    K Number
    K202405

    Validate with FDA (Live)

    Date Cleared
    2021-01-07

    (139 days)

    Product Code
    Regulation Number
    870.1025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BeneVision N12/N15/N17/N19/N22

    The BeneVision N12N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure(IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO2/ScvO2), Carbon Dioxide (CO2), Oxygen (O2), Anesthetic Gas (AG), Impedance Cardiograph (ICG), Bispectral Index (BIS), Respiration Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional Oxygen Saturation (rSO2). The system also provides an interpretation of resting 12-lead ECG.

    All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:

    · BIS, RM, CCO, SvO2/ScvO2, PAWP, and NMT monitoring, PNP, and PNC are intended for adult and pediatric patients only:

    · C.O. monitoring and A-Fib are intended for adult patients only;

    • ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.

    · rSO2 monitoring is intended for use in individuals greater than 2.5kg.

    The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/N22 monitors are not intended for helicopter transport. hospital ambulance, or home use.

    BeneVision N1 Patient Monitor:

    The BeneVision N1 Patient Monitor is intended for monitoring, displaying, storing , alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, and Heart Rate (HR)), Respiration (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP) , Pulmonary Artery Wedge Pressure (PAWP), Carbon Dioxide (CO2) and Oxygen (O2). The system also provides an interpretation of resting 12-lead ECG.

    All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:

    • PAWP, PNP, and PNC are intended for adult and pediatric patients only;
    • A-Fib is intended for adult patients only;

    The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment. It should be used by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for home use.

    Device Description

    The subject BeneVision N Series Patient Monitors includes six monitors:

    • BeneVision N12 Patient Monitor
    • BeneVision N15 Patient Monitor
    • BeneVision N17 Patient Monitor ●
    • . BeneVision N19 Patient Monitor
    • . BeneVision N22 Patient Monitor
    • BeneVision N1 Patient Monitor ●

    Mindray's BeneVision N Series Patient Monitors provide a flexible software and hardware platform to meet the clinical needs of patient monitoring.

    AI/ML Overview

    The provided document describes the BeneVision N Series Patient Monitors and their clearance by the FDA based on substantial equivalence to predicate devices, particularly focusing on changes and new features. The document details the device's intended use, technological comparisons, and performance data from various tests. However, it does not explicitly provide a table of acceptance criteria and reported device performance in the format requested. Instead, it states that "the results of the bench testing show that the subject device meets its accuracy specification and is substantially equivalent to the predicate device."

    Specifically regarding "Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm)" which is the relevant regulation name for the product code MHX (used by the BeneVision N Series Patient Monitors), the document states:
    "Add arrhythmia detection for neonatal patients when used with the MPM 3.0 module." (Page 9, Table 2 ECG section; Page 22, Table 3 ECG section).
    And references "K200015 - ePM series Patient Monitors...: provided as reference devices for ECG algorithm supporting arrhythmia detection in neonate that has been added to the subject BeneVision N Series Patient Monitors." (Page 6, Section 4).

    Given the information in the document, here's a structured response:

    Acceptance Criteria and Device Performance Study for Arrhythmia Detection in Neonates

    The provided FDA 510(k) summary (K202405) for the BeneVision N Series Patient Monitors primarily focuses on demonstrating substantial equivalence to a predicate device (K192972) and a reference device (K200015) for the new feature of neonatal arrhythmia detection. While it explicitly states that the device meets its accuracy specifications, it does not provide a specific table of acceptance criteria or detailed reported performance metrics for neonatal arrhythmia detection. Instead, it relies on general statements about meeting specifications and compliance with standards.

    However, based on the document's information and common regulatory expectations for such devices, we can infer the type of acceptance criteria and the nature of the study.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly provide a table for the acceptance criteria and reported device performance for neonatal arrhythmia detection. It states that the device "meets its accuracy specification." For arrhythmia detection systems, typical performance metrics and acceptance criteria, often derived from standards like IEC 60601-2-27, would generally include:

    MetricAcceptance Criteria (Inferred from regulatory standards and typical arrhythmia detector performance)Reported Device Performance (Implicitly stated as "meets accuracy specifications" and "substantially equivalent")
    Arrhythmia Detection Accuracy (Sensitivity/Specificity)e.g., Sensitivity ≥ 90%, Specificity ≥ 95% for various arrhythmia types (e.g., asystole, bradycardia, tachycardia, PVCs) in applicable age groups (neonates)Met (stated as "meets accuracy specification")
    False Alarm Ratee.g., False QRS detection rate < 1%, False alarm rate per hour < 0.2Met (stated as "meets accuracy specification")
    Response Time to Alarmse.g., Alarm generation within X seconds of event detectionMet (stated as "meets accuracy specification")
    Heart Rate Accuracy±1 bpm or ±1%, whichever is greater (for neonates, 15-350 bpm range)Met (as per device specifications in Table 2 & 3 ECG section)
    Pace DetectionAccurate detection of paced beats; no false QRS detection due to pacing spikesMet (stated as "pace detection enhancement" and "meets accuracy specification")
    PVCs/min Range (Neonates)[0, 350]Met (as per device specifications in Table 2 & 3 ECG section)

    2. Sample Size for Test Set and Data Provenance

    The document does not specify the exact sample size used for the test set for the neonatal arrhythmia detection algorithm. It broadly states that "Mindray conducted functional and system level testing on the subject device." The data provenance is also not explicitly stated in terms of country of origin or whether it was retrospective or prospective data. However, given that it's a 510(k) for a patient monitor and a feature addition (neonatal arrhythmia detection), it's highly likely that a combination of retrospective ECG databases (e.g., MIT-BIH Arrhythmia Database, AHA database, or internal proprietary databases containing neonatal ECGs) and/or prospective testing on simulated or recorded neonatal ECGs would have been used.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not specify the number or qualifications of experts used to establish the ground truth for the test set related to neonatal arrhythmia detection. In studies for arrhythmia detection, ground truth is typically established by certified cardiologists, electrophysiologists, or other medical specialists trained in ECG interpretation, often with years of experience.

    4. Adjudication Method

    The document does not specify any adjudication method (e.g., 2+1, 3+1, none) used for establishing the ground truth or evaluating the test set results. For arrhythmia detection, a common method involves multiple experts reviewing the same data, with discrepancies resolved by a senior expert or consensus voting (e.g., 2 out of 3 agreement).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study directly related to the new neonatal arrhythmia detection feature or the overall performance of the patient monitor with human readers. The submission focuses on the performance of the device itself and its substantial equivalence to predicate devices, not on the improvement of human reader performance with AI assistance. Therefore, no effect size of human reader improvement with AI vs. without AI assistance is reported.

    6. Standalone Performance Study

    Yes, a standalone performance study was done for the algorithm. The document explicitly states: "Mindray conducted functional and system level testing to validate the performance of the devices. The results of the bench testing show that the subject device meets its accuracy specification and is substantially equivalent to the predicate device." This indicates that the algorithm's performance was evaluated independently of human-in-the-loop interaction.

    7. Type of Ground Truth Used

    The document does not explicitly state the type of ground truth used for the testing. For arrhythmia detection algorithms, the ground truth is typically established through expert consensus on annotated ECG waveforms, possibly validated against physiological events or other clinical data where available.

    8. Sample Size for the Training Set

    The document does not provide any information regarding the sample size for the training set used to develop or refine the neonatal arrhythmia detection algorithm.

    9. How Ground Truth for the Training Set Was Established

    Similarly, the document does not describe how the ground truth for the training set was established. Analogous to the test set, it would typically involve manual annotation of ECG data by experts, but this information is not present in the provided text.