Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202983

    Validate with FDA (Live)

    Device Name
    easy Hairfull
    Manufacturer
    Date Cleared
    2020-12-08

    (69 days)

    Product Code
    Regulation Number
    890.5500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The easy Hairfull is intended to promote hair growth in males with androgenetic alopecia who have Norwood Hamilton Classifications of IIa to V and Fitzpatrick Classification of Skin Phototypes I to IV, and in females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II and Fitzpatrick Classification of Skin Phototypes I to IV.

    Device Description

    The easy Hairfull consists of 21 red visible laser diodes (LDs) and 30 red light emitting diodes (LEDs) configured within an outer helmet and protective inner liner. The use of LDs and LEDs provides for a full coverage of the upper 1/3 of the head where commonly covered with stylized hair. The system automatically pauses therapy when the device is not in the correct head position and resumes treatment when the correct head position is re-established. At the end of the therapy cycle, the system signals that therapy is complete and the device will automatically turn off.

    The device could be powered by internal rechargeable li-on battery, and it can also be supplied by specified external adapter with Input AC 100-240V, 50/60Hz and Output DC 5V, 2A.

    AI/ML Overview

    The easy Hairfull device, an infrared lamp for hair growth, received 510(k) clearance based on its substantial equivalence to a predicate device, Hair Up (K180617). The submission refers to non-clinical tests to demonstrate device performance and safety, but does not include any information regarding a clinical study assessing the device's efficacy in promoting hair growth, nor does it provide specific acceptance criteria or reported device performance metrics related to hair growth from such a study.

    Here's an analysis based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided document, there are no specific performance acceptance criteria or reported device performance results related to hair growth efficacy. The document primarily focuses on demonstrating substantial equivalence to a predicate device through technical characteristics and adherence to safety standards.

    The table below summarizes the technical characteristics compared to the predicate device, which implies that matching these characteristics was the "acceptance criteria" for demonstrating substantial equivalence and thus clearance.

    Acceptance Criteria (Technical Characteristic)easy Hairfull Performance (Proposed Device)Predicate Device (Hair Up, K180617) PerformanceDevice Meets Acceptance Criteria?
    ManufacturerY&J BIO Co., Ltd.Y&J BIO Co., Ltd.Yes
    Device Nameeasy HairfullHair UpN/A (different name)
    Classification / RegulationClass 2 / 890.5500Class 2 / 890.5500Yes
    Product CodeOAPOAPYes
    Intended UsePromote hair growth in specific androgenetic alopecia classifications (males IIa-V, females I-II) and Fitzpatrick Skin Phototypes I-IVPromote hair growth in specific androgenetic alopecia classifications (males IIa-V, females I-II) and Fitzpatrick Skin Phototypes I-IVYes
    Mode of OperationLow-level laser diodes and light emitting diodesLow-level laser diodes and light emitting diodesYes
    Wavelength655 nm655 nmYes
    Number of LDs/LEDsLDs: 21, LEDs: 30LDs: 21, LEDs: 30Yes
    Electrical Requirements (Input)AC 100V ~ 240V, 50Hz/60Hz, 5V 2AAC 100V ~ 240V, 50Hz/60Hz, 5V 2AYes
    Battery Typeli-on batteryli-on batteryYes
    How to useHelmet systemHelmet systemYes
    Treatment timeEach 20-25 min, every two days for 16 weeksEach 20-25 min, every two days for 16 weeksYes
    Maximum Power5mW5mWYes

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe an efficacy clinical study using a test set of human subjects for evaluating hair growth. The "Performance Data" section refers to non-clinical tests regarding wavelength, output power, safety, electromagnetic compatibility, photobiological safety, risk management, and usability, all performed on the device itself, not on human subjects for efficacy.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. Since no human test set was described for efficacy evaluation, there is no mention of experts establishing ground truth for such a test. The evaluation for market clearance was based on substantial equivalence for technical and safety parameters.

    4. Adjudication Method for the Test Set

    Not applicable. No human test set for efficacy evaluation was described.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-powered diagnostic or assistive technology for human readers. It's a therapeutic device using light to promote hair growth.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. As mentioned, this is a therapeutic device, not an algorithm, and the provided document does not include a standalone performance study in a clinical context for efficacy. The "Performance Data" section lists non-clinical tests performed on the device in a standalone manner (e.g., measurement of wavelength, power output), but these are not for efficacy.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    For the safety and technical performance data presented (non-clinical tests), the "ground truth" would be established by physical measurements, engineering standards, and regulatory requirements (e.g., IEC standards, 21 CFR regulations).

    For the intended use (promoting hair growth), the document does not present any efficacy data or associated ground truth. The clearance is based on the device being substantially equivalent to a previously cleared device (Hair Up K180617) which presumably had its efficacy established in prior submissions (though details of that are not in this document).

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1