510(k) Data Aggregation

    K Number
    K202893

    Validate with FDA (Live)

    Date Cleared
    2021-06-18

    (263 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Transcutaneous Electrical Nerve Stimulator is an electrical nerve stimulator indicated for use for pain relief by applying an electrical current to electrodes on a patient's skin to treat pain. In particular, this device is indicated for use for symptomatic relief of chronic intractable pain and adjunctive treatment of post surgical and post traumatic pain.

    Device Description

    Transcutaneous Electrical Nerve Stimulator is Transcutaneous Electrical Nerve Stimulator for pain relief. The stimulator sends gentle electrical current to underlying nerves and muscle group via electrodes applied on the skin. The device has standard programs and edit programs. It is a battery-powered portable device, comprising electronic stimulatory module and accessories of lead wires, electrodes and battery. Two outlet sockets are used to connect skin electrodes by lead wires. The accessories of electrodes is 510(k) cleared device (K192568), Size: 50*50mm.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a Transcutaneous Electrical Nerve Stimulator (TENS) device. The purpose of this submission is to demonstrate that the new device, "TENStem eco basic," is substantially equivalent to a legally marketed predicate device (K112288), also a TENS device manufactured by the same company, Wuxi Jiajian Medical Instrument Co., Ltd.

    Here's an analysis of the provided information regarding acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the parameters of the predicate device (K112288) and the relevant IEC standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10). The reported device performance is presented as a comparison between the proposed device ("Modified Device") and the predicate device in the "Substantially Equivalent Comparison Conclusion" table.

    ParameterAcceptance Criteria (Predicate Device K112288)Reported Device Performance (TENStem eco basic)Remark (Meeting Criteria)
    Intended UseSameSameMet
    Type of UsePrescription usePrescription useMet
    Power Source(s)9V Battery type 6F221.5Vx4 AAA alkaline batterySimilar (Note 1)
    Method of Line Current IsolationNAN/AMet
    Patient Leakage Current - Normal Condition (μΑ)2μΑ2μΑMet
    Patient Leakage Current - Single Fault Condition (μΑ)NANAMet
    Average DC current through electrodes when device is on but no pulses are being applied (μΑ)<0.01μΑ<0.01μΑMet
    Number of program1216Similar (Note 1)
    Number of Output channels22Met
    Synchronous or Alternating?SynchronousSynchronousMet
    Method of Channel IsolationBy TransformerBy TransformerMet
    Regulated Current or Regulated Voltage?Voltage controlCurrent controlSimilar (Note 2)
    Software/Firmware/Microprocessor Control?YesYesMet
    Automatic Overload Trip?YesYesMet
    Automatic No-Load Trip?NoNoMet
    Automatic Shut Off?NoNoMet
    User Override Control?YesYesMet
    Indicator/Display - On/Off Status?YesYesMet
    Indicator/Display - Low Battery?YesYesMet
    Indicator/Display - Voltage/Current Level?YesYesMet
    Timer Range (minutes)1~99 min10~90 minSimilar (Note 2)
    Compliance with Voluntary Standards?IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-10Met
    Compliance with 21 CFR 898?YesYesMet
    Weight (grams)170g without batteryApprox.96g without batterySimilar (Note 1)
    Dimensions (mm) [W x H x D]114 x 59 x 27 mm1406428 mmSimilar (Note 1)
    Housing Materials & ConstructionABSABSMet
    WaveformMonophasicMonophasicMet
    ShapeRectangularRectangular pulseMet
    Maximum Output Voltage (volts)35V @500Ω30V±20% @500ΩSimilar (Note 3)
    Maximum Output Current (specify units)70mA @500Ω60mA±20% @500ΩSimilar (Note 3)
    Pulse width (usec)60-300μs±20%75-300μs±20%Similar (Note 3)
    Pulse Period (msec)8.33-2000ms8.33-1000msSimilar (Note 3)
    Max. pulse frequency (Hz) [or Rate (pps)]0.5-120Hz±20%1-120Hz±20%Similar (Note 3)
    Net Charge (µC per pulse)0.83μC @500Ω0.65μC @500ΩSimilar (Note 4)
    Maximum Phase Charge, (µC)23.04µC @500Ω18µC @500ΩSimilar (Note 4)
    Maximum Average Current, (mA)2.76mA @500Ω2.16mA @500ΩSimilar (Note 4)
    Maximum Current Density, (mA/cm²r.m.s.)0.35mA/cm² @500Ω0.09mA/cm² @500ΩSimilar (Note 4)
    Maximum Average Power Density, (mW/cm²)13.27mW/cm² @500Ω2.59mW/cm² @500ΩSimilar (Note 4)
    AccessoriesElectrodes, cables, batteryElectrodes, cables, batteryMet

    Justification for Similarities (Notes from the original document):

    • Note 1: Differences in battery type, number of programs, weight, and dimensions are deemed insignificant in terms of safety or effectiveness, as the device passed relevant IEC tests and the battery is commonly used.
    • Note 2: Differences in "Regulated Current or Regulated Voltage" and "Timer Range" are considered minor and do not raise new safety/effectiveness issues as both devices passed IEC 60601-1, and the timer is user-adjustable.
    • Note 3: Differences in maximum output voltage/current, pulse width, pulse period, and max pulse frequency are analyzed. The proposed device's parameters are within safety limits based on calculation and passed IEC 60601-2-10 tests.
    • Note 4: "Net Charge," "Maximum Phase Charge," and "Maximum Average Current" are similar and comply with IEC 60601-2-10 requirements. The proposed device's maximum average current is smaller, indicating better safety. Differences in maximum current density and maximum average power density are due to electrode area calculation but comply with IEC 60601-2-10 and the 0.25W/cm² limit respectively.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a "test set" in the context of clinical data or patient samples. The testing described is non-clinical bench testing to verify hardware specifications and compliance with safety standards (IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2). The sample size for this type of testing typically refers to the number of devices tested to ensure consistent performance, which is not specified but is assumed to be an adequate number for regulatory compliance.

    The data provenance is from bench tests conducted on the Transcutaneous Electrical Nerve Stimulator, presumably at a testing facility compliant with quality system regulations. The manufacturer is Wuxi Jiajian Medical Instrument Co., Ltd. from Wuxi, Jiangsu, China. This is a retrospective comparison against an existing predicate device, as opposed to a prospective clinical trial.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable to this submission. The "ground truth" for comparative effectiveness in a clinical context (e.g., diagnostic accuracy of an AI algorithm) is established by expert readers or pathology. Here, the "ground truth" for showing substantial equivalence relies on:

    • Compliance with recognized international standards (IEC 60601 series).
    • Direct comparison of technical specifications against a previously cleared device.
    • Risk analysis of any noted differences.

    There is no mention of human experts interpreting test results from the device in a clinical setting to establish "ground truth" for performance.

    4. Adjudication Method for the Test Set

    This information is not applicable. No clinical test set involving human interpretation requiring adjudication is described. The assessment here is based on objective, quantifiable engineering and electrical performance measurements against established standards.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "Clinical data was not including in this submission." MRMC studies are typically performed for diagnostic or prognostic devices to assess the impact of an AI algorithm on human reader performance. This device is a medical device for pain relief, whose performance is primarily evaluated through electrical parameters and safety standards, not diagnostic accuracy requiring human reader studies.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This concept is not applicable in the context of this TENS device. "Standalone performance" or "algorithm-only performance" usually refers to the accuracy or performance of an AI algorithm independent of human intervention. This device is an electrical stimulator, not an AI-driven diagnostic or treatment planning system that would have a "standalone algorithm." Its performance is defined by its electrical output characteristics and safety.

    7. The Type of Ground Truth Used

    The "ground truth" used for demonstrating substantial equivalence is based on:

    • Compliance with recognized international safety and performance standards: Specifically, ANSI AAMI ES60601-1, IEC 60601-2-10, and IEC 60601-1-2. These standards define the acceptable range of electrical parameters, safety features, and electromagnetic compatibility for medical electrical equipment, including TENS devices.
    • Direct comparison to the technical specifications of a legally marketed predicate device (K112288): The predicate device itself serves as a "ground truth" for acceptable performance parameters for this type of device. Any deviations from the predicate are justified by showing they either fall within acceptable variations, are safer, or still comply with relevant standards.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a Transcutaneous Electrical Nerve Stimulator, which is a hardware medical device, not a machine learning model or algorithm that requires a "training set" of data.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable for the reasons stated in point 8.