510(k) Data Aggregation

    K Number
    K201612

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-09-02

    (444 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Elegant™ 2 in 1 Vaginal Moisturizer/ Elegant™ Advanced 5 is a personal lubricant for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with polyurethane condoms. This product is not compatible with natural rubber latex and polyisoprene condoms.

    Device Description

    The subject device is a non-sterile, water-based, personal lubricant that provides lubrication during intimate sexual activity. The formulation contains purified water, polycarbophil, carbomer homopolymer type B, labrafac, emulfree P, sorbic acid and sodium hydroxide. Elegant™ 2 in 1 Vaginal Moisturizer is packaged in an aluminium tube and applied using a reusable plastic applicator marked to deliver 2ml of product. Elegant™ Advanced 5, is packaged in an aluminium tube and applied using a reusable plastic applicator marked to deliver 3ml of product. Elegant™ 2 in 1 Vaginal Moisturizer and Elegant™ Advanced 5 are personal lubricants for over-the-counter (OTC) use.

    AI/ML Overview

    The provided document is a 510(k) summary for the Elegant™ 2 in 1 Vaginal Moisturizer and Elegant™ Advanced 5. This document assesses the substantial equivalence of a new medical device to a legally marketed predicate device. The information required primarily applies to AI/ML medical devices, which this product (a personal lubricant) is not. Therefore, specific details like sample sizes for training/test sets, number of experts for ground truth, adjudication methods, or MRMC studies are not applicable or provided for this type of device.

    However, I can extract the acceptance criteria and a summary of the studies conducted to demonstrate the device meets those criteria, as much as possible from the provided text.

    1. A table of acceptance criteria and the reported device performance

    The document provides "Device Specifications" in Table 1, which serve as the acceptance criteria for the physical and biological properties of the lubricant. It also summarizes the performance studies conducted to demonstrate compliance with these specifications and other safety/effectiveness aspects.

    ParameterAcceptance Criteria (Specification)Reported Device Performance (Summary from studies)
    Physical/Chemical Specifications (from Table 1)
    AppearanceWhite to off-whiteImplied to meet this, as the shelf-life testing demonstrated the device can maintain its specifications.
    Density USP <699>0.9970 - 0.9995 g/cm²Implied to meet this, as the shelf-life testing demonstrated the device can maintain its specifications.
    pH USP <791>2.5-3.5Implied to meet this, as the shelf-life testing demonstrated the device can maintain its specifications.
    Consistency Ph.Eur 2.9.9145 – 200 mmImplied to meet this, as the shelf-life testing demonstrated the device can maintain its specifications.
    Viscosity USP <912>120 - 220 Pa-sImplied to meet this, as the shelf-life testing demonstrated the device can maintain its specifications.
    Osmolality USP <785>15 – 35 mOsm/kgImplied to meet this, as the shelf-life testing demonstrated the device can maintain its specifications.
    Microbiological Specifications (from Table 1)
    Total Aerobic Microbial Count (TAMC) USP <61>≤ 10² CFU/gImplied to meet this, as the shelf-life testing demonstrated the device can maintain its specifications.
    Total combined yeasts/molds count (TYMC) USP <61>≤ 10¹ CFU/gImplied to meet this, as the shelf-life testing demonstrated the device can maintain its specifications.
    Absence of pathogenic organisms: Staphylococcus Aureus, Pseudomonas Aeruginosa and Candida Albicans USP <62>AbsentImplied to meet this, as the shelf-life testing demonstrated the device can maintain its specifications.
    Antimicrobial Effectiveness USP <51>Bacteria: No less than 2.0 log reduction at 14 days and no increase from 14-day count at the 28-day count. Yeast/Mold: No increase from the initial calculated count at 14 and 28 count.Implied to meet this, as the shelf-life testing demonstrated the device can maintain its specifications.
    Biological SafetyNon-cytotoxic, non-irritating, non-sensitizing, and non-systemically toxic.Results of biocompatibility testing (Cytotoxicity, Intra-vaginal tolerance, Penile irritation, Sensitisation, Acute Systemic Toxicity) demonstrated that Elegant™ 2 in 1 Vaginal Moisturizer and Elegant™ Advanced 5 are non-cytotoxic, non-irritating, non-sensitizing, and non-systemically toxic.
    Shelf-LifeMaintain all Table 1 specifications over the stated shelf-life.Long term data at 25°C/60% RH and accelerated data at 35°C/65% RH for the specifications in Table 1 have been generated up to 6 months, supporting an equivalent of a 12-month shelf life.
    Condom CompatibilityCompatible with polyurethane condoms. Not compatible with natural rubber latex and polyisoprene condoms.Testing conducted in accordance with ASTM D7661-10 confirmed the subject device is compatible with polyurethane condoms. It was also found to be not compatible with natural rubber latex and polyisoprene condoms, which is explicitly stated and accepted as a difference from the predicate device that does not raise different questions of safety and effectiveness.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify sample sizes for the biological or chemical tests, which is typical for this type of submission focusing on standard test methodology and results. The studies appear to be laboratory-based and likely prospective, following recognized international and national standards (e.g., ISO, USP, ICH, ASTM). No country of origin for the data is explicitly mentioned, but the submitter is Viramal Limited, located in the United Kingdom.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to a personal lubricant. The "ground truth" for this device type is established by standardized laboratory testing methods (e.g., pH meters for pH, specific assays for microbial counts, etc.), not expert consensus on interpretations.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable to a personal lubricant. Standard laboratory tests do not use adjudication methods in the way clinical studies for diagnostic algorithms do.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable to a personal lubricant. MRMC studies are relevant for medical imaging or similar diagnostic AI/ML devices.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable to a personal lubricant, as it is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance criteria (specifications) of this device is based on objective measurements obtained through standardized laboratory testing methods (e.g., USP monographs for pH, density, viscosity, osmolality, microbial counts; ISO standards for biocompatibility; ASTM standards for condom compatibility).

    8. The sample size for the training set

    This is not applicable to a personal lubricant, as it is not an AI/ML device and does not involve training data.

    9. How the ground truth for the training set was established

    This is not applicable to a personal lubricant.