510(k) Data Aggregation

    K Number
    K202992

    Validate with FDA (Live)

    Date Cleared
    2021-04-14

    (196 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of chest CTs (with or without contrast). The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspect cases of three or more acute Rib fracture (RibFx) pathologies.

    BriefCase uses an artificial intelligence algorithm to analyze images and flag suspect cases on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for suspect cases. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

    The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

    Device Description

    BriefCase is a radiological computer-assisted triage and notification software device. The software svstem is based on an algorithm programmed component and is comprised of a standard off-the-shelf operating system, the Microsoft Windows server 2012 64bit, and additional applications, which include PostgreSQL, DICOM module and the BriefCase Image Processing Application. The device consists of the following three modules: (1) Aidoc Hospital Server (AHS) for image acquisition; (2) Aidoc Cloud Server (ACS) for image processing; and (3) Aidoc Worklist Application for workflow integration, installed on the radiologist' desktop and provides the user interface in which notifications from the BriefCase software are received.

    DICOM images are received, saved, filtered and de-identified before processing. Filtration matches metadata fields with keywords. Series are processed chronologically by running the algorithms on each series to detect suspected cases. The software then flags suspect cases by sending notifications to the Worklist desktop application, thereby prompting triage and prioritization by the attending radiologist. As the BriefCase software platform harbors several triage algorithms, the user may opt to filter out notifications by pathology, e.g., a chest radiologist may choose to filter out notifications on LVO cases, and a neuro-radiologist would opt to divert PE notifications. Where several medical centers are linked to a shared PACS, a user may read cases for a certain center but not for another, and thus may opt to filter out notification by center. Activating the filter does not impact the order in which notifications are presented in the Aidoc worklist application.

    The Worklist Application displays the pop-up text notifications of new suspected studies when they come in. Notifications are in the form of a small pop-up containing patient name, accession number and the relevant pathology (e.g., RibFx). A list of all incoming suspect cases is also displayed. Hovering over a notification or a case in the worklist pops up a compressed, low-quality, grayscale, unannotated image that is captioned "not for diagnostic use" and is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.

    Presenting the radiologist with notification facilitates earlier triaqe by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the Aidoc Medical, Ltd.'s BriefCase for RibFx Triage device, based on the provided document:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    AUC >= 0.95 (lower bound of 95% CI)AUC: 0.976 (95% CI: 0.960, 0.991)
    Sensitivity >= 80% (lower bound of 95% CI)Sensitivity: 96.7% (95% CI: 90.6%, 99.6%)
    Specificity >= 80% (lower bound of 95% CI)Specificity: 90.4% (95% CI: 85.2%, 94.3%)

    Study Details

    2. Sample Size and Data Provenance for Test Set

    • Sample Size (Test Set): 279 cases (91 ground truth positive cases with 3+ rib fractures, 188 ground truth negative cases).
    • Data Provenance:
      • Country of Origin: 3 clinical study sites (2 in the US, 1 outside the US). 70.9% of cases were from US sites.
      • Retrospective/Prospective: Retrospective.

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    • Number of Experts: Two radiologists.
    • Qualifications: Not explicitly stated beyond "radiologists."

    4. Adjudication Method for Test Set

    • Adjudication Method: Two radiologists performed ground truthing, with an additional third radiologist used to resolve inconsistencies. This implies a 2+1 adjudication model.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No, a standalone (algorithm only) performance study was conducted. There is no mention of a human-in-the-loop study to evaluate the improvement of human readers with AI assistance.
    • Effect Size of Human Readers Improvement (if applicable): Not applicable as no MRMC study was performed. However, a secondary endpoint evaluated a "time-to-notification" metric to suggest potential clinical benefit for workflow prioritization.

    6. Standalone (Algorithm Only) Performance Study

    • Standalone Study: Yes, the primary endpoint evaluated the software's performance (AUC, Sensitivity, Specificity) in identifying rib fractures on its own.

    7. Type of Ground Truth Used

    • Type of Ground Truth: Expert consensus (two radiologists with a third for inconsistencies).

    8. Sample Size for Training Set

    • Sample Size (Training Set): Not explicitly provided in the excerpt. The document states "No patient data were reused between the training and the pivotal datasets."

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth Establishment (Training Set): Not explicitly detailed in the excerpt. It states the model was "trained" on RibFx images, but the method for establishing the ground truth for those specific training images is not described. It can be inferred that a similar expert review process would have been used.