    K202850

    Validate with FDA (Live)

    Date Cleared
    2021-02-09

    (134 days)

    Product Code
    Regulation Number
    870.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Concerto Versa™ Detachable Coil is indicated for arterial and venous embolization in the peripheral vasculature.

    Device Description

    The Concerto Versa™ Detachable Coil is an embolization coil that consists of a platinum embolization coil attached to a composite delivery pusher, and a hand-held Instant Detacher (I.D.) which, when activated, detaches the coil from the delivery pusher tip. The Instant Detacher is an accessory sold separately and was most recently cleared in K162704. The Concerto Versa™ Detachable Coil is supplied sterile and is intended for single-use only.

    AI/ML Overview

    The provided document is a 510(k) summary for the Concerto Versa™ Detachable Coil. This document pertains to a medical device (an embolization coil) and does not describe a study involving an AI/Machine Learning device or a diagnostic device. Therefore, many of the requested categories, such as "Adjudication method," "MRMC comparative effectiveness study," "Standalone performance," "Training set," and "Ground truth for training set," are not applicable to this submission.

    The document primarily focuses on demonstrating the substantial equivalence of the new device (Concerto Versa™ Detachable Coil) to a legally marketed predicate device (Concerto Detachable Coil System) through a series of bench tests and leveraging existing data from the predicate. The "acceptance criteria" are implicitly the passing results of these non-clinical tests.

    Here's the information that can be extracted and how it relates to the request:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The acceptance criteria are generally "Pass" for each test, indicating that the device met the specified performance requirements. The document does not provide specific numerical criteria for each test beyond stating they "meet specification."

    Test NameAcceptance Criteria (Implicit)Reported Device Performance
    Friction in CatheterMeet specificationPass
    Coil Force Transfer RatioMeet specificationPass
    Pusher LengthMeet specificationPass
    Coil Visual InspectionNo damage, kinks, 1st loop fullPass
    Friction in SheathMeet specificationPass
    FatigueRe-sheathed/deployed 6+ times without breaking/detachingPass
    PP TensileMeet specificationPass
    Kink resistanceTrack through 0.027" catheter without kinkingPass
    DetachmentMust detach when attemptedPass
    Tip BucklingMeet min/max specificationsPass
    Tinius OlsenMeet specificationPass
    Loop DeformationMeet specificationPass
    Anchor Flow RateMeet specificationPass
    1st Loop OD MeasurementMeet specification based on sizePass
    Flow OcclusionMeet specification in vessel modelPass
    Primary Wind OD MeasurementMeet specificationPass
    Coil LengthMeet specification based on modelPass
    Pusher ElongationMeet specificationPass
    Pusher Skive TensileMeet specificationPass
    Primary wind weld tensile strengthMeet specificationPass
    Couple Tube TensileMeet specificationPass
    Shield Coil TensileMeet specificationPass
    Detachment (time)Less than 5sPass
    Detachment (back-up method)Manually detachable by back-up methodPass
    Fiber Amount CountingMeet specification based on modelPass
    Fiber Pull Out ForceMeet specificationPass
    Distribution SimulationMaintain integrity with packaging through shippingPass

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document does not explicitly state the specific number of coils or test replicates used for each bench test conducted on the Concerto Versa™ Detachable Coil. It states that "A subset of design verification testing was conducted."
    • Data Provenance: The tests were "non-clinical bench testing." The document mentions "clinically relevant tortuosity model of the splenic and gastroduodenal arteries" which implies in-vitro testing. It also states that "GLP acute animal data was also leveraged from the predicate device," indicating some previous animal study data was used, but not for the subject device's new testing.
    • Retrospective or Prospective: The testing performed for this submission would be considered prospective for the subject device, as it was specifically done for this 510(k) submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. This is a medical device submission based on physical and mechanical performance testing, not a diagnostic device requiring expert interpretation for ground truth.

    4. Adjudication Method for the Test Set:

    Not applicable. No expert adjudication process is described as it's not a diagnostic study.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, this was not an MRMC comparative effectiveness study. It is a 510(k) submission for a non-diagnostic medical device.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a medical device, not an algorithm. The performance evaluation is based on mechanical and physical characteristics.

    7. The Type of Ground Truth Used:

    The "ground truth" for the non-clinical performance tests is the defined engineering and material specifications for the device. For example, a coil's length must meet a certain specification, or its detachment time must be below a certain threshold. There's no "ground truth" derived from patient outcomes or expert consensus in the typical sense for a diagnostic device. Biocompatibility was evaluated against ISO 10993 standards.

    8. The Sample Size for the Training Set:

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable.