510(k) Data Aggregation

    K Number
    K202633

    Validate with FDA (Live)

    Device Name
    ScanX Edge
    Manufacturer
    Date Cleared
    2020-10-07

    (26 days)

    Product Code
    Regulation Number
    872.1800
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ScanX Edge is intended to be used for scanning and processing digital images exposed on Phosphor Storage Plates (PSPs) in dental applications.

    Device Description

    The ScanX Edge is a dental device that scans photostimulable phosphor storage plates that have been exposed in place of dental X- Ray film and allows the resulting images to be displayed on a personal computer monitor and stored for later recovery. It will be used by licensed clinicians and authorized technicians for this purpose. The device is an intraoral Plate Scanner, which is designed to read out intraoral Plates of the sizes 0, 1 and 2. intraoral plate Scanners are state of the art since 2002 and are available in various designs from many companies around the world. The intraoral Plates are put into the mouth of the patient, exposed to X-rays and then are read out with the device. The read-out-process is carried out with a 635nm Laser. The laser beam is moved across the surface of the plate by an oscillating MEMS mirror. The laser beam stimulates the top coating of the plates, which consists of x-ray sensitive material. Depending on the exposed dose, the coating emits different levels of light. These light particles are then requisitioned by an optical sensor (Photo Multiplier Tube/ PMT) and transferred into an electrical output signal is digitalized and is the data for the digital X-ray image. The data is transmitted via an Ethernet link to a computer. Before the plate is discharged, the remaining data is erased by a LED-PCB. The user chooses which size of plate he has to use and prepares the device by inserting the appropriate plate insert into the device. He then exposes the plate and then puts the plate directly into the insert by pushing it out of the light protection envelope. The user closes the light protection cover and starts the read out process. After the read out process the picture is transmitted to the connected PC, the picture can be viewed and the IP is erased and ready to use for the next acquisition.

    AI/ML Overview

    The provided text describes the Dürr Dental SE ScanX Edge device (K202633) and its substantial equivalence to a predicate device, the ScanX Intraoral View (K170733). However, it does not contain a study that establishes acceptance criteria and then proves the device meets those criteria with performance data from a specific study.

    Instead, the document focuses on demonstrating that the ScanX Edge maintains similar performance characteristics and safety profiles compared to a previously cleared device. It references technical testing and standards compliance rather than a clinical performance study with defined acceptance criteria and performance metrics for the device itself.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance from a dedicated study as that information is not present in the provided text. I will extract the available performance specifications and the context of their assessment.

    Here's a breakdown of the information that can be extracted from the document, with explanations for what is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    As noted above, the document does not present a specific study that defines acceptance criteria and then reports performance against those criteria. Instead, it compares the ScanX Edge's technical specifications to those of its predicate device, often stating that the values are "similar." These comparisons are made against established standards for general X-ray imaging devices.

    Metric (Acceptance Criteria are inferred as being "Similar" or "More than XX% at YY lp/mm" to the predicate)Predicate Device (ScanX Intraoral View) PerformanceSubject Device (ScanX Edge) Performance
    Image quality - Theoretical resolutions10, 20, 25 or 40 LP/mmMax. theoretical resolution Approx. 16.7 Lp/mm
    MTF (Modulation Transfer Function) at 3 lp/mmMore than 45%More than 40%
    DQE (Detective Quantum Efficiency) at 3 lp/mmMore than 7.5%More than 3.4%
    Image data bit depth16 bits16 bits

    Note: The document states that the "lower DQE did not adversely affect image quality as shown by our SSD testing," implying that despite a numerically lower DQE, the overall image quality was deemed acceptable, likely within the context of substantial equivalence to the predicate in dental applications.


    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not specified. The document refers to "MTF and DQE image performance testing" and "SSD testing" but does not provide details on the number of images or cases used for these tests.
    • Data Provenance: Not specified. The document does not mention the country of origin of the data or whether it was retrospective or prospective. Implicitly, as Dürr Dental SE is a German company, some testing might have occurred in Germany, but this is not stated for the performance data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.

    The performance testing referenced (MTF, DQE, noise power spectrum) are objective technical measurements performed on the device itself, often using phantoms or controlled inputs, not human interpretation of clinical images. Therefore, the concept of "ground truth established by experts" as typically applied in diagnostic AI studies is not directly applicable here for these specific technical metrics. The "SSD testing" mentioned in relation to DQE might involve human assessment, but details are not provided.


    4. Adjudication method for the test set

    Not applicable/Not specified for the referenced technical performance tests (MTF, DQE). These are objective measurements rather than subjective assessments requiring adjudication.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted or reported in this document.
    • Effect Size: Not applicable, as no MRMC study was performed.

    This device is described as an "Extraoral source x-ray system" for scanning phosphor plates and displaying digital images. It is not an AI algorithm intended to assist human readers in diagnosis. Therefore, an MRMC study assessing AI assistance is outside the scope of this device's function as described.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    A "standalone" performance assessment was done in the sense that the intrinsic physical image quality metrics (MTF, DQE, theoretical resolution) of the device were measured and reported. These metrics reflect the device's inherent capability to produce high-quality images, independent of human interpretation or assistance. However, it's not an algorithm that performs a diagnostic task.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the MTF, DQE, and theoretical resolution measurements are based on physical properties and standardized test methods as outlined in IEC 62220-1:2003 ("Medical electrical equipment - Characteristics of digital X-ray imaging devices - Part 1: Determination of the input properties of digital X-ray imaging devices"). This typically involves using phantoms and known physical inputs rather than clinical ground truth like pathology or expert consensus.


    8. The sample size for the training set

    Not applicable. This device is a hardware imaging system, not an artificial intelligence algorithm that requires a training set.


    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this hardware device.