Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202532

    Validate with FDA (Live)

    Date Cleared
    2021-08-21

    (354 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect the surgical patient and operating room personnel from the transfer of microorganisms, body fluids and particulate material. This surgical gown meets the requirements of AAMI Level 3 barrier protection for a surgical gown per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities (AAMI PB70). The Surgical Gowns are single use, disposable medical devices; provided sterile or non-sterile. Non-sterile gowns are to be sold to re-packager/re-labeler establishments for ethylene oxide (EtO) sterilization according to ISO 11135-1 prior to marketing to the end users and Sterile surgical gowns are to be sold directly to users after EtO sterilization validation to ISO 11135-1.

    Device Description

    Surgical Gown is designed for the medical personnel using in operation. The employed material is SMS compound non-woven fabric. The material has many good properties, such as soft, clean, good filtration and uniformity and waterproof, they are not sensitive to human beings, difficult to fluff, they don't have any peculiar smell, other matters and primary color. The gowns are for safe use in the operating room environment, i.e., lint free, free of toxic ingredients and non-fast dyes. It is a kind of Non- Reinforced surgical gown. The chest front and sleeve critical zones of the Disposable Surgical Gowns are constructed from a blue polyolefin SMS (spunbond, meltblown, spunbond) and have been tested according to AAMI PB70:2012 and meet AAMI Level 3 barrier level protection for a surgical gown. The Disposable Surgical Gown is a single use, disposable medical device that will be provided in a variety of nonsterile packaging configurations.

    AI/ML Overview

    The provided text describes the 510(k) summary for Disposable Surgical Gowns (K202532). It details non-clinical testing conducted to demonstrate substantial equivalence to a predicate device (K170762).

    Here's the breakdown of the acceptance criteria and the study proving the device meets them, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Disposable Surgical Gowns are primarily based on meeting the performance requirements for AAMI Level 3 barrier protection as per ANSI/AAMI PB70:2012, along with other physical properties and biocompatibility.

    Testing Methodology / ParameterAcceptance CriteriaReported Device PerformanceResult
    Barrier Protection
    Hydrostatic Pressure (AATCC 127:2014)≥ 50 cm H2OAverage of 203 cm on Front body material, 168 cm on back body material, and 60 cm on seam.PASS
    Water Impact (AATCC 42:2013)≤ 1.0 g0g on Critical Zone (Body material, Sleeve seam and Adhesive part of belt)PASS
    Level (AAMI PB70)Level 3Level 3Same as Predicate
    Resistance to blood and liquid penetration (AAMI PB70)Level 3 AAMI PB70Level 3 AAMI PB70Same as Predicate
    Physical Properties
    Tear strength (N) (Trapezoid Method, ASTM D 5587-2015)MD/CD ≥ 10NMD Mean 60.4N, CD Mean 40.2NPASS
    Breaking Strength (ASTM D 5034-2009(2017))MD/CD ≥ 30NMD Mean (not reported for proposed device), CD Mean 88.2NPASS
    Sewn seam strength (sleeve seam) (ASTM D1683/D1683M-2017(2018))Seam Strength ≥ 30NSeam Strength Mean: 64.5 NPASS
    Water-vapor resistance (ASTM F 1868-2017)≤ 3.0 Pa·m²/WMean: 2.42 Pa·m²/WPASS
    Lint and other particles generation in dry state (ISO 9073-10: 2003)Coefficient of linting Log10 ≤ 4.0Log10 Mean: 3PASS
    Burning Behavior (16 CFR Part 1610)Class IClass IPASS
    Biocompatibility
    Cytotoxicity (ISO 10993-5: 2009)Non-CytotoxicDevice is non-cytotoxic.PASS
    Irritation (ISO 10993-10: 2010)Non-IrritatingDevice is non-irritating.PASS
    Sensitization (ISO 10993-10: 2010)Non-SensitizingDevice is non-sensitizing.PASS

    2. Sample sizes used for the test set and the data provenance

    The document specifies the following sample sizes for the non-clinical tests:

    • Hydrostatic Pressure (AATCC127:2014): 3 non-consecutive lots tested, using a sample size of 32/lot.
    • Water-proof Property (AATCC42:2013): 3 non-consecutive lots tested, using a sample size of 32/lot.
    • Tear strength (Trapezoid Method, ASTM D 5587-2015): Sample size of 5 pcs.
    • Breaking Strength (ASTM D 5034-2009(2017)): Sample size of 8 pcs.
    • Sewn seam strength (sleeve seam) (ASTM D1683/D1683M-2017(2018)): Sample size of 5 pcs.
    • Water-vapor resistance (ASTM F 1868-2017): Sample size of 3 pcs.
    • Lint and other particles generation in the dry state (ISO 9073-10: 2003): Sample size of 5 pcs.
    • Mass per unit area (ASTM D3776/D3776M-2009(2017)): Sample size of 5 pcs.
    • Burning Behavior (16 CFR Part 1610): Sample size of 5 pcs.
    • Biocompatibility (Cytotoxicity, Irritation, Sensitization): Specific sample sizes are not explicitly stated, but the tests were performed according to ISO 10993-5 and ISO 10993-10.

    Data Provenance: The document states that the non-clinical tests were conducted to verify that the proposed device met design specifications. Given the applicant's address (Anhui Medpurest Medical Technology Co.,Ltd, China) and the submission correspondent's address (Shanghai SUNGO Management Consulting Co., Ltd, China), it can be inferred that the testing was performed in China or by labs contracted by the applicant. The data is retrospective in nature, as it's a summary of tests already performed for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This document describes non-clinical performance and biocompatibility testing for a physical medical device (surgical gowns). It does not involve diagnostic image analysis or clinical interpretation. Therefore, the concept of "experts establishing ground truth for a test set" in the context of medical image analysis or similar tasks is not applicable here. The ground truth for these tests is established by standardized testing methodologies and instruments as defined by the listed ASTM, AATCC, ISO, and CPSC standards.

    4. Adjudication method for the test set

    As this is non-clinical performance and material property testing, an adjudication method (like 2+1 or 3+1) is not applicable. The results are quantitative measurements obtained through defined protocols. Pass/Fail criteria are based on exceeding or falling below specified thresholds in the relevant standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices where human interpretation is assisted by AI. This submission is for a physical medical device (surgical gowns), and no AI component is described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone performance study was not done in the context of an algorithm. This submission is for a physical medical device (surgical gowns) and does not involve an algorithm. The "standalone" performance here refers to the physical and barrier properties of the gown itself, which are measured directly through the non-clinical tests.

    7. The type of ground truth used

    The ground truth for this device's performance is established by standardized laboratory testing and measurements based on the referenced national and international standards (e.g., AAMI, ANSI, ASTM, AATCC, ISO, CPSC). For example:

    • For barrier protection, the ground truth is the measured hydrostatic pressure and water impact performance against the specified level (AAMI Level 3).
    • For physical properties like tear strength and breaking strength, the ground truth is the force required to tear or break the material as measured by testing machines.
    • For biocompatibility, the ground truth is the measured biological response (e.g., cell viability for cytotoxicity, skin reaction for irritation/sensitization) compared to established toxicity thresholds.

    8. The sample size for the training set

    Not applicable. This submission is for a physical medical device (surgical gowns) and does not involve AI/machine learning, and therefore does not have a "training set."

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set mentioned or implied for this device, the method for establishing ground truth for a training set is irrelevant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1