510(k) Data Aggregation

    K Number
    K202242

    Validate with FDA (Live)

    Device Name
    HS Fiber
    Date Cleared
    2020-09-10

    (31 days)

    Product Code
    Regulation Number
    878.5000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HS Fiber sutures are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular surgery, and the use of allograft tissues for orthopedic procedures.

    Device Description

    The Riverpoint Medical HS Fiber® sutures are non-absorbable, sterile, surgical sutures composed of multiple single strands of ultra-high molecular weight polyethylene (UHMWPE) braided together to form the implant. HS Fiber sutures are available in common sizes and lengths with or without pre-attached needles.

    AI/ML Overview

    This submission is for a medical device (HS Fiber surgical suture) and not an AI/ML powered device, therefore most of the requested information cannot be extracted from the text. The document describes the device, its intended use, and its substantial equivalence to a previously cleared predicate device.

    However, based on the provided text, here is the information related to the acceptance criteria and performance data for the device:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Standard)Reported Device Performance
    United States Pharmacopeia (USP) requirementsMeets USP performance requirements
    USP performance requirements for needle attachmentTested per USP performance requirements for needle attachment
    USP performance requirements for tensile strengthTested per USP performance requirements for tensile strength
    Endotoxin quantities below recommended limits outlined in FDA Guidance "Pyrogens and Endotoxins Testing: Questions and Answers"LAL endotoxin quantification assessments demonstrate endotoxin quantities below recommended limits
    Biological evaluation per ISO 10993-1:2018 - Biological Evaluation of Medical DevicesMaterials used were evaluated per ISO 10993-1:2018

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the exact sample sizes used for the testing. It generally states that the device "meets requirements" and "is tested," but doesn't provide specific numbers of units tested. The provenance of the data is not mentioned (e.g., country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the submission is for a physical medical device (suture) and not an AI/ML powered device that would require expert-established ground truth for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as the submission is for a physical medical device (suture). Testing involved adherence to established physical and biological standards, not image or data interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable as the submission is for a physical medical device (suture) and not an AI/ML powered device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the submission is for a physical medical device (suture).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the device's performance is based on established, objective, and quantitative performance standards set by:

    • United States Pharmacopeia (USP) standards for physical properties like tensile strength and needle attachment.
    • ISO 10993-1:2018 (Biological Evaluation of Medical Devices) for biocompatibility.
    • FDA Guidance "Pyrogens and Endotoxins Testing: Questions and Answers" for endotoxin levels.

    8. The sample size for the training set

    This is not applicable as the submission is for a physical medical device (suture) and not an AI/ML powered device that requires a training set. The device is manufactured according to specifications and tested for compliance.

    9. How the ground truth for the training set was established

    This is not applicable as the submission is for a physical medical device (suture) and not an AI/ML powered device.