Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202340

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-10-15

    (59 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DROPLET® PEN NEEDLES 30G & 33G are intended for use with pen injector devices for the subcutaneous injection of drugs.

    Device Description

    DROPLET® PEN NEEDLES 30G & 33G 4G are sterile, single use needles intended for use with pen injector devices for the subcutaneous injection of drugs. The pen needles are OTC devices. The pen needle assembly consists of a double-ended cannula that is assembled into an injection molded hub using adhesive. The hub has internal threads, which allow it to be screwed onto the pen injector device. This allows the cartridge end of the cannula to penetrate through the rubber septum of the cartridge. The patient end and the cartridge end of the cannula are lubricated using a silicone -based lubricant for ease of injection and rubber septum penetration. There is an inner needle shield assembled over the patient end of the cannula to protect the needle point from damage and accidental needle sticks. There is also an outer cover. Each pen needle assembly is protected with a peel away seal to provide a sterility barrier.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a medical device: Droplet Pen Needle 30G & 33G.

    This document describes a submission for a new medical device seeking to demonstrate substantial equivalence to a previously cleared predicate device, not a study proving the device meets acceptance criteria for an AI/ML powered device. Therefore, much of the requested information regarding AI/ML device performance metrics, such as ground truth establishment with experts, MRMC studies, and training/test set details, is not applicable to this document.

    However, I can extract the acceptance criteria and non-clinical performance data for this specific medical device, which are based on compliance with an international standard (ISO 11608-2:2012) and biocompatibility testing.

    Here's a breakdown of the information that can be extracted and what cannot:

    Information that can be extracted from the document:

    • Acceptance Criteria (as defined by standard compliance) and Reported Device Performance: This is derived from the "Non-Clinical Performance Data" section.
    • Sample size for the test set: Not explicitly stated as "test set" in the context of AI, but the non-clinical tests would have involved samples of the device. The document does not specify the number of units tested for each parameter.
    • Data Provenance: The tests were performed by HTL-Strefa S.A. (Poland). The data is non-clinical, related to device performance in laboratory settings, not patient data.
    • Type of Ground Truth: For this type of device, the "ground truth" is defined by the technical specifications and requirements outlined in the ISO standard and the results of various physical, chemical, and biological tests.

    Information that CANNOT be extracted from the document (as it's not an AI/ML device submission):

    • Number of experts used to establish ground truth
    • Qualifications of those experts
    • Adjudication method for the test set
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness study
    • Standalone (algorithm only) performance
    • Sample size for the training set
    • How the ground truth for the training set was established

    Acceptance Criteria and Study Proving Device Meets Acceptance Criteria for Droplet Pen Needle 30G & 33G

    As this is a submission for a physical medical device (hypodermic needle) and not an AI/ML powered device, the "acceptance criteria" are defined by compliance with relevant international standards and successful completion of non-clinical performance and biocompatibility testing. There is no AI model or human-in-the-loop performance measurement.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device performance is demonstrated by its compliance with the requirements of ISO 11608-2:2012 Needle-based injection systems for medical use -- Requirements and test methods -- Part 2: Needles.

    Test ParameterClause no. & Requirement of ISO 11608-2:2012Acceptance Criteria (Met if "Meets requirements")Reported Device Performance
    Materials4.1 The needle shall be made of tubing materials specified in ISO 9626.Meets requirementsMeets requirements
    Dimensions4.2 The needles shall fit the test apparatus specified in item 7.3 of ISO 11608-2.Meets requirementsMeets requirements
    Determination of flow rate through the needle4.3 The needle was tested in accordance with Annex A to ISO 11608-2 to determine flow rate through the needle.Meets requirementsMeets requirements
    Bond between hub and needle tube4.4 The union of the hub and needle tube shall not break when tested in accordance with Clause 9 of ISO 11608-2.Meets requirementsMeets requirements
    Needle points4.5 When examined under a magnification of x2.5, needle points shall appear sharp and free from feather edges, burrs and hooks.Meets requirementsMeets requirements
    Freedom from defects4.6 The needle tube shall fulfill the requirements of ISO 7864, 11.3.Meets requirementsMeets requirements
    Lubrication4.7 The needle tube should be lubricated at both the patient end and the cartridge end. The lubricant shall not, under normal or corrected-to-normal vision, be visible as droplets of fluid on the outside surface of the needle tube.Meets requirementsMeets requirements
    Dislocation of measuring point at patient end4.8 Dislocation of the cannula point at the patient end shall be in accordance with Table 2 below when tested as per Clause 8 (of ISO 11608-2).Meets requirementsMeets requirements
    Determination of functional compatibility with needle-based injection systems4.9 Compatibility with any NIS shall be claimed only after testing in accordance with Clause 11.Meets requirementsMeets requirements
    Ease of assembly and disassembly4.10 Attachment of the needle shall be possible without removing the needle from its opened unit packaging. Compliance is checked according to the requirements of Clause 11.Meets requirementsMeets requirements
    Sterility4.11 The needle in its unit packaging shall has been subjected to a validated sterilization process.Meets requirements (SAL = 10^-6)Meets requirements
    Pre-conditioning of needles6 All requirements of the standard related to preconditioning of needles were met.Meets requirementsMeets requirements

    Biocompatibility Testing:

    The device also successfully passed all required biocompatibility tests according to the 2016 FDA guidance Use of International Standard ISO 10993-1. These tests included:

    • Cytotoxicity
    • Sensitization
    • Irritation or Intracutaneous Reactivity
    • Acute Systemic Toxicity
    • ISO Two Week Systemic Toxicity Study in the Rat, Repeated Parenteral Administration of Two Extracts
    • Material-Mediated Pyrogenicity
    • Hemocompatibility
      Reported Biocompatibility Performance: Did not show any adverse biological/biocompatibility reactions.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: The document does not specify the exact number of units tested for each parameter within the non-clinical performance and biocompatibility testing. This information is typically detailed in the full test reports, which are summarized here.
    • Data Provenance: The testing was conducted by HTL-Strefa S.A. based in Poland (ul. Adamówek 7, 95-035 Ozorków, POLAND). The data pertains to non-clinical laboratory testing of the physical device, not patient data, and is thus "prospective" in the sense that the new device was manufactured and then subjected to these tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable for this type of medical device submission. Ground truth for physical device performance is established by standardized testing methods and international standards, not by human expert consensus or annotation of data. The qualifications of the personnel conducting the tests would be subject to internal quality system requirements (e.g., ISO 13485).

    4. Adjudication method for the test set

    • Not applicable for this type of medical device submission. Adjudication methods like 2+1 or 3+1 are used in subjective assessments (e.g., image interpretation) where there might be disagreement among human readers or annotators. For objective physical device tests, results are typically measured and compared against defined thresholds.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a physical hypodermic needle, not an AI-powered device. Therefore, no MRMC study, human reader improvement, or AI assistance is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device does not involve any algorithm.

    7. The type of ground truth used

    • The "ground truth" for this device is defined by the technical specifications and performance requirements outlined in the ISO 11608-2:2012 standard and the established biocompatibility criteria from ISO 10993-1. Performance is measured objectively against these established standards.

    8. The sample size for the training set

    • Not applicable. There is no "training set" as this is a physical device, not an AI model.

    9. How the ground truth for the training set was established

    • Not applicable. There is no "training set" or need for "ground truth establishment" in the AI/ML sense for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1