the CombiDiagnost R90 when compared to the predicate device. (Pages 6-9)
    Adequate Human Factors and Usability.Human Factors and Usability Engineering Test performed. (Page 10)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in terms of patient data or image datasets for the non-clinical performance tests mentioned. The testing refers to conformance to standards and internal verification/validation processes for the device itself. Given it's a 510(k) for an X-ray system, not an AI diagnostic algorithm, the "test set" would typically refer to phantom images, system measurements, and functional tests rather than a patient image dataset.

    • Sample Size: Not explicitly stated for any specific test. The phrase "non-clinical verification and validation tests as well as image quality testing" suggests a range of internal engineering and performance tests.
    • Data Provenance: Not applicable in the context of patient data as this summary focuses on device substantial equivalence rather than a new diagnostic AI a diagnostic AI algorithm. The device manufacturer is Philips Medical Systems DMC GmbH, located in Hamburg, Germany.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided because the submission is not for an AI/CADe device requiring expert ground truth for interpretation of medical images. The "ground truth" for this type of device would be established by engineering specifications, physical measurements, and compliance with recognized standards.

    4. Adjudication Method for the Test Set

    Not applicable, as this is not an AI/CADe device requiring human interpretation adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This type of study is typically conducted for AI/CADe devices to evaluate their impact on human reader performance, which is not the primary focus of this 510(k) for an X-ray imaging system. The document explicitly states: "The CombiDiagnost R90 did not require a clinical study since substantial equivalence... was demonstrated..." (Page 11).

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is an X-ray system, not a standalone AI algorithm. It mentions "Image Processing Software" and features like "UNIQUE 2" and "Bone Suppression," but these are integrated functionalities within the imaging system to enhance image quality, not standalone diagnostic algorithms.

    7. The Type of Ground Truth Used

    For an X-ray imaging system, the "ground truth" for non-clinical testing would typically involve:

    • Engineering Specifications: Conformance to design parameters and functional requirements.
    • Physical Measurements: Using phantoms and test equipment to measure parameters like resolution, contrast, dose, noise, and geometric accuracy.
    • Compliance with Standards: Meeting the requirements outlined in recognized national and international standards (e.g., AAMI ANSI ISO 14971, AAMI ES60601-1, IEC 60601-1-2/3, IEC 60601-2-54, IEC 62220-1-1).

    The document mentions "image quality testing" (Page 10) and verifies that upgraded software features (like UNIQUE 2) aim to provide "improved image processing, reduced noise, and improved contrast" (Page 8), suggesting these metrics are part of the testing ground truth.

    8. The Sample Size for the Training Set

    Not applicable. This device is an X-ray system, not a machine learning model that requires a training set of data. The software components are developed and verified through traditional software engineering processes rather than machine learning training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no machine learning "training set" for this device.