510(k) Data Aggregation

    K Number
    K201936

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-05-06

    (297 days)

    Product Code
    Regulation Number
    880.5025
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SmartSite™ Bag is an empty container used for administration of intravenous solutions to the patient using an intravascular administration set. Medication transfer in and out of the container is done using aseptic technique.

    Device Description

    The SmartSite™ Bag is a single-use, empty IV container, which will be available in 100mL, 250mL, and 500mL product sizes. The SmartSite™ Bag is labeled as sterile fluid path, and contains both an add port and a spike port. The add port facilitates aseptic transfer of medication(s) into or out of the bag. The spike port contains a twist-off protective seal. Either the add port or the spike port can be connected to an intravenous (IV) administration set for medication delivery to the patient. The add port may be repeatedly accessed in accordance with the Directions for Use (DFU). The SmartSite™ Bag may be used for up to 24 hours after the initial access of the add port, consistent with prescribing information for the medications used.

    AI/ML Overview

    The provided text describes the 510(k) submission for the SmartSite Bag, an empty IV container. It outlines the device's indications for use, comparison to a predicate device, and non-clinical tests conducted to demonstrate substantial equivalence. However, the document does NOT contain a formal "study" proving the device meets acceptance criteria in the sense of a clinical trial or a structured performance study with defined acceptance criteria and reported numerical performance measures like sensitivity, specificity, or accuracy for a diagnostic device.

    Instead, the document details various engineering and biocompatibility tests designed to ensure the SmartSite Bag functions as intended and is safe for its stated use, thereby demonstrating equivalence to the predicate device. The acceptance criteria are implicitly met by passing these tests and standards.

    Here's a breakdown of the information that is present, organized as requested where applicable, and noting where the requested information is not available in the given text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present explicit acceptance criteria in a table format with corresponding numerical performance for the device for a specific clinical outcome. Instead, it lists various non-clinical tests and standards the device was evaluated against, implying that meeting these standards constitutes "acceptance."

    Test CategorySpecific Test/StandardAcceptance Criteria (Implied by standard compliance)Reported Device Performance
    BiocompatibilityISO 10993-1 (Evaluation and testing within a risk management process)Acceptable biological risk demonstrated by meeting ISO 10993 requirements for contact characterization."Acceptable biological risk established by demonstrating that the device meets ISO 10993."
    - Extractable and Leachables testing and toxicological risk evaluationCompliance with toxicological risk assessment.Passed
    - CytotoxicityNon-cytotoxic.Passed
    - SensitizationNon-sensitizing.Passed
    - Irritation or Intracutaneous ReactivityNon-irritating.Passed
    - Acute Systemic ToxicityNo acute systemic toxic effects.Passed
    - Subacute/Subchronic ToxicityNo subacute/subchronic toxic effects.Passed
    - Material-Mediated PyrogenicityNon-pyrogenic related to materials.Passed
    - HemocompatibilityBiocompatible with blood.Passed
    Bacterial EndotoxinBacterial Endotoxin testingEndotoxin requirements of <0.5EU/mL.Meets endotoxin requirements.
    SterilizationANSI/AAMI/ISO 11137-1:2006/(R) 2015&A1 2013 (Sterilization of Health Care Products - Radiation - Part 1)Sterility Assurance Level (SAL) of 10^-6."Sterilized using irradiation in accordance with a validated sterilization cycle." (SAL of 10^-6 reported as equivalent to predicate)
    ANSI/AAMI/ISO 11137-2:2013/(R) 2019 (Sterilization of Health Care Products - Radiation - Part 2)Establishment of sterilization dose.Referenced for validation.
    ANSI/AAMI/ISO TIR 13004:2013 (Sterilization of health care products - Radiation - Substantiation of selected sterilization dose: Method VDmaxSD)Substantiation of selected sterilization dose.Referenced for validation.
    Performance TestingISO 15747 (2018) Plastic containers for intravenous injectionsCompliance with relevant sections for plastic containers for IV injections."Tested to verify compliance with relevant sections."
    Particulate TestingUSP <788>Meets particulate requirements."Tested to demonstrate the product meets particulate requirements."
    Additional Testing- Maintenance of performance requirements after infusate exposureMaintained performance after exposure.Passed
    - Luer connection performanceAcceptable Luer connection performance.Passed
    - Bag integrity (leakage)No leakage.Passed
    - Performance after simulated shippingMaintained performance after simulated shipping.Passed
    - Shelf lifeMaintained performance for specified shelf life.Passed
    - SmartSite Microbial Ingress TestingNo microbial ingress.Passed

    2. Sample size used for the test set and the data provenance
    The document does not provide specific sample sizes for each test in a numerical format (e.g., "n=X bags"). It refers to "testing" performed. The data provenance is internal to Gilero, LLC as it refers to non-clinical laboratory testing. There is no mention of country of origin of data or whether it was retrospective or prospective in a clinical sense, as these are non-clinical engineering tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This information is not applicable and not available in the document. The tests described are non-clinical, laboratory-based evaluations against established standards (e.g., ISO, USP), not assessments requiring expert "ground truth" establishment in a diagnostic context.

    4. Adjudication method for the test set
    Not applicable. There is no mention of an adjudication method as the tests are objective, laboratory-based measurements against defined standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This is a medical device (IV bag), not an AI/diagnostic software.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    Not applicable. This is a medical device (IV bag), not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    The "ground truth" for the non-clinical tests is based on established industry standards and specifications (e.g., ISO 10993, USP <788>, ISO 15747, and internal performance specifications for factors like leakage, Luer connection, and shelf life).

    8. The sample size for the training set
    Not applicable. This is a traditional medical device, not an AI/machine learning product that requires a training set.

    9. How the ground truth for the training set was established
    Not applicable. There is no training set for this type of device.