Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diamondback 360® Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

    The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

    The Stealth 360® Peripheral Orbital Atherectorny System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

    The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectorny system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

    The Diamondback 360® Peripheral Orbital Atherectomy System Exchangeable Series is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

    The Exchangeable Series OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

    Device Description

    The DIAMONDBACK 360 Peripheral OAS, Stealth 360 Peripheral OAS, and DIAMONDBACK 360 OAS Exchangeable Series are designed to remove or reduce occlusive material and restore luminal patency by using an orbiting, diamond-coated, eccentrically mounted crown. Each OAS consists of the following main components:

    1. Reusable Saline Pump (provided non-sterile)
    2. Single-use Orbital Atherectomy Device (OAD) (provided sterile). The Exchangeable Series OAD consists of a physician-operated handle and an interchangeable crown cartridge.
    3. Single-use Atherectomy lubricant (provided sterile)
    4. Single-use Atherectomy guide wire (provided sterile)
    AI/ML Overview

    The provided text is a 510(k) summary for the DIAMONDBACK 360 Peripheral Orbital Atherectomy System, DIAMONDBACK 360 Peripheral Orbital Atherectomy System, Exchangeable Series, and Stealth 360 Peripheral Orbital Atherectomy System. This document focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a study proving device performance against those criteria in the context of an AI/ML medical device.

    Therefore, much of the requested information (acceptance criteria table, sample sizes for test and training sets, data provenance, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, and standalone performance) is not present in the provided text.

    The document states that the proposed device differs from the predicate device only in its "manufacturing process and supplier of a device component," and that its "intended use, mechanism of action, and indications for use" are the same as the predicate. The performance data presented is limited to bench testing and biocompatibility testing, which are standard for hardware medical devices to ensure safety and functionality, not to assess AI/ML algorithm performance.

    Here's what can be extracted based on the provided text:

    1. Table of acceptance criteria and the reported device performance:

    The document does not explicitly state acceptance criteria in a table format for performance metrics relevant to AI/ML devices (e.g., sensitivity, specificity, AUC). Instead, it lists performance data based on conventional medical device testing.

    Acceptance Criteria CategoryReported Device Performance
    BiocompatibilityPassed Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Systemic Toxicity, Pyrogenicity, Hemolysis Study, Complement Activation Assay, Platelet and Leukocyte Counts, Partial Thromboplastin Time.
    Bench TestingPassed Corrosion Resistance Testing, Orbit Characterization Testing, Particulate Testing, Radiopacity Assessment, Tensile Verification Testing.

    2. Sample size used for the test set and the data provenance:

    • Sample size for test set: Not mentioned. The testing described is for physical characteristics and biocompatibility, not for an AI/ML algorithm's performance on a dataset.
    • Data provenance: Not mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of experts: Not applicable, as this device is not an AI/ML algorithm requiring expert ground truth for image interpretation or similar tasks. The ground truth for biocompatibility and bench testing would be established by laboratory standards and measurements.
    • Qualifications of experts: Not applicable.

    4. Adjudication method for the test set:

    • Adjudication method: Not applicable for the type of testing described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC study: No, an MRMC study was not done, as this is not an AI/ML-driven device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Standalone performance: No, this is not an AI/ML algorithm.

    7. The type of ground truth used:

    • Type of ground truth: For biocompatibility, the ground truth would be established by standardized biological assays and observation of cellular/tissue responses against established safety thresholds. For bench testing, the ground truth is established through physical measurements, engineering specifications, and adherence to relevant industry standards.

    8. The sample size for the training set:

    • Sample size for training set: Not applicable, as this is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • How ground truth was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1