    Device Name
    EYE-SYNC
    Manufacturer
    Date Cleared
    2021-10-02

    (368 days)

    Product Code
    Regulation Number
    882.1455
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EYE-SYNC® is intended for recording, viewing, and analyzing eye movements in support of identifying visual tracking impairment in human subjects.

    The EYE-SYNC® is intended to record, measure, and analyze eye movements as an aid in the diagnosis of concussion, also known as mild traumatic brain injury (mTBI), within three days of sport-related head injury in patients 17-24 years of age in conjunction with a standard neurological assessment, for use by medical professionals qualified to interpret the results of a concussion assessment examination.

    A negative EYE-SYNC® classification corresponds to eye movements that are consistent with a lack of concussion.

    A positive EYE-SYNC® classification corresponds to eye movements that may be present in patients with concussion.

    Device Description

    The SyncThink EYE-SYNC device is a portable, fully enclosed eye tracking environment with three primary components:

      1. Eye Tracker (head-mounted device) with eye tracking sensor
      1. Eye tracking display
      1. Android Tablet

    The eye tracker is a modified Samsung GearVR provided by SensoMotoric Instruments (SMI). It is mounted to the subject's face and held either by hands placed on the side or using a strap. The eye tracking sensor includes two high-speed infrared cameras (for each eye) connected to a visual display for battery, computation, and display. Camera lighting is provided by 12 high-quality Light-emitting Diodes (LEDs) centered at 850 nanometers. Eye gaze tracking is performed using a proprietary implementation of the pupil- corneal reaction method. The eye tracker display is a nonnetworked mobile device that fits within the eye tracking sensor and connects over USB. The display receives eye tracking sensor information for post-processing, manages sensor calibration, provides binocular visual display to the subject, and interfaces with the Android tablet over Bluetooth. Eye gaze tracking and visual display is combined to provide several assessment paradigms to characterize subject eye tracking performance:

    • Smooth Pursuit
    • · Saccades
    • · Vestibular-Ocular Reflex (VOR)
    • · VOR Cancellation (VORx)

    An EYE-SYNC software app on the display device manages these functions.

    The Android tablet is a standard off-the-shelf 9.7" mobile tablet from Samsung with Verizon 4G cellular connectivity. A second EYE-SYNC software app designed to provides an integrative platform for data collected on the HMD eye tracker:

    • · Patient, administrator, and records management
    • · Eye tracker assessment Bluetooth control
    • · Assessment Vestibular/Ocular-Motor Screening (VOMS) self-report tools
    • · Eye tracker assessment real-time Bluetooth monitoring
    • Eye tracker data log Bluetooth transfer
    • Eye tracker assessment analysis using visual synchronization metrics
    • · Analysis report generation with visualizations
    • · Background diagnostics to verify device health
    • · Cloud connectivity for data synchronization

    Internal batteries from the eye tracking display and Android tablet provide power for remote use, away from power source. Each device has an EYE-SYNC software app installed to provide the described functionality. EYE-SYNC is provided as a complete system and both apps are managed a single software project with identical version numbers. EYE-SYNC is provided in a standalone carry-case with user manual, strap, cleaning, and charging accessories.

    AI/ML Overview

    The provided text describes the regulatory clearance of the EYE-SYNC device. Here's a breakdown of the acceptance criteria and study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device PerformanceComments
    Sensitivity (Detection of Concussion)82% (95% CI: 74%, 89%)Meets the performance demonstrated by the predicate device (80.4%).
    Specificity (Non-detection of Concussion)93% (95% CI: 91%, 94%)Significantly higher than the predicate device (66.1%), indicating better ability to correctly identify non-concussed individuals.
    Negative Predictive Value (NPV)98% (95% CI: 97%, 99%)Higher than the predicate device (94.5%), indicating excellent ability to rule out concussion when the test is negative.
    Positive Predictive Value (PPV)56% (95% CI: 48%, 64%)Higher than the predicate device (31.6%), indicating improved ability to identify concussion when the test is positive, though still with a notable rate of false positives. (Note: PPV and NPV calculated based on observed study prevalence of 10% concussion).
    Test-Retest Reliability (for specific metrics)SD tangential error: ICC = 0.86 (0.82, 0.90) SD radial error: ICC = 0.78 (0.71, 0.84) Mean phase error: ICC = 0.83 (0.77, 0.87)High ICC values indicate strong reliability of the measurements.
    Electrical SafetyComplied with IEC 60601-1:2012 Ed. 3.0 and IEC 60601-2-57:2011 Ed1.0.
    Electromagnetic Compatibility (EMC)Complied with IEC 60601-1-2:2014 Ed 4.0.
    Software Verification and ValidationConducted according to FDA Guidance, considered "moderate" level of concern.
    CybersecurityAssessed and documented according to FDA Guidance.
    Light Safety (Photobiological Safety)Conformed to EN62471.
    BiocompatibilityCytotoxicity, Sensitization, and Irritation tests conducted and passed.(Specific results not given, but stated to be conducted in accordance with ISO 10993-1).

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 1,069 subjects
    • Data Provenance: The study was a retrospective analysis of subjects ages 17-24 years actively engaged in competitive athletics. The location/country of origin is not explicitly stated, but the context of FDA clearance for a US market implies the data is relevant to a US population or sufficiently representative thereof. The study involved patient evaluations conducted by healthcare practitioners blinded to EYE-SYNC device output "+3 days from injury," suggesting it was collected progressively for the purpose of the study.

    3. Number of Experts and Qualifications for Ground Truth

    • Number of Experts: Not explicitly stated as a specific number of individual experts.
    • Qualifications of Experts: The ground truth for concussion was established using the SCAT-5 clinical reference standard definition of concussion, with evaluations conducted by "a healthcare practitioner blinded to EYE-SYNC device output." While specific qualifications like "radiologist with 10 years of experience" are not provided, "healthcare practitioner" implies a medical professional (e.g., physician, athletic trainer, nurse practitioner) trained in concussion assessment.

    4. Adjudication Method for the Test Set

    • The text states that the SCAT-5 clinical reference standard definition of concussion was used, with evaluations conducted by a single healthcare practitioner who was blinded to the EYE-SYNC device output. There is no mention of multiple readers or an explicit adjudication method (e.g., 2+1, 3+1). The "ground truth" seems to have been based on this single blinded practitioner's assessment using the SCAT-5.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was done to evaluate how much human readers improve with AI vs. without AI assistance. The study focuses on the standalone algorithm's performance in aid of diagnosis. The device's use is explicitly "in conjunction with a standard neurological assessment," implying it's an aid to a human clinician, but the study itself does not measure the human-in-the-loop performance change.

    6. Standalone (Algorithm Only) Performance

    • Yes, a standalone performance evaluation was clearly done. The reported sensitivity, specificity, PPV, and NPV are for the EYE-SYNC device's classification algorithm. The healthcare practitioner who established the ground truth was "blinded to EYE-SYNC device output," which supports this being a standalone performance assessment of the algorithm against clinical ground truth.

    7. Type of Ground Truth Used

    • The ground truth used was expert consensus / clinical reference standard: The SCAT-5 clinical reference standard definition of concussion, as determined by a blinded healthcare practitioner. This is a widely accepted clinical tool for concussion assessment.

    8. Sample Size for the Training Set

    • The text does not explicitly state the sample size for the training set. The provided data focuses solely on the "validation data analysis" performed on 1,069 subjects, which represents the test set.

    9. How Ground Truth for Training Set Was Established

    • Since the training set size is not provided, the method for establishing its ground truth is also not explicitly described in this document. It is typically assumed that training data ground truth would be established through similar expert labeling or clinical diagnosis processes, but no details are given here.